1.Structural design and mechanical analysis of a "drum-shaped" balloon-expandable valve stent in expanded configuration.
Youzhi ZHAO ; Qianwen HOU ; Jianye ZHOU ; Shiliang CHEN ; Hanbing ZHANG ; Aike QIAO
Journal of Biomedical Engineering 2025;42(5):945-953
Stent migration is one of the common complications following transcatheter valve implantation. This study aims to design a "drum-shaped" balloon-expandable aortic valve stent to address this issue and conduct a mechanical analysis. The implantation process of the stent was evaluated using a method that combines numerical simulation and in vitro experiments. Furthermore, the fatigue process of the stent under pulsatile cyclic loading was simulated, and its fatigue performance was assessed using a Goodman diagram. The process of the stent migrating toward the left ventricular side was simulated, and the force-displacement curve of the stent was extracted to evaluate its anti- migration performance. The results showed that all five stent models could be crimped into a 14F sheath and enabled uniform expansion of the native valve leaflets. The stress in each stent was below the ultimate stress, so no fatigue fracture occurred. As the cell height ratio decreased, the contact area fraction between the stent and the aortic root gradually decreased. However, the mean contact force and the maximum anti-migration force first decreased and then increased. Specifically, model S5 had the smallest contact area fraction but the largest mean contact force and maximum anti-migration force, reaching approximately 0.16 MPa and 10.73 N, respectively. The designed stent achieves a "drum-shaped" change after expansion and has good anti-migration performance.
Stents
;
Prosthesis Design
;
Heart Valve Prosthesis
;
Humans
;
Aortic Valve/surgery*
;
Stress, Mechanical
;
Transcatheter Aortic Valve Replacement/instrumentation*
2.Impact of elevated arterial blood pressure on bioprosthetic valve calcification and failure after transcatheter aortic valve replacement.
Wenjing SHENG ; Qifeng ZHU ; Hanyi DAI ; Dao ZHOU ; Xianbao LIU
Journal of Zhejiang University. Medical sciences 2025;54(2):154-160
Transcatheter aortic valve replacement (TAVR) has emerged as the standard treatment for severe aortic stenosis, demonstrating comparable efficacy to traditional surgery in low and intermediate-risk patients. However, the bioprosthetic valves utilized in TAVR have a limited lifespan, and bioprosthetic valve failure, including calcification, rupture or infection may develop, leading to poor clinical outcomes. Elevated blood pressure has been identified as a key factor in aortic valve calcification, and its role in bioprosthetic valve failure is gaining increasing attention. Hypertension may accelerate the calcification process and exacerbate valve failure due to increased mechanical stress on the valve, activation of the renin-angiotensin system, and enhanced thrombus formation. Furthermore, elevated blood pressure interacts with prosthesis mismatch and paravalvular leak, jointly affecting valve durability. This review explores the impact of elevated blood pressure on bioprosthetic valve calcification and failure after TAVR, and emphasizes the importance of blood pressure control, optimized preoperative assessment, and appropriate valve selection in reducing valve failures.
Humans
;
Transcatheter Aortic Valve Replacement/adverse effects*
;
Calcinosis/etiology*
;
Bioprosthesis
;
Heart Valve Prosthesis/adverse effects*
;
Prosthesis Failure
;
Aortic Valve Stenosis/surgery*
;
Aortic Valve/surgery*
;
Hypertension/physiopathology*
3.Single-center experience in the treatment of severe aortic stenosis with XcorTM transcatheter aortic valve replacement system: 1-year follow-up results.
Shengwen WANG ; Haozhong LIU ; Haijiang GUO ; Tong TAN ; Hanxiang XIE ; Xiang LIU ; Hailong QIU ; Jimei CHEN ; Huiming GUO ; Jian LIU
Journal of Zhejiang University. Medical sciences 2025;54(2):141-148
OBJECTIVES:
To analyze the early clinical outcomes of the XcorTM transcatheter aortic valve replacement (TAVR) system in treating severe aortic stenosis. This study has been registered at Chinese Clinical Trial Registry (ChiCTR2200065593).
