Aptamer-Based Microcantilever Sensor for O-ethyl S-[2-(diisopropylamino) ethyl] methylphosphonothiolate, Sarin Detection and Kinetic Analysis
	    		
		   		
		   			
		   		
	    	
    	 
    	10.11895/j.issn.0253-3820.140191
   		
        
        	
        		- VernacularTitle:微悬臂梁适配子传感器检测维埃克斯、沙林及动力学分析
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Zhiwei LIU
			        		
			        		;
		        		
		        		
		        		
			        		Zhaoyang TONG
			        		
			        		;
		        		
		        		
		        		
			        		Lanqun HAO
			        		
			        		;
		        		
		        		
		        		
			        		Bing LIU
			        		
			        		;
		        		
		        		
		        		
			        		Xihui MU
			        		
			        		;
		        		
		        		
		        		
			        		Jinping ZHANG
			        		
			        		;
		        		
		        		
		        		
			        		Chuan GAO
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Microcantilever;
			        		
			        		
			        		
				        		Aptasensor;
			        		
			        		
			        		
				        		O-Ethyl S-[2-( diisopropylamino) ethyl] methylphosphonothiolate;
			        		
			        		
			        		
				        		Sarin;
			        		
			        		
			        		
				        		Kinetic analysis
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Analytical Chemistry
	            		
	            		 2014;(8):1144-1149
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	A new method for O-ethyl S-[2-( diisopropylamino) ethyl] methylphosphonothiolate ( VX) , sarin detection and its kinetic analysis based on piezoresistive microcantilever aptasensor was developed, where VX, sarin aptamers were immobilized on the microcantilever surface by biotin-avidin binding system. A linear relationship between the response voltage and the concentration of VX in the range of 2-60μg/L was obtained. The linear regression equation was △Ue=0. 886C-1. 039 (n=5, R=0. 984, p<0. 001) and the detection limit was 2μg/L ( S/N≥3 ) . A linear relationship between the response voltage and the concentration of sarin in the range of 10-60 μg/L was obtained, the linear regression equation was △Ue=0. 716C-2. 304 ( n=5, R=0 . 996 , p<0 . 001 ) and the detection limit was 10 μg/L ( S/N≥3 ) . The sensor showed no response for O-butyl methylphosphonochloridate, a structural analog of VX and sarin, which indicated high specificity and good anti-interference ability. On this basis, a reaction kinetic model based on receptor-ligand binding and the relationship with output voltage change was established. Response voltage (△Ue ) and response time( t0 ) were obtained from the fitting equation on different concentrations of VX, sarin fitted well with the measured values.