Assessment of aortic stenosis with dual-source CT
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3760/cma.j.issn.1005-1201.2014.06.009
   		
        
        	
        		- VernacularTitle:双源 CT 对主动脉瓣狭窄的诊断价值
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Pufei WANG
			        		
			        		;
		        		
		        		
		        		
			        		Liang LYU
			        		
			        		;
		        		
		        		
		        		
			        		Gang WANG
			        		
			        		;
		        		
		        		
		        		
			        		Lipeng YANG
			        		
			        		;
		        		
		        		
		        		
			        		Bo SHE
			        		
			        		;
		        		
		        		
		        		
			        		Hongfei AN
			        		
			        		;
		        		
		        		
		        		
			        		Ziyi ZHOU
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Aortic valve stenosis;
			        		
			        		
			        		
				        		Tomography,X-ray computed;
			        		
			        		
			        		
				        		Echocardiography
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Radiology
	            		
	            		 2014;(6):472-475
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To discuss the feasibility and accuracy of dual-source CT ( DSCT) in the evaluation of aortic stenosis ( AS) with transthoracic echocardiography ( TTE) as reference.Methods A total of 53 patients who underwent both DSCT and TTE were prospectively evaluated.All of them were assessed by TTE for aortic stenosis.Maximum aortic valve area ( AVA) in systolic phase was measured with DSCT , and was compared to that index obtained from the continuity equation on TTE.The severity of AS was graded as mild , moderate , or severe according to the AVA.Linear regression analysis and Bland-Altman plots were used to compare the AVA measured by using CT and TTE.Agreement on semi-quantitative grades of AS severity between the two methods was tested by using Kappa statistics.Results The mean AVA using DSCT was (1.45 ±0.35 ) cm2 compared to the mean AVA of ( 1.33 ±0.36 ) cm2 using TTE, with a significant correlation between them (r=0.92,P<0.01).Bland-Altman analysis demonstrated good inter-modality consistency between DSCT and TTE.However , DSCT demonstrated a slight overestimation of the AVA compared to TTE.As identified by TTE, there were 53 patients with AS, 13 with mild AS, 21 with moderate AS , and 19 with severe AS.In 3 patients DSCT showed no AS , TTE detected mild AS.In 6 cases, TTE had graded the stenosis as moderate , but the stenosis degree was graded as mild using DSCT.Kappa analysis showed a good agreement between the two methods on semi -quantitative grades of aortic stenosis severity (Kappa=0.75,P<0.01).Conclusion AVA measurements using DSCT is feasible and reasonably accurate for those patients with moderate to severe aortic stenosis.