Accuracy of Three Intraocular Lens-power Formulas in Predicting Refractive Outcomes in Different Intraocular Lenses.
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3341/jkos.2016.57.12.1891
   		
        
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Sung Il KANG
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Kun MOON
			        		
			        		;
		        		
		        		
		        		
			        		Jong Hwa JUN
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. Department of Ophthalmology, Keimyung University School of Medicine, Daegu, Korea. junjonghwa@gmail.com
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Original Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Intraocular lenses;
			        		
			        		
			        		
				        		Intraocular lens power formula;
			        		
			        		
			        		
				        		Mean absolute error;
			        		
			        		
			        		
				        		Mean numeric error
			        		
			        		
	        			
        			
        		
 
        	
            
            	- MeSH:
            	
	        			
	        				
	        				
				        		
					        		Biometry;
				        		
			        		
				        		
					        		Cataract;
				        		
			        		
				        		
					        		Lenses, Intraocular*;
				        		
			        		
				        		
					        		Medical Records;
				        		
			        		
				        		
					        		Refractometry;
				        		
			        		
				        		
					        		Retrospective Studies
				        		
			        		
	        			
	        			
            	
            	
 
            
            
            	- From:Journal of the Korean Ophthalmological Society
	            		
	            		 2016;57(12):1891-1896
	            	
            	
 
            
            
            	- CountryRepublic of Korea
 
            
            
            	- Language:Korean
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	PURPOSE: To compare the accuracy of different power-calculation formulas in predicting the postoperative refraction of three-piece and one-piece intraocular lenses (IOL). METHODS: We retrospectively reviewed the medical records of 74 eyes (62 patients) that had undergone cataract surgery involving implantation of one of two IOLs―the SENSAR® AAB00 1-Piece Acrylic IOL (44 eyes), or the Hoya® VA60BB 3-Piece Acrylic IOL (30 eyes)―between October 2014 and March 2015. Axial length was measured using an optical low-coherence refractometry (Lenstar®), and biometry was then calculated by the pre-installed Lenstar program, which used the SRK/II, Sanders-Retzlaff-Kraff/Theoretical (SRK/T), and Hoffer Q formulas. Mean absolute error (MAE) and mean numeric error (MNE) were measured 1 day, 1 week, 1 month, and 2 months after surgery. RESULTS: Using the SRK/T and Hoffer Q formulas, the one-piece IOL group differed significantly from the three-piece IOL group in terms of the MNE obtained 1 month and 2 months after surgery. Across all formulas and time points, there were no significant differences between the groups in terms of MAE. CONCLUSIONS: There was no significant difference between the different power-calculation formulas. Starting 1 month after surgery, the three-piece IOL group showed myopic postoperative refraction compared to the predictive spherical equivalent using the SRK/T and Hoffer Q formulas.