Evaluation of ambulatory oxygen uptake monitor.
	    		
	    			
	    			
		        		
			        		
		        		
			        
		   		
		   		
	    	
    	 
    	
   		
        
        	
        		- VernacularTitle:携帯型酸素摂取量測定装置の評価と応用
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		TAKANAO HIGUCHI
			        		
			        		;
		        		
		        		
		        		
			        		TOSHIYO TAMURA
			        		
			        		;
		        		
		        		
		        		
			        		TATSUO TOGAWA
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Portable device;
			        		
			        		
			        		
				        		Oxygen uptake;
			        		
			        		
			        		
				        		Flow-through system;
			        		
			        		
			        		
				        		Servo-control system;
			        		
			        		
			        		
				        		Respiratory monitoring
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:Japanese Journal of Physical Fitness and Sports Medicine
	            		
	            		 1991;40(2):195-201
	            	
            	
 
            
            
            	- CountryJapan
 
            
            
            	- Language:Japanese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	A portable device for long-term monitoring of oxygen uptake without discomfort has been developed based on a flow-through principle. The oxygen content of the exhaust gas was kept constant by a servo-controlled blower, so that the flow rate was essentially proportional to the oxygen uptake. This system was evaluated by both a bench test and in exercising human subjects. The limiting current-type oxygen sensor used this portable device produced results which were in good agreement those obtained using a zirconia oxygen sensor (r= 0.999), and stable values were obtained for at least 15 h. The response time for a stepped oxygen change was 4 s. The air flow rate produced by the blower was proportional to the supplied voltage. The results using a time-of-flight flowmeter were highly correlated with those for a hot-wire anemometer (r=0, 999) . The oxygen uptake of exercising human subjects measured with this portable device showed good agreement with that obtained by a conventional oxygen monitor (r=0.973 for cycle ergometer, r=0, 932 for treadmill) . This portable device was able to measure oxygen uptake in subjects during daily activities, and also while driving a car, without disturbance.