Monopolar Radiofrequency Ablation Using a Dual-Switching System and a Separable Clustered Electrode: Evaluation of the In Vivo Efficiency.
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3348/kjr.2014.15.2.235
   		
        
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Jeong Hee YOON
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Jeong Min LEE
			        		
			        		;
		        		
		        		
		        		
			        		Eui Jin HWANG
			        		
			        		;
		        		
		        		
		        		
			        		In Pyung HWANG
			        		
			        		;
		        		
		        		
		        		
			        		Jeehyun BAEK
			        		
			        		;
		        		
		        		
		        		
			        		Joon Koo HAN
			        		
			        		;
		        		
		        		
		        		
			        		Byung Ihn CHOI
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. Department of Radiology, Seoul National University Hospital, Seoul 110-744, Korea. jmsh@snu.ac.kr
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Original Article ; Evaluation Studies ; Research Support, Non-U.S. Gov't
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Liver;
			        		
			        		
			        		
				        		Interventional procedures;
			        		
			        		
			        		
				        		Thermal ablation;
			        		
			        		
			        		
				        		Radiofrequency ablation;
			        		
			        		
			        		
				        		Experimental study
			        		
			        		
	        			
        			
        		
 
        	
            
            	- MeSH:
            	
	        			
	        				
	        				
				        		
					        		Animals;
				        		
			        		
				        		
					        		Catheter Ablation/*instrumentation/*methods;
				        		
			        		
				        		
					        		*Electrodes;
				        		
			        		
				        		
					        		Feasibility Studies;
				        		
			        		
				        		
					        		Liver/*surgery;
				        		
			        		
				        		
					        		Male;
				        		
			        		
				        		
					        		Sus scrofa;
				        		
			        		
				        		
					        		Time Factors
				        		
			        		
	        			
	        			
            	
            	
 
            
            
            	- From:Korean Journal of Radiology
	            		
	            		 2014;15(2):235-244
	            	
            	
 
            
            
            	- CountryRepublic of Korea
 
            
            
            	- Language:English
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	OBJECTIVE: To determine the in vivo efficiency of monopolar radiofrequency ablation (RFA) using a dual-switching (DS) system and a separable clustered (SC) electrode to create coagulation in swine liver. MATERIALS AND METHODS: Thirty-three ablation zones were created in nine pigs using a DS system and an SC electrode in the switching monopolar mode. The pigs were divided into two groups for two experiments: 1) preliminary experiments (n = 3) to identify the optimal inter-electrode distances (IEDs) for dual-switching monopolar (DSM)-RFA, and 2) main experiments (n = 6) to compare the in vivo efficiency of DSM-RFA with that of a single-switching monopolar (SSM)-RFA. RF energy was alternatively applied to one of the three electrodes (SSM-RFA) or concurrently applied to a pair of electrodes (DSM-RFA) for 12 minutes in in vivo porcine livers. The delivered RFA energy and the shapes and dimensions of the coagulation areas were compared between the two groups. RESULTS: No pig died during RFA. The ideal IEDs for creating round or oval coagulation area using the DSM-RFA were 2.0 and 2.5 cm. DSM-RFA allowed more efficient RF energy delivery than SSM-RFA at the given time (23.0 +/- 4.0 kcal vs. 16.92 +/- 2.0 kcal, respectively; p = 0.0005). DSM-RFA created a significantly larger coagulation volume than SSM-RFA (40.4 +/- 16.4 cm3 vs. 20.8 +/- 10.7 cm3; p < 0.001). Both groups showed similar circularity of the ablation zones (p = 0.29). CONCLUSION: Dual-switching monopolar-radiofrequency ablation using an SC electrode is feasible and can create larger ablation zones than SSM-RFA as it allows more RF energy delivery at a given time.