Graphite furnace atomic absorption spectrometry for determination of thallium in blood
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3760/cma.j.issn.1001-9391.2016.04.019
   		
        
        	
        		- VernacularTitle:血中铊测定的石墨炉原子吸收光谱法
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Qinlong ZHANG
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Ge GAO
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. 610041,成都市疾病预防控制中心
			        		
		        		
	        		
        		 
        	
        	
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Colloidal palladium;
			        		
			        		
			        		
				        		Spectrometry,atomic absorption;
			        		
			        		
			        		
				        		Blood;
			        		
			        		
			        		
				        		Thallium
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Industrial Hygiene and Occupational Diseases
	            		
	            		 2016;34(4):302-304
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective Colloidal palladium was used as chemical modifier in the determination of blood thallium by graphite furnace atomic absorption spectrometry.Methods Blood samples were precipitated with 5% (V/V) nitric acid,and then determined by GFAAS with colloidal palladium used as a chemical modifier.0.2% (W/V) sodium chloride was added in the standard series to improve the matrix matching between standard solution and sample.Results The detection limit was 0.2 μg/L.The correlation coefficient was 0.9991.The recoveries were between 93.9% to 101.5%.The relative standard deviations were between 1.8% to 2.7%.The certified reference material of whole blood thallium was determined and the result was within the reference range Conclusion The method is accurate,simple and sensitive,and it can meet the needs of detection thallium in blood entirely.