Determination of 24 elemental impurities in dizocine injection by ICP-MS
	    		
		   		
		   			
		   		
	    	
    	 
    	10.19778/j.chp.2024.02.008
   		
        
        	
        		- VernacularTitle:ICP-MS测定地佐辛注射液中24种元素杂质
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Xiaoqing HUANG
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Yibo TAO
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. 泰州市药品检验院,江苏泰州 225300
			        		
		        		
	        		
        		 
        	
        	
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		dizocine injection;
			        		
			        		
			        		
				        		ICH Q3D;
			        		
			        		
			        		
				        		ICP-MS;
			        		
			        		
			        		
				        		elemental impurities;
			        		
			        		
			        		
				        		quality control
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Drug Standards of China
	            		
	            		 2024;25(2):159-163
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective:To establish a method of inductively coupled plasma mass spectrometry(ICP-MS)for the simultaneous determination of 24 elemental impurities in dizocine injection in accordance with the relevant regula-tions of ICH Q3D.Methods:ICP-MS method was adopted to analyze the sample solutions after being diluted and then corrected by spiking indium and bismuth as internal standards,the plasma output power was set at 1 550 W,the sampling depth was set at 5 mm,the maximum load volume was set at 1.038 L·min-1,the cooling gas was set at 14 L·min-1,the auxiliary gas was 0.8 L·min-1,the peristaltic pump speed was 40 r·min-1,the injection system was hydrofluoric acid resistant system,the sampling cone and the pattern of the intercept cone were nickel cone and the sampling mode was manual and repeated 3 times.Results:All 24 elements were linear(all r were more than 0.99),the average recoveries were between 80.46%and 102.95%,the repeatability were less than 10.88%,and the intermediate precision were not more than 6.49%.The results of 24 elemental impurities in 3 different batches of dizocine injection met requirements of ICH Q3D.Conclusion:The ICP-MS method is conven-ient,rapid,sensitive and accurate,and can be used to detect the concentrations of 24 metal elements in dizocine injection,which may provide a basis for more strict quality management of the injection.