Accuracy of BIS value for monitoring depth of sevoflurane-nitrous oxide anesthesia in children
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3760/cma.j.issn.0254-1416.2010.10.024
   		
        
        	
        		- VernacularTitle:BIS值监测患儿七氟醚复合氧化亚氮麻醉深度的准确性
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Shoudong PAN
			        		
			        		;
		        		
		        		
		        		
			        		Yi FENG
			        		
			        		;
		        		
		        		
		        		
			        		Xubo MA
			        		
			        		;
		        		
		        		
		        		
			        		Baxian YANG
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Electroencephalography;
			        		
			        		
			        		
				        		Monitoring,intraoperative;
			        		
			        		
			        		
				        		Anesthetics,inhalation;
			        		
			        		
			        		
				        		Nitrous oxide;
			        		
			        		
			        		
				        		Child
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Anesthesiology
	            		
	            		 2010;30(10):1233-1235
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To evaluate the accuracy of BIS value for monitoring the depth of sevoflurane-nitrous oxide in children. Methods Seventy-two ASA Ⅰ or Ⅱ children aged 1-14 yr undergoing abdominal surgery were randomly assigned into 3 groups ( n = 24 each) :sevoflurane group (group S), sevoflurane + 30% N2O group (group SN1 )and sevoflurane + 60% N2O (group SN2 ). Anesthesia was induced with atropine, lidocaine, propofol, rocuronium and remifentanil. After tracheal intubation, the patients were mechanically ventilated. PETCO2 was maintained at 35-45 mm Hg. Anesthesia was maintained with 2.5% sevoflurane combined with 0, 30% and 60% nitrous oxide in group S, SN1 and SN2 respectively. The end-tidal sevoflurane concentration (CETSev) was maintained at 2.5%, 2.0% and 1.5%. Each CETSev was maintained for at least 10 min after the begining of the surgery. Then the CETSev was modified to maintain BIS value at 40-60. BIS value was recorded before anesthesia induction and each stable CETSev. CETSev was recorded at maintaining BIS value of 40-60 for at least 20 min ( C50 ).Results BIS value and C50 were significantly lower in group SN2 than in group S and SN1 ( P < 0.05), while no significant difference was found between group S and SN1 ( P > 0.05). BIS value was negatively correlated with CETSev in all 3 groups (r = -0.736, -0.817, -0.729, P < 0.01).There was no significant difference in the correlation coefficients among the 3 groups ( P > 0.05 ). Conclusion BIS value can accurately reflect the depth of sevoflurane-nitrous oxide anesthesia in children.