Dosimetric verification for radiotherapy quality audit under reference and non-reference conditions in Jiangsu province
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3760/cma.j.issn.0254-5098.2014.06.014
   		
        
        	
        		- VernacularTitle:江苏省核查放射治疗参考与非参考条件下剂量学参数方法验证
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Jin WANG
			        		
			        		;
		        		
		        		
		        		
			        		Ningle YU
			        		
			        		;
		        		
		        		
		        		
			        		Chunyong YANG
			        		
			        		;
		        		
		        		
		        		
			        		Xiang DU
			        		
			        		;
		        		
		        		
		        		
			        		Wei CHEN
			        		
			        		;
		        		
		        		
		        		
			        		Suming LUO
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Thermoluminescent dosimeter;
			        		
			        		
			        		
				        		Reference and non-reference conditions;
			        		
			        		
			        		
				        		Finger ionization chamber;
			        		
			        		
			        		
				        		Plane parallel ionization chamber;
			        		
			        		
			        		
				        		Radiotherapy dose
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Radiological Medicine and Protection
	            		
	            		 2014;34(6):451-455
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To verify the methodology for auditing dosimetric parameters in reference and non-reference conditions with thermoluminescent dosimeters (TLDs).Methods Under reference and non-reference conditions,the established TLD methods were used to observe the absorbed dose variations with depth,SSD,field size and 45 wedges for 10 photon beams at 5 hospitals.Dosimetric parameters,including doses at Dmax points in axis,on 5 electron beams of 9 MeV were measured.The measurement results were compared between the TLDs and plane parallel ionization chambers.Results For 6 MV photon beams,the relative deviation of between finger ionization chamber method and TLD chips was in the range of-1.7% to 5.4% under on-axis non-reference conditions,and-6.3% to-0.6% under off-axis non-reference conditions,respectively,all within the range of ≤ ± 7% as required by the IAEA.The relative deviation between plane parallel chamber and TLD method was-2.3% to 3.7%,within ± 5% as required by the IAEA.Conclusions It is convenient and feasible to use TLD method for quality audits of dosimetric parameters in radiotherapy.