METHODS:
This single-arm, prospective clinical trial enrolled patients with severe aortic stenosis treated with the XcorTM TAVR system at the Section of Heart Valve & Coronary Artery Surgery, Guangdong Provincial People's Hospital. Perioperative and follow-up parameters were compared to evaluate differences in hemodynamic outcomes. All-cause mortality, aortic regurgitation, paravalvular leakage, cerebrovascular events, and reoperation were analyzed.
RESULTS:
Thirty-two patients with severe aortic stenosis were included (20 males, 12 females), with (70.9±4.3) years old and a Society of Thoracic Surgeons (STS) score of 6.45% (6.07%, 7.28%). Notably, 87.5% of patients had New York Heart Association (NYHA) class≥Ⅲ. All patients underwent successful XcorTM bioprosthesis implantation, achieving an immediate technical success rate of 100.0% and device success rate of 96.9%. Mean aortic valve gradient decreased from (55.21±23.17) mmHg (1 mmHg=0.133 kPa) to (8.45±5.30) mmHg, peak aortic jet velocity decreased from (4.66±0.85) m/s to (1.99±0.48) m/s, aortic valve area increased from (0.66±0.21) cm² to (2.09±0.67) cm² (all P<0.01). Intraoperative ventricular fibrillation occurred in one patient, while one case exhibited moderate prosthetic valve regurgitation and paravalvular leakage post-procedure. At 12-month follow-up, sustained improvements were observed in cardiac function, left ventricular ejection fraction, hemodynamic parameters, and SF-12 quality-of-life scores (all P<0.01). All-cause mortality was 12.5% (4/32), with 13.8% (4/29) developing moderate paravalvular leakage.
CONCLUSIONS
The XcorTM TAVR system demonstrated favorable early outcomes in severe aortic stenosis patients, significantly improving symptoms and hemodynamics while exhibiting excellent performance in preventing malignant arrhythmias and coronary obstruction.
Humans
;
Male
;
Female
;
Aortic Valve Stenosis/surgery*
;
Transcatheter Aortic Valve Replacement/methods*
;
Aged
;
Follow-Up Studies
;
Prospective Studies
;
Treatment Outcome
;
Aged, 80 and over
;
Heart Valve Prosthesis
;
Middle Aged
4.Left bundle branch pacing in a patient with decreased cardiac function after transcatheter aortic valve replacement.
Xinghong LI ; Jubo JIANG ; Sheng'an SU ; Fang ZHOU
Journal of Zhejiang University. Medical sciences 2025;54(2):149-153
A case of an elderly patient with severe aortic insufficiency who carried high risks for surgical valve replacement. After a detailed preoperative evaluation, the patient successfully received transapical transcatheter aortic valve replacement. Postoperatively, complete left bundle branch block developed, resulting in impaired left ventricular function. Despite guideline-directed medical therapy for heart failure, cardiac function showed no significant recovery. At 4.5 months post-surgery, left bundle branch pacing was performed, leading to a marked improvement in cardiac function.
Aged
;
Humans
;
Male
;
Aortic Valve Insufficiency/surgery*
;
Bundle-Branch Block/etiology*
;
Cardiac Pacing, Artificial
;
Postoperative Complications/therapy*
;
Transcatheter Aortic Valve Replacement/adverse effects*
5.A case of coronary artery protection in transcatheter aortic valve replacement of quadricuspid aortic valve.
Zhipeng CHEN ; Dong YANG ; Han ZHANG
Journal of Zhejiang University. Medical sciences 2025;54(2):161-166
A 72-year-old patient with quadricuspid aortic valve underwent transcatheter aortic valve replacement due to severe valve stenosis accompanied by moderate insufficiency. As initially planned, the right coronary artery was protected during the procedure. However, after the artificial valve was released, the left coronary artery was found to be blocked, so a coronary protection stent was implanted in the left coronary artery ostium under the guidance of intravascular ultrasonography. This case indicates that for patients with a quadricuspid aortic valve undergoing transcatheter aortic valve replacement, in addition to preoperative measurement of the aortic root, attention should also be paid to the coronary artery obstruction caused by the displacement of the artificial valve frame during the procedure.
