Determination of Carbon Monoxide in Seawater by Headspace Analysis
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3724/SP.J.1096.2010.00352
   		
        
        	
        		- VernacularTitle:顶空法测定海水中一氧化碳
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Xiaolan LU
			        		
			        		;
		        		
		        		
		        		
			        		Guipeng YANG
			        		
			        		;
		        		
		        		
		        		
			        		Xiaomeng WANG
			        		
			        		;
		        		
		        		
		        		
			        		Weilei WANG
			        		
			        		;
		        		
		        		
		        		
			        		Chunyan REN
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Carbon monoxide;
			        		
			        		
			        		
				        		Seawater;
			        		
			        		
			        		
				        		Headspace analysis method
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Analytical Chemistry
	            		
	            		 2010;38(3):352-356
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Oceanic carbon monoxide(CO) has been of biogeochemical interest due to its significant role in global carbon cycle and the greenhouse effect. A headspace method coupled with ta3000 trace gas analyzer system for the determination of CO in seawater was developed. The effects of temperature, equilibrium time and water/gas volume ratio on the sensitivity of headspace analysis were studied in detail. The results showed that CO concentrations in seawater were measured successfully by the 50-mL glass-only syringes with a water/gas volume ratio of 44∶ 6 and an equilibrium time of 5 min at 20 ℃ room temperature. Under the optimized conditions, the linear range of concentrations of CO was 0-2.7×10~(-6), r=0.999,p<0.0001. The relative standard deviation of the analysis method was <4.4%, with a detection limit of 0.02 nmol/L. The average recovery of CO was 90.5%. The concentrations of CO in surface waters of the North Yellow Sea were measured using this method and ranged from 0.20 nmol/L to 3.13 nmol/L, indicating that this method can be successfully applied to the detection of the in situ CO concentrations in seawater.