1.Analysis of the Influence of Different Scanning Conditions of Medical Linear Accelerator CBCT on Image Quality.
Li LIU ; Chengwei YE ; Jianjun YUAN ; Yingui LUO ; Zhiyao LUO ; Wei ZENG ; Ling LI ; Huan LIU ; Yan LIU
Chinese Journal of Medical Instrumentation 2025;49(2):176-180
OBJECTIVE:
To investigate the influence of different scanning conditions on the image quality of medical electron accelerator cone-beam computed tomography (CBCT) and provide a reference for the selection of scanning conditions for different body parts. Methods Set different scanning conditions, the Catphan 503 phantom was scanned using CBCT parameters to analyze the influence of spatial resolution, noise, uniformity, spatial geometric accuracy, and low-contrast resolution on the image quality of CBCT.
RESULTS:
For the head, chest, and abdomen, with the increase in scanning parameter values, the noise value decreased by 47.4%, 26.1%, and 51.3% respectively, and the uniformity values decreased by 30.2%, 26.6%, and 47.9% respectively. The low-contrast resolution values decreased by 50.6%, 34.2%, and 12.0%. The influence of different scanning conditions on spatial geometric accuracy and spatial resolution is not significant.
CONCLUSION
Different scanning parameters have a certain influence on the image quality of medical electron accelerator CBCT. Lower scanning parameters can be selected based on individual patients to reduce the additional radiation dose, providing a reference for the safe application of CBCT image guidance in radiotherapy.
Cone-Beam Computed Tomography/instrumentation*
;
Phantoms, Imaging
;
Particle Accelerators
2.Comparative study of different image-guided systems in positioning error and placement outside the target area of radiotherapy for cervical cancer
Li LIU ; Chengwei YE ; Jianjun YUAN ; Yingui LUO ; Zhiyao LUO ; Wei ZENG ; Ling LI ; Huan LIU ; Yan LIU ; Miao HE
China Medical Equipment 2025;22(4):18-22
Objective:To compare and analyze the differences in positioning errors and placement outside the target area between two different image-guided systems(Halcyon and Synergy)in radiotherapy for cervical cancer.Methods:The data of 40 patients with cervical cancer who received treatment in Dezhou People's Hospital from March 2023 to June 2023 were retrospectively selected.Each patient was treated respectively with Halcyon linear accelerator and Synergy linear accelerator,with 20 cases for each treatment method.All patients underwent cone beam computed tomography(CBCT)at kV level to conduct image scanning before radiotherapy,and the obtained CBCT images were matched with the positioned CT images by online.The registration results of left and right(x),head and foot(y)and front and back(z)of three translation direction,and pitch(Rx),rotation(Ry)and deflection(Rz)directions of rotation direction were respectively obtained.The margin of planning target volume(MPTV)was calculated according to positioning expansion formula.The differences of positioning errors and target area expansion between Halcyon and Synergy image-guided systems were compared and analyzed.Results:Halcyon and Synergy image-guided systems obtained respectively 479 and 447 sets of CBCT images.The translational errors of Halcyon image-guided system on x,y,and z directions were respectively(-0.03±0.21),(0.12±0.63)and(0.02±0.15)cm,and the rotational errors of that on Rx,Ry and Rz directions were respectively(0.38±1.26)°,(-0.40±1.02)°,and(0.07±1.07)°.The translational errors of Synergy image-guided system were respectively(-0.07±0.18),(0.08±0.49)and(0.11±0.22)cm on the three directions,and the rotational errors of that were respectively(0.28±1.29)°,(-0.29±1.29)° and(-0.16±1.20)°.The translational error values on the y direction of both image-guided system were larger than that on x and z directions,and the absolute values of the range of translational errors of them were basically within 5 mm.There were no statistically significant difference in the positioning errors of rotation on Rx and Ry directions between Halcyon and Synergy image-guided systems(P>0.05),while there was a statistically significant difference in the positioning error of rotation on Rz direction between them(t=-3.06,P<0.05).The MPTV values of Halcyon image-guided system on x,y and z directions were respectively 0.4,1.4 and 0.3 cm,and those of Synergy image-guided system were respectively 0.4,0.9 and 0.5 cm.The expansion boundary of target area of Halcyon image-guided system on the y direction was larger than that of Synergy image-guided system,and that on other directions were all less or equal to 0.5 cm.Conclusion:The MPTV values of Halcyon and Synergy image-guided systems can meet clinical requirements in radiotherapy.The positioning errors and the tube current of the scan of Halcyon image-guided system were larger than those of Synergy image-guided system,but the scanning time and treatment time of that were significantly lower than those of Synergy image-guided system.Different treatment systems can be selected individually for patients according to their different conditions.