Aged
;
Humans
;
Aortic Valve/surgery*
;
Aortic Valve Stenosis/surgery*
;
Coronary Vessels
;
Stents
;
Transcatheter Aortic Valve Replacement/methods*
6.Advances in stroke after transcatheter aortic valve replacement.
Journal of Zhejiang University. Medical sciences 2025;54(2):167-174
With the clinical generalization and popularization of transcatheter aortic valve replacement (TAVR), cerebrovascular events related to TAVR occur more frequently, which significantly impairs neurocognitive function, increases mortality, and seriously affects prognosis and quality of life in these patients. However, the reported incidence rates of TAVR-related stroke differ in literature due to inconsistent diagnostic criteria. According to the onset time, TAVR-related stroke can be divided into acute (≤24 h), subacute (24 h-30 d), early (31 d-1 year) and late (>1 year) types, and the cause of stroke generally varies according to the onset time. Both surgical (balloon aortic valvuloplasty, types of transcatheter heart valve, alternative access) and non-surgical (valvular calcium burden, bicuspid aortic valve, subclinical leaflet thrombosis, postoperative new-onset atrial fibrillation) can be related to the occurrence of TAVR-related stroke. Postprocedural monitoring, postprocedural antithrombotic therapy, and cerebral embolic protection devices are important for the prevention of TAVR-related stoke. This article reviews the research progress on TAVR-related stroke, focusing on its epidemiology, risk factors and preventive measures, aiming to provide reference for the clinical management of stroke in TAVR.
Humans
;
Transcatheter Aortic Valve Replacement/adverse effects*
;
Stroke/epidemiology*
;
Postoperative Complications/etiology*
;
Aortic Valve Stenosis/surgery*
;
Risk Factors
7.Risk of coronary obstruction and protection strategies in transcatheter aortic valve replacement.
Yanren PENG ; Ruqiong NIE ; Haifeng ZHANG
Journal of Zhejiang University. Medical sciences 2025;54(2):175-182
Transcatheter aortic valve replacement (TAVR) has emerged as the first-line treatment for aortic valve stenosis. Coronary obstruction is a severe complication of TAVR, with mortality rates exceeding 30%. Coronary obstruction can be classified as acute or delayed based on the timing of the onset, and as direct or indirect obstruction according to the underlying mechanism. Risk factors for predicting coronary obstruction include a small sinus of Valsalva diameter, excessively long native leaflets, low coronary height, and small sinotubular junction height and diameter. Accurate preoperative assessment of these anatomical parameters using CT is crucial for selecting the appropriate valve type, size, and implantation depth. Preventive technical strategies for coronary obstruction include intraoperative interventional treatments (such as the "Chimney" stenting technique), leaflet modification (such as the BASILICA technique), and alignment of the annulus and coronaries. These techniques have demonstrated significant efficacy in reducing the incidence of coronary obstruction and associated mortality. This paper reviews the epidemiology, classification, and mechanisms of coronary obstruction, with a particular focus on the identification, prevention, and treatment of high-risk patients. The aim is to highlight the importance of recognizing and managing coronary risks during TAVR and to provide actionable recommendations for the prevention and treatment of coronary obstruction in clinical practice.
Humans
;
Transcatheter Aortic Valve Replacement/methods*
;
Risk Factors
;
Aortic Valve Stenosis/surgery*
;
Postoperative Complications/prevention & control*
;
Coronary Occlusion/etiology*
8.Structural valve deterioration after transcatheter aortic valve replacement: a research update.
