Study on MSO/GO-based determination method for trace amount of aqueous Hg2+.
- Author:
	        		
		        		
		        		
			        		Tao SUN
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Jue HE
			        		
			        		;
		        		
		        		
		        		
			        		Xu-wu XIANG
			        		
			        		;
		        		
		        		
		        		
			        		Xiao-lan HONG
			        		
			        		;
		        		
		        		
		        		
			        		Xiao-yan YAO
			        		
			        		;
		        		
		        		
		        		
			        		Lin-zhi ZHANG
			        		
			        		;
		        		
		        		
		        		
			        		Yi-yun WANG
			        		
			        		;
		        		
		        		
		        		
			        		Yan-ne XIE
			        		
			        		;
		        		
		        		
		        		
			        		Wen-he WU
			        		
			        		;
		        		
		        		
		        		
			        		Jian-xin LU
			        		
			        		
		        		
		        		
		        		
			        		
			        		Author Information
			        		
 - Publication Type:Journal Article
 - MeSH: Biosensing Techniques; Fluorometry; Graphite; Mercury; analysis; Nanotechnology; Oligonucleotide Probes; Water
 - From: Chinese Journal of Industrial Hygiene and Occupational Diseases 2013;31(10):783-786
 - CountryChina
 - Language:Chinese
 - 
		        	Abstract:
			       	
			       		
				        
				        	
OBJECTIVETo establish a highly sensitive fluorometric nanobiosensor for determination of aqueous mercury ions (Hg(2+)) using optimized mercury-specific oligonucleotide (MSO) probes and graphene oxide (GO).
METHODSThe nanobiosensor was assembled by attaching the self-designed MSO(1) (5' end labeled with fluorophore carboxyfluorescein (FAM), denoted as FAM-MSO(1)) and MSO(2) to the surface of GO through strong non-covalent bonding forces. Upon the addition of Hg(2+), the formation of the T-Hg(2+)-T configuration desorbed the FAM-MSO(1) and MSO(2) from the surface of GO, resulting in a restoration of the fluorescence of FAM-MSO(1). Using the specific mispairing of T-Hg(2+)-T and the changes in fluorescent signals in solutions, quantitative analysis of Hg(2+) could be performed.
RESULTSThe average thickness of the prepared GO sheets was only 1.4 nm. For the Hg(2+) nanobiosensor, the optimum concentrations of FAM-MSO(1) and MSO(2) were both 1 µmol/L, the optimum volume of 0.5 g/L GO was 5 µL, and the limit of detection was 10 pmol/L; it had low cross-reactivity with 10 other kinds of non-specific metal ions; the fluorescence recovery efficiency was up to 65% in the re-determination of Hg(2+) after addition of Na(2)S(2)O(3).