Comparison of anterior chamber depth measured by anterior segment optical coherence to-mography and ultrasound biomicroscopy:a meta-analysis
	    		
		   		
	    	
    	
    	
   		
        
        	
        		- VernacularTitle:光学相干断层扫描及超声生物显微镜测量眼前房深度的meta分析
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Linjiang CHEN
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Ke XIONG
			        		
			        		;
		        		
		        		
		        		
			        		Jing WU
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. 南方医科大学南方医院 眼科
			        		
		        		
	        		
        		 
        	
        	
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		optical coherence tomography;
			        		
			        		
			        		
				        		ultrasound biomicroscopy;
			        		
			        		
			        		
				        		anterior chamber;
			        		
			        		
			        		
				        		meta-analysis
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Journal of Southern Medical University
	            		
	            		 2013;(10):1533-1537
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To compare the differences in the anterior chamber depth (ACD) measured by anterior segment optical coherence tomography (AS-OCT) and ultrasound biomicroscopy (UBM). Methods All studies pertaining to ACD measured by AS-OCT and UBM were collected from online databases. The assessment of methodological quality and data extraction from the included studies were performed independently by two reviewers for meta-analysis. Results Eight studies involving 710 eyes were included in the analysis. The difference of ACD measurements between AS-OCT and UBM was not statistically significant in the overall patients included for analysis (SMD=0.19, 95%CI[0.00, 0.39]) or in the patients with primary angle-closed glaucoma (SMD=0.02,95%CI[-0.04,0.19]). Conclusion The ACD measurements do not differ significantly between AS-OCT and UBM. Due to the relatively small number of the included studies and the patients involved, this conclusion needs further confirmation by high-quality studies involving larger sample sizes.