Dosimetric Characteristics on Penumbra Regions of the multileaf Collimator as Compared with the Lead Alloy Block.
	    		
		   		
	    	
    	
    	
   		
        
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Sang Wook LEE
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Young Tack OH
			        		
			        		;
		        		
		        		
		        		
			        		Woo Cheol KIM
			        		
			        		;
		        		
		        		
		        		
			        		Ki Chang KEUM
			        		
			        		;
		        		
		        		
		        		
			        		Seong Ick YOON
			        		
			        		;
		        		
		        		
		        		
			        		Hyun Soo KIM
			        		
			        		;
		        		
		        		
		        		
			        		Won PARK
			        		
			        		;
		        		
		        		
		        		
			        		Seong Sil CHU
			        		
			        		;
		        		
		        		
		        		
			        		Gwi Eon KIM
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. Department of Radiation Oncology, Yonsei University College of Medicine, Seoul, Korea.
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Original Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Multileaf collimator;
			        		
			        		
			        		
				        		Effective penumbra;
			        		
			        		
			        		
				        		Film dosimetry
			        		
			        		
	        			
        			
        		
 
        	
            
            	- MeSH:
            	
	        			
	        				
	        				
				        		
					        		Alloys*;
				        		
			        		
				        		
					        		Film Dosimetry;
				        		
			        		
				        		
					        		Pectinidae;
				        		
			        		
				        		
					        		Radiotherapy, Conformal
				        		
			        		
	        			
	        			
            	
            	
 
            
            
            	- From:Journal of the Korean Society for Therapeutic Radiology
	            		
	            		 1995;13(4):391-396
	            	
            	
 
            
            
            	- CountryRepublic of Korea
 
            
            
            	- Language:Korean
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	PURPOSE: The Conformal Radiation Therapy has been widely used under favour of development of computer technologies. The delivery of a large number of static radiation fields are being necessary for the conformal irradiation. In this paper, we investigate dosimetric characteristics on penumbra regions of a multileaf collimator(MLC), and compare to those of lead alloy block for he optimal use of the system in 3-D conformal radiotherapy. MATERIALS AND METHODS: The measurement of penumbra by MLC or lead alloy block was performed with 6 or 10 MV X-rays. The film was positioned at a dmax depth and 10 cm depth, and its optical density was determined using a scanning videodensitometer. The effective penumbra, the distance from 80% to 20% isodose lines and 90 to 10 were analyzed as a function of the angle between the direction of leaf motion and the edge defined by leaves. RESULTS: Increasing MLC angle (0-75degree) was observed with increasing the penumbra widths and the scalloping effect. There was no definite differences of penumbra width from 80% to 20% isodose lines, while being the small increase of penumbra width of lead alloy block are agree reasonably with those of MLC within 4.8mm. CONCLUSION: The comparative qualitative study of the penumbra between MLC and lead alloy block demonstrate the clinical acceptability and suitability of the multileaf collimator for 3-D conformal radiotherapy.