Recommendation of a highly sensitive method for measuring hemoglobin in hemolytic test.
	    		
		   		
	    	
    	
    	
   		
        
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Wangping HU
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Yingying HU
			        		
			        		;
		        		
		        		
		        		
			        		Fuying FENG
			        		
			        		;
		        		
		        		
		        		
			        		Jinyao HUANG
			        		
			        		;
		        		
		        		
		        		
			        		Rongsheng ZHANG
			        		
			        		;
		        		
		        		
		        		
			        		Rongren CHEN
			        		
			        		;
		        		
		        		
		        		
			        		Changshao ZHOU
			        		
			        		;
		        		
		        		
		        		
			        		Hailin WANG
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. Clinical Chemistry Department,Fuzhou Genenal Hospital, Fuzhou 350025, China. HWP9999@sohu.com
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
            
            	- MeSH:
            	
	        			
	        				
	        				
				        		
					        		Chlorophenols;
				        		
			        		
				        		
					        		Coombs Test;
				        		
			        		
				        		
					        		methods;
				        		
			        		
				        		
					        		Hemoglobins;
				        		
			        		
				        		
					        		analysis;
				        		
			        		
				        		
					        		Hemolysis;
				        		
			        		
				        		
					        		Humans;
				        		
			        		
				        		
					        		Sensitivity and Specificity
				        		
			        		
	        			
	        			
            	
            	
 
            
            
            	- From:
	            		
	            			Journal of Biomedical Engineering
	            		
	            		 2007;24(3):664-666
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	In this paper is recommended a highly sensitive and reagent-safe method to determine plasma heamoglobin (FHb) in viscacha hemolytic test. The 2,4-dichlorophenol method (2,4-DCP) of Trinder reaction has been improved. The performance of 2,4-DCP is verified. The sensitivity of 2,4-DCP is 2.39 times that of phenol method. It is well used with run precision and day-to-day precision. The reaction color is stable. The reference value FHb is 1-36.7 mg/L. Sodium citric is an excellent anticoagulant liquid to keep erythrocyte. The 2,4-DCP method is neither carcinogenic nor poisonous;it is suitable for viscacha hemolytic test in clinical and biomedical engineering.