The analytical performance assessment of high-sensitivity cardiac troponin T reference intervals investigation
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3760/cma.j.issn.1009-9158.2010.09.003
   		
        
        	
        		- VernacularTitle:高敏感方法检测心肌肌钙蛋白T的分析性能评价
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Lingyan SONG
			        		
			        		;
		        		
		        		
		        		
			        		Jiong WU
			        		
			        		;
		        		
		        		
		        		
			        		Binbin SONG
			        		
			        		;
		        		
		        		
		        		
			        		Wenqi SHAO
			        		
			        		;
		        		
		        		
		        		
			        		Chunyan ZHANG
			        		
			        		;
		        		
		        		
		        		
			        		Beili WANG
			        		
			        		;
		        		
		        		
		        		
			        		Wei GUO
			        		
			        		;
		        		
		        		
		        		
			        		Baishen PAN
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Troponin T;
			        		
			        		
			        		
				        		Reference values;
			        		
			        		
			        		
				        		Diagnostic techniques and procedure;
			        		
			        		
			        		
				        		Evaluation studies
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Laboratory Medicine
	            		
	            		 2010;33(9):814-818
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To assess the analytical performance of hs-cTnT and biological variations in healthy population as well as establish hs-cTnT reference intervals. Methods The serum samples from 100 acute myocardial infraction patients and 474 apparently healthy subjects were collected. The functional sensitivity,within- and between-run imprecision were determined. The hs-cTnT assay and con-cTnT assay were evaluated. The serum hs-cTnT levels were detected in apparently healthy subjects to establish reference intervals. Moreover,the long-term and short-term biological variations for hs-cTnT in healthy volunteers were assessed. Results The functional sensitivity of hs-cTnT was 0. 005 μg/L. The within- and between-run precision for lower level control(0. 014 μg/L) and higher level control(2. 500 μg/L) was 2. 97% vs 3. 64%and 0. 66% vs 1.01% ,respectively. The correlation between hs-cTnT assay and con-cTnT assay was good ( R2 =0. 972 ,P <0. 01 ). The 99th percentile in apparently healthy subjects was 0. 003 μg/L for women less than60 years, 0.008 μg/L for men less than 60 years, 0.015 μg/L for women above 60 years and 0. 021 μg/L for men above 60 years. The CVa, CVi, CVg and CVt of short-term biological variations in detecting hs-cTnT from 22 apparently healthy subjects were 3.8%, 4. 8%, 49.9% and 58.5%,respectively. The CVa,CVi ,CVg and CVt of long-term biological variations were 5. 3% ,6. 4% ,56. 6% and 68. 3% respectively. Conclusions The analytical performance of the hs-cTnT is better than con-cTnT assay,achieving acceptable level according to guideline. Our experimental result could provide the basis for the new high sensitivity cTnT assay in the diagnosis of acute myocardial infarction.