The influence of individualized low-dose CT imaging parameters on liver fat content measurement by quantitative CT
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3969/j.issn.1002-1671.2024.10.032
   		
        
        	
        		- VernacularTitle:个体化低剂量CT成像参数对定量CT肝脏脂肪含量测量的影响
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Shiyu WANG
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Yijun LIU
			        		
			        		;
		        		
		        		
		        		
			        		Wei WEI
			        		
			        		;
		        		
		        		
		        		
			        		Beibei LI
			        		
			        		;
		        		
		        		
		        		
			        		Xu WANG
			        		
			        		;
		        		
		        		
		        		
			        		Xiaoyu TONG
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. 大连医科大学附属第一医院放射科,辽宁 大连 116011
			        		
		        		
	        		
        		 
        	
        	
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		fat content;
			        		
			        		
			        		
				        		automatic tube voltage modulation;
			        		
			        		
			        		
				        		computed tomography;
			        		
			        		
			        		
				        		radiation dose;
			        		
			        		
			        		
				        		image quality
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Journal of Practical Radiology
	            		
	            		 2024;40(10):1717-1721
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To investigate the influence of individualized low-dose CT imaging parameters on quantitative computed tomography(QCT)liver fat content measurement and image quality by using the combined technique of Auto-prescription and adaptive statistical iterative reconstruction(ASIR-V)algorithm with different weights.Methods A total of 231 patients who underwent both chest and abdominal CT scans were prospectively selected.The overlapping liver of two-position scans in the same case was studied,in which the chest was scanned with an individualized low-dose protocol(group A)and the upper abdomen was scanned with a routine 120 kVp protocol(group B).Patients with group A were scanned by using 80 kVp and 100 kVp based on Auto-prescription technique,and divided into the subgroups of A1 and A2;meanwhile the same patients with group B were scanned by using 120 kVp and divided into the subgroups of B1 and B2.The images were transferred to AW4.6 and QCTpro workstations for measuring CT values,standard deviation(SD)values and Fat%QCT values of the left lobe,right anterior lobe,and right posterior lobe of the liver in the overlapping regions.The signal-to-noise ratio(SNR)and contrast-to-noise ratio(CNR)were also calculated using the same-level erector spinae as background.Paired sample t-test was employed to compare the differences in Fat%QCT measurements between groups A and B.Spearman correlation analysis was conducted,and linear regression equations were established.Two observers evaluated image quality using a five-point scale and Kappa test was conducted to test inter-observer consistency.Results In group A,the Fat%QCT values showed no statistically significant difference among different weights of ASIR-V.However,the statistically significant difference was found between groups A and B(P<0.05),and there were highly positive correlations between subgroups A1 and B1 and between subgroups A2 and B2(r=0.939 7,0.987 2,P<0.05).The linear regression equation under 80 kVp and 100 kVp were as follows:y=0.976x+3.119,y=1.007x+2.041.The SNR and CNR of group A combined with post-60%ASIR-V were higher than those of group B combined with post-40%ASIR-V under conventional tube voltage.The subjective scores between groups A and B had no statistically significant difference.Conclusion Low-dose chest imaging based on Auto-prescription technique combined with post-60%ASIR-V can not only satisfy clinical diagnostic requirement,but also realize quantitative analysis of Fat%QCT under low tube voltage(80 kVp/100 kVp).