Determination of Warfarin in Human Plasma by HPLC
	    		
		   		
	    	
    	
    	
   		
        
        	
        		- VernacularTitle:HPLC法测定人血浆中华法林药物浓度
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Xiaohui HUANG
			        		
			        		;
		        		
		        		
		        		
			        		Ping YANG
			        		
			        		;
		        		
		        		
		        		
			        		Yan LIU
			        		
			        		;
		        		
		        		
		        		
			        		Huaiping TIAN
			        		
			        		;
		        		
		        		
		        		
			        		Jian ZHANG
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Warfarin;
			        		
			        		
			        		
				        		Drug concentration detection;
			        		
			        		
			        		
				        		HPLC
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			China Pharmacist
	            		
	            		 2014;(2):199-201
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective:To establish a simple HPLC method for the determination of warfarin in human plasma and analyze the re-sults of 100 monitoring reports. Methods:Warfarin was extracted from the plasma samples by acetonitrile. The extract was separated on a ZORBAX Eclipse XDB-C8 column (150 mm × 4. 6 mm, 5 μm)with a mobile phase consisting of methanol and 0. 1% acetic acid (65∶35) at flow rate of 1. 0 ml·min-1 and detected by a diode array detector. The detection wavelength was at 308nm and the sample size was 20 μl. Results:The calibration curve was linear within the range of 0. 1-5. 12μg·ml-1 . The average recovery and RSD was ranged from 85. 0% to 102. 3% and 1. 2% to 3. 9%(n=6), respectively. Inter-and intra-day RSD were 1. 1%-1. 8% and 1. 6%-2. 9%, respectively. Conclusion:The proposed method is rapid, convenient and accurate in the detection of warfarinin concentration in plasma.