3.Comparative study of different image-guided systems in positioning error and placement outside the target area of radiotherapy for cervical cancer
Li LIU ; Chengwei YE ; Jianjun YUAN ; Yingui LUO ; Zhiyao LUO ; Wei ZENG ; Ling LI ; Huan LIU ; Yan LIU ; Miao HE
China Medical Equipment 2025;22(4):18-22
Objective:To compare and analyze the differences in positioning errors and placement outside the target area between two different image-guided systems(Halcyon and Synergy)in radiotherapy for cervical cancer.Methods:The data of 40 patients with cervical cancer who received treatment in Dezhou People's Hospital from March 2023 to June 2023 were retrospectively selected.Each patient was treated respectively with Halcyon linear accelerator and Synergy linear accelerator,with 20 cases for each treatment method.All patients underwent cone beam computed tomography(CBCT)at kV level to conduct image scanning before radiotherapy,and the obtained CBCT images were matched with the positioned CT images by online.The registration results of left and right(x),head and foot(y)and front and back(z)of three translation direction,and pitch(Rx),rotation(Ry)and deflection(Rz)directions of rotation direction were respectively obtained.The margin of planning target volume(MPTV)was calculated according to positioning expansion formula.The differences of positioning errors and target area expansion between Halcyon and Synergy image-guided systems were compared and analyzed.Results:Halcyon and Synergy image-guided systems obtained respectively 479 and 447 sets of CBCT images.The translational errors of Halcyon image-guided system on x,y,and z directions were respectively(-0.03±0.21),(0.12±0.63)and(0.02±0.15)cm,and the rotational errors of that on Rx,Ry and Rz directions were respectively(0.38±1.26)°,(-0.40±1.02)°,and(0.07±1.07)°.The translational errors of Synergy image-guided system were respectively(-0.07±0.18),(0.08±0.49)and(0.11±0.22)cm on the three directions,and the rotational errors of that were respectively(0.28±1.29)°,(-0.29±1.29)° and(-0.16±1.20)°.The translational error values on the y direction of both image-guided system were larger than that on x and z directions,and the absolute values of the range of translational errors of them were basically within 5 mm.There were no statistically significant difference in the positioning errors of rotation on Rx and Ry directions between Halcyon and Synergy image-guided systems(P>0.05),while there was a statistically significant difference in the positioning error of rotation on Rz direction between them(t=-3.06,P<0.05).The MPTV values of Halcyon image-guided system on x,y and z directions were respectively 0.4,1.4 and 0.3 cm,and those of Synergy image-guided system were respectively 0.4,0.9 and 0.5 cm.The expansion boundary of target area of Halcyon image-guided system on the y direction was larger than that of Synergy image-guided system,and that on other directions were all less or equal to 0.5 cm.Conclusion:The MPTV values of Halcyon and Synergy image-guided systems can meet clinical requirements in radiotherapy.The positioning errors and the tube current of the scan of Halcyon image-guided system were larger than those of Synergy image-guided system,but the scanning time and treatment time of that were significantly lower than those of Synergy image-guided system.Different treatment systems can be selected individually for patients according to their different conditions.
4.Clinical study of the effect of different CBCT scanning parameters on image quality and positioning error of cervical cancer
Li LIU ; Chengwei YE ; Jianjun YUAN ; Yingui LUO ; Zhiyao LUO ; Wei ZENG ; Ling LI ; Jiang LIU
China Medical Equipment 2024;21(11):18-24
Objective:To assess the effects of the scanning parameters of cone beam computed tomography(CBCT)on image quality,positioning error and additional radiation dose for patients with cervical cancer after tube voltage was reduced,so as to explore scanning conditions that was suitable to patients with cervical cancer.Methods:Seventy-two patients with cervical cancer who received radiotherapy in Deyang People's Hospital from January 2020 to August 2022 were selected.According to different scanning parameters,they were divided into low-dose group A(120 kV,154.4 mAs),low-dose group B(120 kV,228.8 mAs)and conventional group[adopted business-supplied scan sequences(120 kV,675.8 mAs)].The CBCT scan image was obtained by decreasing the tube voltage and tube current,and reducing scanning angle.The registration results of positioning error of low-dose group A,B and conventional group on three directions included left and right(x-axis),head and foot(y-axis)and abdominal and back(z-axis)were respectively obtained,and the results were analyzed by variance analysis.The Catphan503 phantom of X-ray volumetric imaging(XVI)was used to measure the image quality of the corresponding 3 groups,and the further comparison was conducted.Results:There were no significant differences between the three groups in the scanning parameters on x,y and z axis directions for positioning error of patients with cervical cancer(P>0.05).The scanning times of low-dose group A and B were approximately 1 minute,and the scanning time of conventional group was approximately 2 minutes.The noises of low-dose group A and B,and conventional group were respectively 36.51%,26.09%and 18.37%,and the radiation dose that was generated by CBCT image scanning in group A was correspondingly reduced.Conclusion:The reductions of scanning speed and collimator size of CBCT scanning parameters do not affect the positioning error and image quality,and it can decrease the scanning time and additional radiation dose of patients in undergoing radiotherapy.