Determination of sildenafil citrate content by capillary zone electrophoresis
	    		
		   		
	    	
    	
    	
   		
        
        	
        		- VernacularTitle:毛细管区带电泳法测定枸橼酸昔多芬的含量
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Xinrong ZHANG
			        		
			        		;
		        		
		        		
		        		
			        		Yifeng CHAI
			        		
			        		;
		        		
		        		
		        		
			        		Xiangfeng WU
			        		
			        		;
		        		
		        		
		        		
			        		Lili LIU
			        		
			        		;
		        		
		        		
		        		
			        		Genjin YANG
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
            
            
            	- From:
	            		
	            			Academic Journal of Second Military Medical University
	            		
	            		 2001;22(5):427-428
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective: To establish capillary zone electrophoresis method for determination of sildenafil citrate (Viagra) content in its troche. Methods: Using tetrandrine as internal standard(IS), the electrophoretic separation was achieved with 25 mmol/L borate (pH=7.89) running buffer. And a voltage of 14 kV was applied to the 40 cm×75 μm(i.d) capillary. The analytes were introduced into capillary by siphon (1 s) and determined with on-column monitoring at 214 nm. Results:The determination could be completed within 4 min and the minimum concentration of detection was 5 μg/ml.The analytical results of sildenafil citrate samples demonstrated a good linear relationship within the range of 24-480 μg/ml.The relative standard deviations (RSD) of intra-day was less than 1.58% and that of inter-day was less than 2.46%.The present recoveries were between 95%-105%. Conclusion:The CZE method is accurate, simple, rapid and reliable, so it can be applied to the determination of sildenafil citrate content.