Evaluation of measurement uncertainty of laboratory endocrine test items
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3969/j.issn.1673-9701.2024.03.010
   		
        
        	
        		- VernacularTitle:实验室内分泌项目测量不确定度的评定研究
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Luyun ZENG
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Xiaojun YANG
			        		
			        		
			        		
			        			2
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Xiaowen JIN
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. Department of Clinical Laboratory, Beijing Xicheng District Maternal and Child Health Hospital, Beijing 100054, China
			        		
			        			2. Department of Epidemiology and Health Statistics, School of Public Health, Capital Medical University, Beijing Municipal  Key Laboratory of Clinical Epidemiology, Beijing 100069, China
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Evaluation;
			        		
			        		
			        		
				        		Measurement uncertainty;
			        		
			        		
			        		
				        		Top-down;
			        		
			        		
			        		
				        		Endocrine project
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			China Modern Doctor
	            		
	            		 2024;62(3):42-46
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To evaluate the measurement uncertainty of endocrine test items by"top-down"method.Methods The"top-down"method in the technical report of CNAS-CL01-G003:2021"Medical Laboratory-Evaluation and Expression of Measurement Uncertainty"was used,collecting the laboratory's internal quality control data and the external quality assurance results of Beijing Clinical Laboratory Center from 2019 to 2021.To evaluate the measurement uncertainty of 11 endocrine tests including free triiodothyronine(FT3),total triiodothyronine(TT3)and free thyroxine(FT4).Results The relative expanded uncertainties of 11 endocrine test items in laboratory were FT3:12.658%,TT3:13.372%,FT4:10.468%,total thyroxine(TT4):32.382%,thyroid stimulating hormone(TSH):8.594%,estradiol(E2):18.656%,follicle-stimulating hormone(FSH):14.650%,luteinizing hormone(LH):29.384%,progesterone(PRO):28.806%,prolactin(PRL):13.810%and testosterone(TESTO):31.610%.At the normal concentration level,the relative expanded uncertainties were FT3:12.424%,TT3:12.462%,FT4:8.606%,TT4:14.130%,TSH:12.536%,E2:19.586%,FSH:14.382%,LH:23.400%,PRO:38.346%,PRL:16.014%and TESTO:39.352%.The relative expanded uncertainties at high concentration levels are FT3:13.882%,TT3:14.096%,FT4:11.040%,TT4:8.614%,TSH:7.782%,E2:14.366%,FSH:13.436%,LH:14.804%,PRO:11.574%,PRL:17.742%,TESTO:39.322%.The relative extended measurement uncertainty of TT4,LH near low concentration levels,PRO near low and normal concentration levels,and TESTO near three concentration levels was higher than the target uncertainty(25%),and the remaining items are less than the target uncertainty.Conclusion The quality control data in the laboratory can reflect the measurement uncertainty of 11 endocrine detection items including FT3,TT3 and FT4 in Abbott i2000SR automatic chemiluminescence immunoassay analyzer,which has important clinical application value.