Evaluation of the performance of the electrochemiluminescent immunoassay system of Elecsys 2010.
	    		
		   		
	    	
    	
    	
   		
        
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Weishi QIN
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. Department of Nuclear Medicine, West China Hospital, Sichuan University, Chengdu
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
            
            	- MeSH:
            	
	        			
	        				
	        				
				        		
					        		Electrochemistry;
				        		
			        		
				        		
					        		instrumentation;
				        		
			        		
				        		
					        		Evaluation Studies as Topic;
				        		
			        		
				        		
					        		Humans;
				        		
			        		
				        		
					        		Immunoassay;
				        		
			        		
				        		
					        		instrumentation;
				        		
			        		
				        		
					        		standards;
				        		
			        		
				        		
					        		Luminescent Measurements;
				        		
			        		
				        		
					        		Sensitivity and Specificity;
				        		
			        		
				        		
					        		Thyrotropin;
				        		
			        		
				        		
					        		blood
				        		
			        		
	        			
	        			
            	
            	
 
            
            
            	- From:
	            		
	            			Journal of Biomedical Engineering
	            		
	            		 2002;19(4):673-675
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Performance of the electrochemiluminescent immunoassay system of Elecsys 2010 was evaluated by quantitative assay. The functional sensitivity of the TSH assay was 0.004 microU/ml. The extra-assay imprecision was RSD = 1.45%-4.65% in the range of TSH concentrations of 0.43-88.15 microU/ml. The intra-assay imprecision was RSD = 1.38%-3.14% in the range of TSH concentrations of 0.41-87.28 microU/ml. The recovery of TSH was in the range of 98.6%-104.4%. The correlation between theoretical values and obtained values of TSH was adequate(r = 0.999). We conclude that the Elecsys 2010 is provably satisfactory in sensitivity, precision, accuracy, reliability, practicability and rapidity.