1.Powered Tool for the Removal of a Well-Fixed Acetabular Cup: A Comparative Experimental Study
Jun Young CHUNG ; Yonghan CHA ; Chang-Ho JUNG ; Jin-Woo KIM ; Jun-Il YOO ; Jung-Taek KIM ; Yongho JEON
Clinics in Orthopedic Surgery 2025;17(3):372-380
Background:
The removal of a well-fixed acetabular cup is a challenging, labor-intensive, and time-consuming step during revision hip arthroplasty. Although the advent of the manual osteotome, Explant, has simplified the procedure, it is still a stressful process as it dissipates the surgeon’s strength and time and risks an iatrogenic pelvic fracture. Recently, EZX, a powered tool for extraction of well-fixed acetabular cups with semicircular blade was invented. This study aimed to compare Explant and EZX in an experimental condition for their efficacy and safety.
Methods:
Cementless acetabular cups were press-fitted to 20 hemipelvic polyurethane models using foam adhesives. Ten cups were removed with each tool for comparison of the elapsed time, loads on the entire hemipelvis, periacetabular strain and temperature, volume of periacetabular bone removed, and diameter of the remaining acetabular rim. Strains and loads were quantitatively assessed using strain gauges and load cells for precise and reliable measurements.
Results:
The mean duration required to remove a well-fixed cup with EZX was 38.5 seconds (range, 25–55), whereas that with Explant was 543.7 seconds (range, 214–1,051) (p < 0.001). The load on the entire hemipelvis with EZX (mean, 9.1 kgf; range, 6.4–11.3) was 33% lower than that with Explant (mean, 13.6 kgf; range, 9.2–17.1) (p < 0.001). The periacetabular peak strains at the 3 positions with EZX were significantly lower than those with Explant (p < 0.001). The temperature during the removal did not differ significantly between the 2 tools. Although the mean volume of bone loss with Explant was 2.4 mL more than that with EZX (p < 0.001), the mean diameters of the remaining acetabular rim were not significantly different, measuring 54.1 mm with both tools.
Conclusions
The present experiment revealed that a well-fixed cup could be removed using a powered tool with less strength and time and less load on the entire pelvis. Although the powered tool removed a larger volume of bone, the diameters of the remaining acetabular rims were equivalent. This tool may help surgeons remove well-fixed cups in a short time and reduce the deforming load on the bone around the cup without increasing the size of the subsequent reconstruction cup.
2.Early Administration of Nelonemdaz May Improve the Stroke Outcomes in Patients With Acute Stroke
Jin Soo LEE ; Ji Sung LEE ; Seong Hwan AHN ; Hyun Goo KANG ; Tae-Jin SONG ; Dong-Ick SHIN ; Hee-Joon BAE ; Chang Hun KIM ; Sung Hyuk HEO ; Jae-Kwan CHA ; Yeong Bae LEE ; Eung Gyu KIM ; Man Seok PARK ; Hee-Kwon PARK ; Jinkwon KIM ; Sungwook YU ; Heejung MO ; Sung Il SOHN ; Jee Hyun KWON ; Jae Guk KIM ; Young Seo KIM ; Jay Chol CHOI ; Yang-Ha HWANG ; Keun Hwa JUNG ; Soo-Kyoung KIM ; Woo Keun SEO ; Jung Hwa SEO ; Joonsang YOO ; Jun Young CHANG ; Mooseok PARK ; Kyu Sun YUM ; Chun San AN ; Byoung Joo GWAG ; Dennis W. CHOI ; Ji Man HONG ; Sun U. KWON ;
Journal of Stroke 2025;27(2):279-283
3.Powered Tool for the Removal of a Well-Fixed Acetabular Cup: A Comparative Experimental Study
Jun Young CHUNG ; Yonghan CHA ; Chang-Ho JUNG ; Jin-Woo KIM ; Jun-Il YOO ; Jung-Taek KIM ; Yongho JEON
Clinics in Orthopedic Surgery 2025;17(3):372-380
Background:
The removal of a well-fixed acetabular cup is a challenging, labor-intensive, and time-consuming step during revision hip arthroplasty. Although the advent of the manual osteotome, Explant, has simplified the procedure, it is still a stressful process as it dissipates the surgeon’s strength and time and risks an iatrogenic pelvic fracture. Recently, EZX, a powered tool for extraction of well-fixed acetabular cups with semicircular blade was invented. This study aimed to compare Explant and EZX in an experimental condition for their efficacy and safety.
Methods:
Cementless acetabular cups were press-fitted to 20 hemipelvic polyurethane models using foam adhesives. Ten cups were removed with each tool for comparison of the elapsed time, loads on the entire hemipelvis, periacetabular strain and temperature, volume of periacetabular bone removed, and diameter of the remaining acetabular rim. Strains and loads were quantitatively assessed using strain gauges and load cells for precise and reliable measurements.
