Analysis of structure principle and common fault maintenance of Allura XPer FD20 digital subtraction angiography equipment
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3969/j.issn.1672-8270.2024.08.036
   		
        
        	
        		- VernacularTitle:Allura XPer FD20型数字减影血管造影设备结构原理分析及故障维修
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Lingling QU
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Jian LI
			        		
			        		;
		        		
		        		
		        		
			        		Fang QIAO
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. 上海健康医学院附属周浦医院采购办公室 上海 201318
			        		
		        		
	        		
        		 
        	
        	
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Digital subtraction angiography(DSA);
			        		
			        		
			        		
				        		Attenuation coefficient;
			        		
			        		
			        		
				        		Flat panel detector;
			        		
			        		
			        		
				        		Driving circuit;
			        		
			        		
			        		
				        		Temperature control
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			China Medical Equipment
	            		
	            		 2024;21(8):194-197
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Mastering the structure,composition and working principle of digital subtraction angiography(DSA)equipment is an important prerequisite for medical engineers to carry out daily maintenance and repair management.By analyzing the structural composition and functional components of DSA from the aspects of X-ray source,mechanical drive,host control and image output,the working process of subtraction imaging was described according to Beer-Lambert law,and the influencing factors of DSA image quality were summarized.At the same time,taking the Allura XPer FD20 DSA equipment as an example,the failure cases of examination bed,rack and temperature control that occurred in clinical use were analyzed,and the fault investigation and processing were carried out from the hardware structure and circuit module of the equipment,and the equipment fault alarm information was solved in time to ensure the operation quality of the equipment and improve the efficiency of fault maintenance.