1.Research on coagulation effect of cold atmospheric plasma jet device and its mechanism of action
Yan LI ; Hong-ye ZHENG ; Ao-xi XU ; Ya-jun ZHAO ; Shan-shan JIN ; Xu ZHANG ; Yu-fan WEI ; Yi-heng ZHANG ; Li ZHU ; Xi-ru LI
Chinese Medical Equipment Journal 2025;46(6):20-27
Objective To investigate the coagulation effect of a cold atmospheric plasma(CAP)jet device with helium as the working gas and to study its coagulation mechanism preliminarily.Methods A CAP jet device treatment group,a helium airflow treatment group,a hot air treatment group(60℃)and a natural coagulation group were formed according to the treatment modes of the blood samples,with 10 μL of blood samples involved in each group,in order to validate the coagulation effect of the CAP jet device in vitro;the coagulation mechanism of the CAP jet device was explored by its application to the treatment of anticoagulated whole blood,platelet-rich plasma and platelet-depleted plasma;the coagulation effect of the CAP jet device in vivo was verified with a mouse liver punctate hemorrhage model and a rabbit mesenteric hemorrhage model.Results The CAP jet device can significantly accelerate the coagulation of anticoagulated blood droplets,and the coagulation time of anticoagulated blood droplets in the CAP jet device-treated group was shortened from 28 min in the natural coagulation group to(23±1.56)s,with the difference statistically significant(P<0.05),and the CAP jet device treatment group gained advantages significantly over the helium airflow treatment group(P<0.05)and the hot air(60℃)treatment group(P<0.05)in coagulation-promoting effect;the procoagulant effect of the CAP jet device rose with the increase of platelet content in blood droplets,and the coagulation effect of platelet-rich blood droplets was significantly better than that of whole blood(P<0.05),while no coagulation was observed in platelet-poor droplets.The CAP jet device could rapidly stop hemostasis of punctate hemorrhage in mouse liver and mesenteric hemorrhage in rabbits without delayed hemorrhage occurring within 10 min,and no obvious structural abnormality of the liver and thermal damage of the tissue were found microscopically.Conclusion The CAP jet device plays procoagulant and hemostatic effects in vivo and in vitro,and its effect is not dependent on temperature and airflow evaporation effects and is considered to be related to platelet activation,with low thermal damage to living tissue.[Chinese Medical Equipment Journal,2025,46(6):20-27]
2.Research on coagulation effect of cold atmospheric plasma jet device and its mechanism of action
Yan LI ; Hong-ye ZHENG ; Ao-xi XU ; Ya-jun ZHAO ; Shan-shan JIN ; Xu ZHANG ; Yu-fan WEI ; Yi-heng ZHANG ; Li ZHU ; Xi-ru LI
Chinese Medical Equipment Journal 2025;46(6):20-27
Objective To investigate the coagulation effect of a cold atmospheric plasma(CAP)jet device with helium as the working gas and to study its coagulation mechanism preliminarily.Methods A CAP jet device treatment group,a helium airflow treatment group,a hot air treatment group(60℃)and a natural coagulation group were formed according to the treatment modes of the blood samples,with 10 μL of blood samples involved in each group,in order to validate the coagulation effect of the CAP jet device in vitro;the coagulation mechanism of the CAP jet device was explored by its application to the treatment of anticoagulated whole blood,platelet-rich plasma and platelet-depleted plasma;the coagulation effect of the CAP jet device in vivo was verified with a mouse liver punctate hemorrhage model and a rabbit mesenteric hemorrhage model.Results The CAP jet device can significantly accelerate the coagulation of anticoagulated blood droplets,and the coagulation time of anticoagulated blood droplets in the CAP jet device-treated group was shortened from 28 min in the natural coagulation group to(23±1.56)s,with the difference statistically significant(P<0.05),and the CAP jet device treatment group gained advantages significantly over the helium airflow treatment group(P<0.05)and the hot air(60℃)treatment group(P<0.05)in coagulation-promoting effect;the procoagulant effect of the CAP jet device rose with the increase of platelet content in blood droplets,and the coagulation effect of platelet-rich blood droplets was significantly better than that of whole blood(P<0.05),while no coagulation was observed in platelet-poor droplets.The CAP jet device could rapidly stop hemostasis of punctate hemorrhage in mouse liver and mesenteric hemorrhage in rabbits without delayed hemorrhage occurring within 10 min,and no obvious structural abnormality of the liver and thermal damage of the tissue were found microscopically.Conclusion The CAP jet device plays procoagulant and hemostatic effects in vivo and in vitro,and its effect is not dependent on temperature and airflow evaporation effects and is considered to be related to platelet activation,with low thermal damage to living tissue.[Chinese Medical Equipment Journal,2025,46(6):20-27]
3.Study of debridement effects of multi shapes of plasma scalpels in explosion injury model
Hong-ye ZHENG ; Yu LI ; Zi-heng XU ; Yu-fan WEI ; Bo-ya ZHANG ; Yan LI ; Li ZHU ; Xi-ru LI
Chinese Medical Equipment Journal 2025;46(2):31-38