Mengyun YAN ; Zhengang ZHAO ; Mao CHEN
Journal of Zhejiang University. Medical sciences 2025;54(2):183-190
Structural valve deterioration (SVD) refers to intrinsic and irreversible pathological changes in the components of prosthetic heart valves, manifesting as fibrosis, calcification, wear and tear, loosening, as well as strut fracture or deformation of the valve framework. These changes ultimately lead to valve stenosis and/or regurgitation.The mechanisms may be related to mechanical stress, immune response and abnormal calcium-phosphorus metabolism. Studies have shown that risk factors for SVD include patient factors (such as age, underlying cardiovascular disease and comorbidities), valve factors (such as material properties, processing techniques, and valve type), and surgical factors (such as valve injury, suboptimal stent expansion, and irregular stent release morphology). Clinical imaging assessment of SVD demonstrates complementary advantages among echocardiography, multi-detector spiral CT and cardiac magnetic resonance imaging, with distinct diagnostic objectives. The primary management strategies for SVD after trans-catheter aortic valve replacement (TAVR) include drug therapy, redo-TAVR, surgical aortic valve replacement (SAVR) and the novel SURPLUS technique. Among them, redo-TAVR has become a common method because of its minimally invasive nature, but it is still necessary to further clarify the patient indications and optimize the surgical strategy. SAVR is reserved for young, low-risk patients; SURPLUS combines the advantages of SAVR and TAVR, making it suitable for cases where redo-TAVR is unfeasible or contraindicated, while the risk of SAVR is excessively high. This article reviews the latest progress of SVD following TAVR treatment to provide reference for research into the durability of bioprosthetic valve and clinical intervention of SVD.
Humans
;
Transcatheter Aortic Valve Replacement/adverse effects*
;
Heart Valve Prosthesis/adverse effects*
;
Prosthesis Failure
;
Aortic Valve/pathology*
;
Aortic Valve Stenosis/surgery*
9.Clinical efficacy and safety of transcatheter aortic valve replacement for patients with severe pure native aortic regurgitation.
Jiantao CHEN ; Yi ZHANG ; Kangni FENG ; Suiqing HUANG ; Hanri XIAO ; Mengya LIANG ; Zhongkai WU
Journal of Zhejiang University. Medical sciences 2025;54(4):529-540
OBJECTIVES:
To evaluate the early clinical efficacy and safety of trans-catheter aortic valve replacement (TAVR) for patients with severe pure native aortic regurgitation (PNAR) who are not suitable for conventional surgical aortic valve replace-ment.
METHODS:
A retrospective analysis was conducted on 48 patients with PNAR who underwent TAVR at the Department of Cardiac Surgery, the First Affiliated Hospital of Sun Yat-sen University between March 2019 and February 2025. These included 25 cases with transfemoral approach (TF-TAVR group) and 23 cases with transapical approach (TA-TAVR group). Efficacy and safety were assessed by analyzing baseline characteristics, all-cause mortality, and procedure-related complications.
RESULTS:
Compared with the TA-TAVR group, the TF-TAVR group exhibited significantly smaller aortic annulus circumference and diameter, left ventricular outflow tract circumference and diameter, diameters of the left, right, and non-coronary sinuses, and sinotubular junction (STJ) diameter, along with a shorter distance from the STJ to the aortic annular plane ring plane, a smaller annulus angle (all P<0.05). Additionally, the TF-TAVR group showed a deeper prosthesis implantation depth relative to the aortic annular plane (P<0.01). The overall technical success rate was 91.67%, and the device success rate was 83.33%. Post-TAVR, both groups demonstrated significant improvement in left ventricular end-diastolic diameter (both P<0.05), while only the TA-TAVR group showed significant reduction in left ventricular end-systolic diameter (P<0.05). For primary outcomes, in-hospital mortality occurred in 2 patients (4.17%). No additional deaths were reported at 60 or 90 d after surgery. During 90-180 d after surgery, one patient in the TF-TAVR group died of sudden cardiac death, and one in the TA-TAVR group died of gastroin-testinal bleeding. During 180 d-1 year after surgery, one patient in the TF-TAVR group died of low cardiac output syndrome. No statistically significant differences were observed in 1-year Kaplan-Meier survival curves between the two groups (P>0.05). No conduction block events occurred in TA-TAVR group during hospitalization or 1-year follow-up, while high-grade atrioventricular block, left bundle branch block, permanent pacemaker implantation occurred in TF-TAVR group during hospitalization (12.00%, 4.00%, and 12.00%, respectively).
