Analysis of the effect of dosimeter wearing position on effective dose estimation among interventional radiology workers
10.13491/j.issn.1004-714X.2025.05.011
- VernacularTitle:介入放射工作人员剂量计佩戴位置对有效剂量评估的影响分析
- Author:
Xuanrong ZHANG
1
;
Wen GUO
1
;
Xian XUE
1
;
Pin GAO
1
;
Kaiyi WANG
1
;
Xuan ZHANG
2
;
Yanqiu DING
1
;
Xiao LUO
3
;
Wenfang MENG
3
;
Jun CHAO
3
Author Information
1. Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China.
2. Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China;National Institude of Metrology, Beijing 100029, China.
3. Hainan Provincial Center for Disease Control and Prevention, Haikou 570203, China.
- Publication Type:OriginalArticles
- Keywords:
Interventional radiology worker;
Personal dose monitoring;
Effective dose assessment method;
Wearing position
- From:
Chinese Journal of Radiological Health
2025;34(5):687-694
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the influence of the wearing position of dosimeters outside lead aprons on effective dose estimation for interventional radiology workers, analyze the differences between single and double dosimeter methods in effective dose estimation, and provide a reference for the personal dose monitoring of interventional radiology workers. Methods This study employed a combined approach of on-site monitoring and Monte Carlo simulation to evaluate the impact of the wearing position of dosimeters outside lead aprons on effective dose estimation, as well as the differences between effective doses measured using single and double dosimeters. Interventional radiology workers wore dosimeters at three positions: the neck outside the lead collar, the left chest outside the lead apron, and inside the lead apron. Effective doses were estimated using the single and double dosimeter methods specified in GBZ 128-2019 Specifications for individual monitoring of occupational external exposure, and the impact of different wearing positions on the estimation results was compared. Geant4 Monte Carlo simulations were used to model dose distributions at the neck outside the lead collar and at the left chest outside the lead apron for operators performing cardiovascular interventions under tube voltages of 70, 80, 90, and 100 kVp and exposure angles of posteroanterior (PA), anteroposterior (AP), and left anterior oblique 45° (LAO45°) positions. The study assessed the impact of dosimeter wearing position on effective dose estimation. Results Monte Carlo simulations demonstrated that neck doses consistently exceeded left chest doses across different tube voltages and exposure angles, with neck-to-chest dose ratios of 0.80-0.90. Under identical tube voltage conditions, AP showed the highest doses, followed by LAO45°, and PA demonstrated the lowest doses. The single and double dosimeter methods exhibited consistent patterns in effective dose estimation. Single dosimeter method generally yielded higher effective doses with relative deviations of 9.9% to 83%, though these deviations decreased under high tube voltages. Field monitoring data indicated that most interventional radiology workers maintained relative deviations between single and double dosimeter calculations below 6%, with neck-to-chest dose ratios of 0.95-1.1. The estimation patterns remained consistent across both methods, though single dosimeter method showed slightly higher results. Conclusion Under PA, AP, or LAO45°, the doses at the neck consistently exceeded those at the left chest. Therefore, when wearing lead protective equipment, the dosimeter should be properly positioned at the neck outside the lead collar to accurately reflect the radiation doses of surgeons. Some interventional radiology workers improperly positioned the dosimeter (intended at the neck outside the lead collar) at the left chest outside the lead apron, and this may result in an underestimation of the effective dose.