The impact of programming optimization for atrioventricular synchrony after Micra AV leadless pacemakers implantation
10.3969/j.issn.1004-8812.2024.02.002
- VernacularTitle:Micra AV无导线起搏器植入术后随访优化对房室同步率的影响
- Author:
Ze ZHENG
1
;
Yu-Chen SHI
;
Song-Yuan HE
;
Shao-Ping WANG
;
Shi-Ying LI
;
Shu-Juan CHENG
;
Jing-Hua LIU
Author Information
1. 首都医科大学附属北京安贞医院心内科 北京市心肺血管疾病研究所,北京 100029
- Keywords:
Leadless pacemaker;
Programming;
Optimization;
Atrioventricular synchronization rate
- From:
Chinese Journal of Interventional Cardiology
2024;32(2):71-75
- CountryChina
- Language:Chinese
-
Abstract:
Objective To analyze the atrioventricular synchronization rate after implantation of Micra AV leadless pacemaker,and the impact of postoperative programming optimization on atrioventricular synchronization rate.Methods A prospective cohort study was conducted to select patients with complete atrioventricular block who underwent Micra AV leadless pacemaker implantation at Beijing Anzhen Hospital from August 2022 to June 2023.Programming optimization were performed at 1 week,1 month,and 3 months postoperatively,and atrioventricular synchronization rate,electrical parameters,and echocardiography were recorded.Results A total of 68 patients with complete atrioventricular block implanted with Micra AV were selected,with an average age of(68.2±9.7)years,including 47 males(69.1%).All patients were successfully implanted with Micra AV,and there were no serious postoperative complications;The average threshold,sense,and impedance parameters were stable during 1 week,1 month,and 3 months after the procedure;There was no significant difference in the EF value of postoperative echocardiography(P=0.162);The average atrioventricular synchronization rates at 1 week,1 month,and 3 months postoperatively were(75.2%vs.83.8%vs.91.6%,P=0.001).Conclusions As an mechanical atrial sensing,Micra AV requires personalized adjustment of relevant parameters;Postoperative follow-up programming optimization plays an important role in the atrioventricular synchronization and comfort level in patients with complete atrioventricular block after implantation of Micra AV leadless pacemaker.