Results:
The mean duration required to remove a well-fixed cup with EZX was 38.5 seconds (range, 25–55), whereas that with Explant was 543.7 seconds (range, 214–1,051) (p < 0.001). The load on the entire hemipelvis with EZX (mean, 9.1 kgf; range, 6.4–11.3) was 33% lower than that with Explant (mean, 13.6 kgf; range, 9.2–17.1) (p < 0.001). The periacetabular peak strains at the 3 positions with EZX were significantly lower than those with Explant (p < 0.001). The temperature during the removal did not differ significantly between the 2 tools. Although the mean volume of bone loss with Explant was 2.4 mL more than that with EZX (p < 0.001), the mean diameters of the remaining acetabular rim were not significantly different, measuring 54.1 mm with both tools.
Conclusions
The present experiment revealed that a well-fixed cup could be removed using a powered tool with less strength and time and less load on the entire pelvis. Although the powered tool removed a larger volume of bone, the diameters of the remaining acetabular rims were equivalent. This tool may help surgeons remove well-fixed cups in a short time and reduce the deforming load on the bone around the cup without increasing the size of the subsequent reconstruction cup.
4.Independent and Combined Effects of Particulate Matter and Sleep Deprivation on Human Skin Barrier
Il Joo KWON ; Eun Jung LEE ; Jong Ho PARK ; Ji Young KIM ; Seohyun PARK ; Yu Jeong BAE ; Shinwon HWANG ; Hye-won NA ; Nari CHA ; Geunhyuk JANG ; Hyoung-June KIM ; Hae Kwang LEE ; Sang Ho OH
Annals of Dermatology 2025;37(3):131-139
Background:
The exposome encompasses all factors people encounter through life, with the skin constantly exposed. While particulate matter (PM) and sleep deprivation are known to contribute to barrier dysfunction, their combined effects remain unclear.
Objective:
To evaluate the independent and combined effects of PM exposure and short-term sleep deprivation on skin barrier function.
Methods:
Forty healthy Korean women (aged 24–58 years) were enrolled in this study. Forearms were divided into 4 sites: control, PM exposure, sleep deprivation, and PM plus sleep deprivation. Parameters such as trans-epidermal water loss (TEWL), hydration, elasticity, roughness, and redness were measured at baseline and post-exposure. RNA sequencing and reverse transcription-polymerase chain reaction were conducted on tape-stripped skin samples.
Results:
PM exposure significantly increased TEWL (+25.59%, p<0.01), roughness (+21.9%, p<0.01), and redness (+13.7%, p<0.0001) while reducing elasticity (−3.98%, p<0.01). Sleep deprivation modestly reduced elasticity (−1.39%, p<0.05) without affecting other parameters.Combined PM and sleep deprivation did not further exacerbate barrier dysfunction compared to PM alone. RNA sequencing revealed reduced FLG and LORICRIN expression and upregulated endoplasmic reticulum (ER) stress markers (HSP90B1, CANX) in both PM and sleep deprivation conditions.
Conclusion
PM exposure impaired skin barrier function, while short-term sleep deprivation alone did not significantly affect the barrier, either independently or in combination with PM.However, it was observed that the sleep deprivation-only, while not directly causing barrier damage, induced changes in ER stress-related gene expression in tape-stripped skin samples, like the PM exposure-only. This suggests that such signaling pathways could potentially exacerbate skin barrier deterioration.
5.Powered Tool for the Removal of a Well-Fixed Acetabular Cup: A Comparative Experimental Study
Jun Young CHUNG ; Yonghan CHA ; Chang-Ho JUNG ; Jin-Woo KIM ; Jun-Il YOO ; Jung-Taek KIM ; Yongho JEON
Clinics in Orthopedic Surgery 2025;17(3):372-380
Background:
The removal of a well-fixed acetabular cup is a challenging, labor-intensive, and time-consuming step during revision hip arthroplasty. Although the advent of the manual osteotome, Explant, has simplified the procedure, it is still a stressful process as it dissipates the surgeon’s strength and time and risks an iatrogenic pelvic fracture. Recently, EZX, a powered tool for extraction of well-fixed acetabular cups with semicircular blade was invented. This study aimed to compare Explant and EZX in an experimental condition for their efficacy and safety.
Methods:
Cementless acetabular cups were press-fitted to 20 hemipelvic polyurethane models using foam adhesives. Ten cups were removed with each tool for comparison of the elapsed time, loads on the entire hemipelvis, periacetabular strain and temperature, volume of periacetabular bone removed, and diameter of the remaining acetabular rim. Strains and loads were quantitatively assessed using strain gauges and load cells for precise and reliable measurements.
