Comparison of glycosylated hemoglobin levels detected by 3 kinds of analytic instruments
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3969/j.issn.1673-4130.2014.19.044
   		
        
        	
        		- VernacularTitle:3种仪器检测糖化血红蛋白结果分析
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Weijun ZHANG
			        		
			        		;
		        		
		        		
		        		
			        		Shengqi HUANG
			        		
			        		;
		        		
		        		
		        		
			        		Lianmei LUO
			        		
			        		;
		        		
		        		
		        		
			        		Xiujuan LI
			        		
			        		;
		        		
		        		
		        		
			        		Yanfen ZHENG
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		glycosylated hemoglobin;
			        		
			        		
			        		
				        		ion exchange chromatography;
			        		
			        		
			        		
				        		affinity chromatography
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			International Journal of Laboratory Medicine
	            		
	            		 2014;(19):2672-2673
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To compare levels of glycosylated hemoglobin detected by 3 kinds of analytic instruments .Methods 75 samples were measured by D-10 glycosylated hemoglobin automatic analyzer(ionexchange chromatography) ,HA-8160 glycosylated hemoglobin automatic analyzer(affinity chromatography) and 7170A automatic analyzer(immune turbidimetry) .Results were tested by the homogeneity of variance ,the one-way analysis of variance and correlation analysis .Results There was a positive correlation between D-10 glycosylated hemoglobin automatic analyzer and HA-8160 glycosylated hemoglobin automatic analyzer(r2 =0 .996) , the linear equation was Y=0 .953X+0 .519 .There was a positive correlation between D-10 glycosylated hemoglobin automatic ana-lyzer and 7170A automatic analyzer(r2 =0 .996) ,the linear equation was Y=0 .925X+0 .576 .There was a positive correlation be-tween HA-8160 glycosylated hemoglobin automatic analyzer and 7170A automatic analyzer(r2 =0 .998) ,the linear equation was Y=0 .969 X+0 .081 .Conclusion In the premise of quality control in laboratory ,three different instrument can use to detect the level of glycosylated hemoglobin .