Evaluation of NT-proBNP assay performance on electro-chemiluminescence immunoassay system
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3760/cma.j.issn.1009-9158.2011.12.025
   		
        
        	
        		- VernacularTitle:电化学发光免疫法检测NT-proBNP的分析性能评价
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Xiuming ZHANG
			        		
			        		;
		        		
		        		
		        		
			        		Dongmei WEN
			        		
			        		;
		        		
		        		
		        		
			        		Weijia WANG
			        		
			        		;
		        		
		        		
		        		
			        		Fei LI
			        		
			        		;
		        		
		        		
		        		
			        		Lijuan KAN
			        		
			        		;
		        		
		        		
		        		
			        		Haizhong YAN
			        		
			        		;
		        		
		        		
		        		
			        		Minghuan SUO
			        		
			        		;
		        		
		        		
		        		
			        		Nengliang OUYANG
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Natriuretic peptide,brain;
			        		
			        		
			        		
				        		Electrochemistry;
			        		
			        		
			        		
				        		Luminescence;
			        		
			        		
			        		
				        		Immunoassay;
			        		
			        		
			        		
				        		Evaluation study
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Laboratory Medicine
	            		
	            		 2011;34(12):1152-1157
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	ObjectiveTo evaluate analytical performance of NT-proBNP on Electro-Chemiluminescence Immunoassay system.MethodsThe precision,accuracy,limit of blank ( LoB ),limit of detection (LoD),functional sensitivity (FS),analytical measure range (AMR),maximal dilution rate,clinical reportable range(CRR) and the analytical anti-interference ability of NT-proBNP were evaluated according to EP documents issued by CLSI and related references.The analytical performance data were compared to quality standards declared by the manusfacturers.According to CLSI C28-A2,80 healthy volunteers,aged from 18 to 74, were chosen and divided into 4 groups on average for biological reference intervals verification.Results The within-run CV and total CV were 1.1% -2.2%and 1.5% -2.9% respectively.The deviations from controls distributed by National Center for Clinical Laboratory and affiliated calibrators were 2.7% -5.9% and 2.7% -7.5%,respectively.The results of LoB,LoD and FS were 2.5,7.8 and 8.8 pg/ml,respectively.AMR was 8 -35 126 pg/ml,and the most suitable dilution rate was 1∶ 2,so the CRR was 9 -70 252 pg/ml.428 μmol/L bilirubin,2 g/L haematoglobin and 2 200 FIU chyle didn't interfere with the NT-proBNP assay.Moreover,almost all the data from different age groups were in the range of biological reference intervals declared by the manusfacturers, except one test data (167 pg/ml).Conclusions The analyticalperformance of NT-proBNP analyzed on Roche Cobas E601electrochemiluminescence immunoassay systemisconsistentwiththestandarlswhichmanufacturershas proclaimed.The establishment of LoD,FS,maximal dilution and CRR for NT-proBNP assay could provide the quality assurance for clinical use and the biological reference intervals declared by manusfacturers could meet the clinical needs.