Objective To explore the debridement effects of 3 types of plasma scalpels for the animal model of explosion injury,and to compare them with the steel scalpel and high-frequency electrosurgical scalpel.Methods Firstly,blast wounds were constructed in the right inguinal regions of 9 Landrace pigs by high-pressure gas impact combined with preset metal shrapnel.Secondly,debridement was carried out in experimental groups with wide-,arrow-or needle-type plasma scalpel and in control groups with steel and high-frequency electrisurgical scalpel,with the operating temperature and debridement time recorded during the procedure and trauma specimens analyzed pathologically after the debridement;comparisons were performed among the five types of scalpels in terms of debridement effect,and among the four ones in terms of maximum operating temperature and depth of tissue thermal damage under electrocutaneous cutting and electrocoagulation modes with the steel scalpel excluded because it did not generate any heat.GraphPad Prism 9.5.1 software was used for statistical analysis.Results There were no significant differences in debridement effect found between the three plasma scalpels and the steel and high-frequency electrosurgical scalpels(P>0.05).The three types of plasma scalpels had the maximum operating temperature lower significantly than that of the high-frequency electrosurgical scalpel during debridement(P<0.05).Under electrosection and electrocoagulation modes the three plasma scalpels had the depths of tissue thermal damage statistically less than that by the high-frequency electrosurgical scalpel under electrosection and electrocoagulation modes(P<0.05).The depths of tissue thermal damage by the four scalpels under electrocoagulation mode were obviously greater than those under electrosection mode(P<0.05).Conclusion Multi shapes of plasma scalpels behave well in debridement with low operating temperature,little tissue thermal damage and high efficiency for wound protection and the same efficacy with the steel scalpel and high-frequency electrosur-gical scalpel.[Chinese Medical Equipment Journal,2025,46(2):31-38]
4.Expert Consensus on the Ethical Requirements for Generative AI-Assisted Academic Writing
You-Quan BU ; Yong-Fu CAO ; Zeng-Yi CHANG ; Hong-Yu CHEN ; Xiao-Wei CHEN ; Yuan-Yuan CHEN ; Zhu-Cheng CHEN ; Rui DENG ; Jie DING ; Zhong-Kai FAN ; Guo-Quan GAO ; Xu GAO ; Lan HU ; Xiao-Qing HU ; Hong-Ti JIA ; Ying KONG ; En-Min LI ; Ling LI ; Yu-Hua LI ; Jun-Rong LIU ; Zhi-Qiang LIU ; Ya-Ping LUO ; Xue-Mei LV ; Yan-Xi PEI ; Xiao-Zhong PENG ; Qi-Qun TANG ; You WAN ; Yong WANG ; Ming-Xu WANG ; Xian WANG ; Guang-Kuan XIE ; Jun XIE ; Xiao-Hua YAN ; Mei YIN ; Zhong-Shan YU ; Chun-Yan ZHOU ; Rui-Fang ZHU
Chinese Journal of Biochemistry and Molecular Biology 2025;41(6):826-832
With the rapid development of generative artificial intelligence(GAI)technologies,their widespread application in academic research and writing is continuously expanding the boundaries of sci-entific inquiry.However,this trend has also raised a series of ethical and regulatory challenges,inclu-ding issues related to authorship,content authenticity,citation accuracy,and accountability.In light of the growing involvement of AI in generating academic content,establishing an open,controllable,and trustworthy ethical governance framework has become a key task for safeguarding research integrity and maintaining trust within the academic community.This expert consensus outlines ethical requirements across key stages of AI-assisted academic writing-including topic selection,data management,citation practices,and authorship attribution.It aims to clarify the boundaries and ethical obligations surrounding AI use in academic writing,ensuring that technological tools enhance efficiency without compromising in-tegrity.The goal is to provide guidance and institutional support for building a responsible and sustainable research ecosystem.
5.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
6.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
7.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
8.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
9.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
10.Expert Consensus on the Ethical Requirements for Generative AI-Assisted Academic Writing
You-Quan BU ; Yong-Fu CAO ; Zeng-Yi CHANG ; Hong-Yu CHEN ; Xiao-Wei CHEN ; Yuan-Yuan CHEN ; Zhu-Cheng CHEN ; Rui DENG ; Jie DING ; Zhong-Kai FAN ; Guo-Quan GAO ; Xu GAO ; Lan HU ; Xiao-Qing HU ; Hong-Ti JIA ; Ying KONG ; En-Min LI ; Ling LI ; Yu-Hua LI ; Jun-Rong LIU ; Zhi-Qiang LIU ; Ya-Ping LUO ; Xue-Mei LV ; Yan-Xi PEI ; Xiao-Zhong PENG ; Qi-Qun TANG ; You WAN ; Yong WANG ; Ming-Xu WANG ; Xian WANG ; Guang-Kuan XIE ; Jun XIE ; Xiao-Hua YAN ; Mei YIN ; Zhong-Shan YU ; Chun-Yan ZHOU ; Rui-Fang ZHU
Chinese Journal of Biochemistry and Molecular Biology 2025;41(6):826-832
With the rapid development of generative artificial intelligence(GAI)technologies,their widespread application in academic research and writing is continuously expanding the boundaries of sci-entific inquiry.However,this trend has also raised a series of ethical and regulatory challenges,inclu-ding issues related to authorship,content authenticity,citation accuracy,and accountability.In light of the growing involvement of AI in generating academic content,establishing an open,controllable,and trustworthy ethical governance framework has become a key task for safeguarding research integrity and maintaining trust within the academic community.This expert consensus outlines ethical requirements across key stages of AI-assisted academic writing-including topic selection,data management,citation practices,and authorship attribution.It aims to clarify the boundaries and ethical obligations surrounding AI use in academic writing,ensuring that technological tools enhance efficiency without compromising in-tegrity.The goal is to provide guidance and institutional support for building a responsible and sustainable research ecosystem.