Results:
The mean duration required to remove a well-fixed cup with EZX was 38.5 seconds (range, 25–55), whereas that with Explant was 543.7 seconds (range, 214–1,051) (p < 0.001). The load on the entire hemipelvis with EZX (mean, 9.1 kgf; range, 6.4–11.3) was 33% lower than that with Explant (mean, 13.6 kgf; range, 9.2–17.1) (p < 0.001). The periacetabular peak strains at the 3 positions with EZX were significantly lower than those with Explant (p < 0.001). The temperature during the removal did not differ significantly between the 2 tools. Although the mean volume of bone loss with Explant was 2.4 mL more than that with EZX (p < 0.001), the mean diameters of the remaining acetabular rim were not significantly different, measuring 54.1 mm with both tools.
Conclusions
The present experiment revealed that a well-fixed cup could be removed using a powered tool with less strength and time and less load on the entire pelvis. Although the powered tool removed a larger volume of bone, the diameters of the remaining acetabular rims were equivalent. This tool may help surgeons remove well-fixed cups in a short time and reduce the deforming load on the bone around the cup without increasing the size of the subsequent reconstruction cup.
6.Early Administration of Nelonemdaz May Improve the Stroke Outcomes in Patients With Acute Stroke
Jin Soo LEE ; Ji Sung LEE ; Seong Hwan AHN ; Hyun Goo KANG ; Tae-Jin SONG ; Dong-Ick SHIN ; Hee-Joon BAE ; Chang Hun KIM ; Sung Hyuk HEO ; Jae-Kwan CHA ; Yeong Bae LEE ; Eung Gyu KIM ; Man Seok PARK ; Hee-Kwon PARK ; Jinkwon KIM ; Sungwook YU ; Heejung MO ; Sung Il SOHN ; Jee Hyun KWON ; Jae Guk KIM ; Young Seo KIM ; Jay Chol CHOI ; Yang-Ha HWANG ; Keun Hwa JUNG ; Soo-Kyoung KIM ; Woo Keun SEO ; Jung Hwa SEO ; Joonsang YOO ; Jun Young CHANG ; Mooseok PARK ; Kyu Sun YUM ; Chun San AN ; Byoung Joo GWAG ; Dennis W. CHOI ; Ji Man HONG ; Sun U. KWON ;
Journal of Stroke 2025;27(2):279-283
7.Powered Tool for the Removal of a Well-Fixed Acetabular Cup: A Comparative Experimental Study
Jun Young CHUNG ; Yonghan CHA ; Chang-Ho JUNG ; Jin-Woo KIM ; Jun-Il YOO ; Jung-Taek KIM ; Yongho JEON
Clinics in Orthopedic Surgery 2025;17(3):372-380
Background:
The removal of a well-fixed acetabular cup is a challenging, labor-intensive, and time-consuming step during revision hip arthroplasty. Although the advent of the manual osteotome, Explant, has simplified the procedure, it is still a stressful process as it dissipates the surgeon’s strength and time and risks an iatrogenic pelvic fracture. Recently, EZX, a powered tool for extraction of well-fixed acetabular cups with semicircular blade was invented. This study aimed to compare Explant and EZX in an experimental condition for their efficacy and safety.
Methods:
Cementless acetabular cups were press-fitted to 20 hemipelvic polyurethane models using foam adhesives. Ten cups were removed with each tool for comparison of the elapsed time, loads on the entire hemipelvis, periacetabular strain and temperature, volume of periacetabular bone removed, and diameter of the remaining acetabular rim. Strains and loads were quantitatively assessed using strain gauges and load cells for precise and reliable measurements.
Results:
The mean duration required to remove a well-fixed cup with EZX was 38.5 seconds (range, 25–55), whereas that with Explant was 543.7 seconds (range, 214–1,051) (p < 0.001). The load on the entire hemipelvis with EZX (mean, 9.1 kgf; range, 6.4–11.3) was 33% lower than that with Explant (mean, 13.6 kgf; range, 9.2–17.1) (p < 0.001). The periacetabular peak strains at the 3 positions with EZX were significantly lower than those with Explant (p < 0.001). The temperature during the removal did not differ significantly between the 2 tools. Although the mean volume of bone loss with Explant was 2.4 mL more than that with EZX (p < 0.001), the mean diameters of the remaining acetabular rim were not significantly different, measuring 54.1 mm with both tools.
Conclusions
The present experiment revealed that a well-fixed cup could be removed using a powered tool with less strength and time and less load on the entire pelvis. Although the powered tool removed a larger volume of bone, the diameters of the remaining acetabular rims were equivalent. This tool may help surgeons remove well-fixed cups in a short time and reduce the deforming load on the bone around the cup without increasing the size of the subsequent reconstruction cup.
