Measurement and calculation of X-ray small fields' data for Varian accelerator
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3760/cma.j.issn.1004-4221.2012.06.022
   		
        
        	
        		- VernacularTitle:瓦里安加速器6 MV X线小野数据测量与计算比较
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Xuetao WANG
			        		
			        		;
		        		
		        		
		        		
			        		Shaowen CHEN
			        		
			        		;
		        		
		        		
		        		
			        		Zhenhui DAI
			        		
			        		;
		        		
		        		
		        		
			        		Lin ZHU
			        		
			        		;
		        		
		        		
		        		
			        		Xiaowei LIU
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Ionization chamber measurement;
			        		
			        		
			        		
				        		Monte Carlo calculation;
			        		
			        		
			        		
				        		Small field's data
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Radiation Oncology
	            		
	            		 2012;(6):557-559
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective The beam data is compared with those obtained from Monte Carlo (MC)simulation and measurement to investigate their feasibility and reliability for X-ray small fields.MethodsThe beam data,including the total scatter factor (Scp),percentage depth dose (PDD) was acquired byneasurement and calculation with the field size ranging from 0.5 cm × 0.5 cm to 10 cm x 10 cm.The resultswere compared and analyzed.Results All the data is most consistent for the fields size of ≥3.5 cmx 3.5cm,but they are obvious different for the fields size of ≤ 3.0 cm × 3.0 cm.The measurements seem toreliable using the chambers of CC04 and CC13 for the fields size of ≥2.0 cm x 2.0 cm.Conclusions It isdemonstrated that the accurate measurements and calculations of Scp and PDD can be obtained for the fieldssize of ≥2.0 cm ×2.0 cm,but they needed morc rcscarchcs for thc smaller fields.