Verification of dosimetric methodology for auditing radiotherapy quality under non-reference condition in Hubei province
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3760/cma.j.issn.0254-5098.2014.06.016
   		
        
        	
        		- VernacularTitle:湖北省核查放射治疗非参考条件剂量学参数方法的验证
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Xinxing MA
			        		
			        		;
		        		
		        		
		        		
			        		Suming LUO
			        		
			        		;
		        		
		        		
		        		
			        		Zhijian HE
			        		
			        		;
		        		
		        		
		        		
			        		Wenshan ZHOU
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Non-reference condition;
			        		
			        		
			        		
				        		Finger ionization chamber;
			        		
			        		
			        		
				        		Photon beams;
			        		
			        		
			        		
				        		Electron beams;
			        		
			        		
			        		
				        		Absorbed dose
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Radiological Medicine and Protection
	            		
	            		 2014;34(6):461-465
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To verify the reliability of TLD-based quality audit for radiotherapy dosimetry of medical electron accelerator in non-reference condition by monitoring the dose variations from electron beams with different field sizes and 45° wedge and the dose variations from photon beams with different field sizes and source-skin distance.Methods Both TLDs and finger ionization chambers were placed at a depth of 10 cm in water to measure the absorbed dose from photon beams,and also placed at the depth of maximum dose from electron beams under non-reference condition.TLDs were then mailed to National Institute for Radiological Protection,China CDC for further measurement.Results Among the 70 measuring points for photon beams,58 points showed the results with a relative error less than ± 7.0% (IAEA's acceptable deviation:± 7.0%) between TLDs and finger ionization chambers measurements,and the percentage of qualified point numbers was 82.8%.After corrected by P,value,62 points were qualified and the percentage was up to 88.6%.All of the measuring points for electron beams,with the total number of 24,presented a relative error within ± 5.0% (IAEA's acceptable deviation:± 5.0%) between TLDs and finger ioization cylindrical chambers measurements.Conclusions TLD-based quality audit is convenient for determining radiotherapy dosimetric parameters of electron beams in non-reference condition and can improve the accuracy of the measuring parameters in connection with finger chambers.For electron beams of 5 MeV < E0 < 10 MeV,the absorbed dose parameters measured by finger ionization chambers,combined with TLD audit,can help obtain the precise and reliable results.