8.Early Administration of Nelonemdaz May Improve the Stroke Outcomes in Patients With Acute Stroke
Jin Soo LEE ; Ji Sung LEE ; Seong Hwan AHN ; Hyun Goo KANG ; Tae-Jin SONG ; Dong-Ick SHIN ; Hee-Joon BAE ; Chang Hun KIM ; Sung Hyuk HEO ; Jae-Kwan CHA ; Yeong Bae LEE ; Eung Gyu KIM ; Man Seok PARK ; Hee-Kwon PARK ; Jinkwon KIM ; Sungwook YU ; Heejung MO ; Sung Il SOHN ; Jee Hyun KWON ; Jae Guk KIM ; Young Seo KIM ; Jay Chol CHOI ; Yang-Ha HWANG ; Keun Hwa JUNG ; Soo-Kyoung KIM ; Woo Keun SEO ; Jung Hwa SEO ; Joonsang YOO ; Jun Young CHANG ; Mooseok PARK ; Kyu Sun YUM ; Chun San AN ; Byoung Joo GWAG ; Dennis W. CHOI ; Ji Man HONG ; Sun U. KWON ;
Journal of Stroke 2025;27(2):279-283
9.2024 Consensus Statement on Coronary Stenosis and Plaque Evaluation in CT Angiography From the Asian Society of Cardiovascular Imaging-Practical Tutorial (ASCI-PT)
Cherry KIM ; Chul Hwan PARK ; Bae Young LEE ; Chan Ho PARK ; Eun-Ju KANG ; Hyun Jung KOO ; Kakuya KITAGAWA ; Min Jae CHA ; Rungroj KRITTAYAPHONG ; Sang Il CHOI ; Hwan Seok YONG ; Sung Min KO ; Sung Mok KIM ; Sung Ho HWANG ; Nguyen Ngoc TRANG ; Whal LEE ; Young Jin KIM ; Jongmin LEE ; Dong Hyun YANG
Cardiovascular Imaging Asia 2024;8(2):21-31
The Asian Society of Cardiovascular Imaging-Practical Tutorial (ASCI-PT) is an instructional initiative of the ASCI School designed to enhance educational standards. In 2021, the ASCI-PT was convened with the goal of formulating a consensus statement on the assessment of coronary stenosis and coronary plaque using coronary CT angiography (CCTA). Nineteen experts from four countries conducted thorough reviews of current guidelines and deliberated on eight key issues to refine the process and improve the clarity of reporting CCTA findings. The experts engaged in both online and on-site sessions to establish a unified agreement. This document presents a summary of the ASCI-PT 2021 deliberations and offers a comprehensive consensus statement on the evaluation of coronary stenosis and coronary plaque in CCTA.
10.Deep Learning-Based Automatic Classification of Ischemic Stroke Subtype Using Diffusion-Weighted Images
Wi-Sun RYU ; Dawid SCHELLINGERHOUT ; Hoyoun LEE ; Keon-Joo LEE ; Chi Kyung KIM ; Beom Joon KIM ; Jong-Won CHUNG ; Jae-Sung LIM ; Joon-Tae KIM ; Dae-Hyun KIM ; Jae-Kwan CHA ; Leonard SUNWOO ; Dongmin KIM ; Sang-Il SUH ; Oh Young BANG ; Hee-Joon BAE ; Dong-Eog KIM
Journal of Stroke 2024;26(2):300-311
Background:
and Purpose Accurate classification of ischemic stroke subtype is important for effective secondary prevention of stroke. We used diffusion-weighted image (DWI) and atrial fibrillation (AF) data to train a deep learning algorithm to classify stroke subtype.
Methods:
Model development was done in 2,988 patients with ischemic stroke from three centers by using U-net for infarct segmentation and EfficientNetV2 for subtype classification. Experienced neurologists (n=5) determined subtypes for external test datasets, while establishing a consensus for clinical trial datasets. Automatically segmented infarcts were fed into the model (DWI-only algorithm). Subsequently, another model was trained, with AF included as a categorical variable (DWI+AF algorithm). These models were tested: (1) internally against the opinion of the labeling experts, (2) against fresh external DWI data, and (3) against clinical trial dataset.
Results:
In the training-and-validation datasets, the mean (±standard deviation) age was 68.0±12.5 (61.1% male). In internal testing, compared with the experts, the DWI-only and the DWI+AF algorithms respectively achieved moderate (65.3%) and near-strong (79.1%) agreement. In external testing, both algorithms again showed good agreements (59.3%–60.7% and 73.7%–74.0%, respectively). In the clinical trial dataset, compared with the expert consensus, percentage agreements and Cohen’s kappa were respectively 58.1% and 0.34 for the DWI-only vs. 72.9% and 0.57 for the DWI+AF algorithms. The corresponding values between experts were comparable (76.0% and 0.61) to the DWI+AF algorithm.
Conclusion
Our model trained on a large dataset of DWI (both with or without AF information) was able to classify ischemic stroke subtypes comparable to a consensus of stroke experts.

Result Analysis
Print
Save
E-mail