Repair and maintenance multi-leaf collimator optical system of Elekta medical linear accelerator
	    		
		   		
		   			
		   		
	    	
    	 
    	10.19745/j.1003-8868.2024013
   		
        
        	
        		- VernacularTitle:医科达医用直线加速器MLC光学系统维修与维护研究
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Qing-She DING
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Li-Qing ZHONG
			        		
			        		;
		        		
		        		
		        		
			        		Liang-Gang SUN
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. 联勤保障部队第902医院医学工程科,安徽蚌埠 233015
			        		
		        		
	        		
        		 
        	
        	
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Elekta medical linear accelerator;
			        		
			        		
			        		
				        		multi-leaf collimator;
			        		
			        		
			        		
				        		optical system;
			        		
			        		
			        		
				        		fault repair;
			        		
			        		
			        		
				        		maintenance
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Medical Equipment Journal
	            		
	            		 2024;45(1):76-83
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	The structure and working principle of the multi-leaf collimator(MLC)optical system of Elekta linear accelerator were introduced.The causes for the faults of the MLC optical system were analyzed,and the fault repair and maintenance of the MLC optical system were summarized.References were provided for medical engineers to repair and maintain the MLC optical system.[Chinese Medical Equipment Journal,2024,45(1):76-83]