Preparation and evaluation of a time-resolved fluoroimmunoassay kit for CA50 determination.
- Author:
	        		
		        		
		        		
			        		Jian-feng HANG
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Ying-song WU
			        		
			        		;
		        		
		        		
		        		
			        		Wei-wen XU
			        		
			        		;
		        		
		        		
		        		
			        		Wei-hong YU
			        		
			        		;
		        		
		        		
		        		
			        		Ying HUANG
			        		
			        		;
		        		
		        		
		        		
			        		Ming LI
			        		
			        		
		        		
		        		
		        		
			        		
			        		Author Information
			        		
 - Publication Type:Journal Article
 - MeSH: Adolescent; Adult; Aged; Aged, 80 and over; Antigens, Tumor-Associated, Carbohydrate; blood; Female; Fluoroimmunoassay; instrumentation; methods; Humans; Male; Middle Aged; Radioimmunoassay; Reagent Kits, Diagnostic; standards; Reference Standards; Reproducibility of Results
 - From: Chinese Journal of Oncology 2007;29(3):189-192
 - CountryChina
 - Language:Chinese
 - 
		        	Abstract:
			       	
			       		
				        
				        	
OBJECTIVETo prepare a rapid and sensitive diagnostic kit for detection of CA50 based on time-resolved fluoroimmunoassay.
METHODSA sandwich-TRFIA diagmostic kit was developed using anti-CA50 monoclonal antibody and all parameters of the kit were evaluated.
RESULTSThe linear measurement range of the kit was (5 - 300) U/ml. The sensitivity was 0.2 U/ml. The intra- and inter-assay coefficients of variation were 4.3% - 8.2% and 7.7% - 11.2%, respectively. There was no cross-reaction with CEA, CA12-5, CA15-3 and AFP. The cross reactivity with CA19-9 was 0.7 U/ml. The correlation coefficient of detection results of 107 blood samples between this newly developed kit and commercially available CA50 RIA kit was 0.901.
CONCLUSIONThis newly developed CA50-TRFIA kit is a valuable test tool for clinical application with even better sensitivity, specificity and accuracy.
 
            