Design of Pulse and Respiratory Signals Detection System
	    		
		   		
		   			
		   		
	    	
    	 
    	10.12455/j.issn.1671-7104.230288
   		
        
        	
        		- VernacularTitle:脉搏与呼吸信号检测系统设计
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Rui ZHANG
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Lidong XING
			        		
			        		;
		        		
		        		
		        		
			        		Xiao ZHANG
			        		
			        		;
		        		
		        		
		        		
			        		Zhiyu QIAN
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. 南京航空航天大学自动化学院,南京市,210016
			        		
		        		
	        		
        		 
        	
        	
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		pulse signal;
			        		
			        		
			        		
				        		respiratory signal;
			        		
			        		
			        		
				        		synchronous acquisition;
			        		
			        		
			        		
				        		feature parameter extraction
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Medical Instrumentation
	            		
	            		 2024;48(2):167-172
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	A pulse and respiration synchronous detection system is designed to explore the changes of physiological signals in different situations.The system obtains the corresponding signal through STM32 control pulse and respiratory acquisition circuit,calculates and displays real-time parameters such as heart rate and respiratory rate,and transmits the data to the upper computer for storage in the database.The experimental test results show that the system can monitor pulse and respiratory waveform in different situations,and the waveform is in good condition.Compared with medical pulse oximeter,the error of measured heart rate and blood oxygen concentration is less than 3%,and the error of respiratory rate is less than 5%compared with the actual value,which verifies the accuracy of system signal acquisition.The system is small in size,low in cost,and comfortable to wear,and can be applied in experimental research related to pulse and respiratory signals.