CONCLUSIONS
TAVR demonstrates high feasibility and acceptable safety for severe PNAR patients who are not suitable for conventional SAVR. Both TF-TAVR and TA-TAVR show comparable early postoperative efficacy and safety profiles.
Humans
;
Transcatheter Aortic Valve Replacement/adverse effects*
;
Aortic Valve Insufficiency/surgery*
;
Retrospective Studies
;
Male
;
Female
;
Aged
;
Treatment Outcome
;
Aortic Valve/surgery*
;
Aged, 80 and over
;
Heart Valve Prosthesis
10.Experience of the use of cerebral embolic protection device during transcatheter aortic valve replacement.
Lan ZHANG ; Xinmin LIU ; Ziwei XI ; Fei YUAN ; Jing YAO ; Zhengming JIANG ; Yunfeng YAN ; Guangyuan SONG
Journal of Zhejiang University. Medical sciences 2025;54(4):541-548
OBJECTIVES:
To evaluate the feasibility, efficacy and safety of Sentinel cerebral embolic protection device (CEPD) during transcatheter aortic valve replacement (TAVR). This study is a subgroup analysis of the China Moderate to Severe Valvular Heart Disease Registry, which has been registered at the Chinese Clinical Trial Registry (ChiCTR2300075006).
METHODS:
Patients undergoing TAVR with the Sentinel CEPD from October 2023 to September 2024 were retrospectively enrolled. A total of 80 patients were included, with a median age of 72 (68, 76) years, including 52 males (65.0%) and 28 females (35.0%); 62 patients (77.5%) with tricuspid valves, and 18 patients (22.5%) with bicuspid valves; 34 patients (42.5%) with type Ⅰ aortic arch, 24 patients (30.0%) with type Ⅱ aortic arch, 12 patients (15.0%) with type Ⅲ aortic arch, and 10 patients (12.5%) with bovine-type aortic arch. Clinical data of the patients were summarized and analyzed. The primary endpoints were success rate of Sentinel CEPD implantation, as well as all-cause death, symptomatic stroke, transient ischemic attack, and Sentinel CEPD access vessel complications during hospitalization and within 30 days postoperatively.
RESULTS:
In the 80 patients, self-expanding valves were used in 68 cases (85.0%) and balloon-expandable valves in 12 cases (15.0%). Seventy-nine patients (98.8%) successfully underwent TAVR with Sentinel CEPD deployment. Macroscopically visible debris was captured in 92.5% (74/80) by filters of Sentinel CEPD. Although the procedure time for Sentinel CEPD placement was slightly longer in patients with bovine-type aortic arch, there was no statistically significant difference in deployment time among different aortic arch types (P>0.05). During hospitalization and within 30 days postoperatively, only one case of transient ischemic attack occurred, and there was no all-cause mortality, symptomatic stroke, or access-site vascular complications related to the Sentinel CEPD observed.
CONCLUSIONS
The Sentinel CEPD demonstrates high feasibility across aortic arch types, potential efficacy in embolic capture, and excellent safety in TAVR.
Humans
;
Transcatheter Aortic Valve Replacement/adverse effects*
;
Male
;
Female
;
Aged
;
Embolic Protection Devices
;
Retrospective Studies
;
Intracranial Embolism/prevention & control*
;
Aged, 80 and over
;
Aortic Valve Stenosis/surgery*
;
Aortic Valve/surgery*

Result Analysis
Print
Save
E-mail