Design of Non-Invasive Blood Oxygen Measurement Based on AFE4490
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3969/j.issn.1671-7104.2015.05.009
   		
        
        	
        		- VernacularTitle:基于血氧模拟前端AFE4490的无创血氧测量模块设计
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Jinsong ZHU
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Shouhao WU
			        		
			        		;
		        		
		        		
		        		
			        		Wenxiu GUO
			        		
			        		;
		        		
		        		
		        		
			        		Hui ZHENG
			        		
			        		;
		        		
		        		
		        		
			        		Dong TANG
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. 深圳清华大学研究院数字电视系统重点实验室
			        		
		        		
	        		
        		 
        	
        	
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		non-invasive oximetry;
			        		
			        		
			        		
				        		oxygen analog front-end AFE4490;
			        		
			        		
			        		
				        		Lambert-Beer;
			        		
			        		
			        		
				        		Photo Plethysmo Graphy(PPG) signal
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Medical Instrumentation
	            		
	            		 2015;(5):341-343
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	From the perspective of portable monitoring devices,we use an analog front-end AFE4490 design a module of Non-invasive blood oxygen measurement,used to colect human pulse wave signal and peak (valey) value detection and then use the principles of non-invasive oximetry calculated oxygen saturation (SPO2). This design of noninvasive oximetry module has the characteristics of smal size, low power consumption, and the results of test show that the measurement of oxygen saturation are correct.