Application value of computed radiography energy subtraction in detecting pulmonary soft-tissue:comparison with direct digital radiography
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3760/cma.j.issn.0254-5098.2010.04.032
   		
        
        	
        		- VernacularTitle:计算机X线摄影能量减影与直接数字摄影对胸部软组织病变检测能力比较
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Menglong ZHANG
			        		
			        		;
		        		
		        		
		        		
			        		Kai LIU
			        		
			        		;
		        		
		        		
		        		
			        		Liangan ZHANG
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Energy subtraction;
			        		
			        		
			        		
				        		Computed radiography ( CR );
			        		
			        		
			        		
				        		Direct digital radiography (DR);
			        		
			        		
			        		
				        		Comparative study
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Radiological Medicine and Protection
	            		
	            		 2010;30(4):483-485
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To compare the efficiency of computed radiography energy subtraction (ES-CR) with direct digital radiography (DR) in detecting the pulmonary soft-tissue signals.Methods The adult chest phantom including TRG was imaged by ES-CR and DR separately and was estimated by 6 radiologists with 5-value-differentiation method. ROC was used to analyze the performance of two systems. Images of DR and ES-CR of 28 patients with small lung cancer were diagnosed by 3 radiologists. Results The area under the ROC curves was 0. 727 in ES-CR and 0.827 in DR (Z= 2.96,P < 0. 05). 28 small lung cancer patients were significantly different between two systems (x2 = 5.14,P <0. 05). Conclusions The true positive proportion of DR might be higher than that of ES-CR. ES-CR is inferior to DR in detecting pulmonary soft-tissue signals.