1.Axial biomechanical performance evaluation of locally-developed modular external fixator.
Jan Francois B. Severo ; Miguel Sandino O. Aljibe ; Anjenina U. Durana ; Jason Pechardo ; Dionella Jitka B. Quinagoran ; Eduardo R. Magdaluyo Jr. ; Emmanuel P. Estrella
Acta Medica Philippina 2026;60(9):25-32
BACKGROUND
In the Philippines and other developing countries, access to high-stability external fixators for traumainduced bone fracture management is limited, as modular external fixators, especially those with snap-on features, are manufactured overseas and are prohibitively expensive for most Filipino patients.
OBJECTIVEThis study aimed to assess the biomechanical performance of a locally-developed modular external fixator prototype for tibial diaphyseal fractures in comparison to available external fixators, such as Roger Anderson and Hoffmann. This provides an initial evaluation for the use of the external fixator as an alternative in terms of its stability.
METHODSUsing axial compression testing compliant with the ASTM F1541-24 standards, the ultimate strength, yield strength, safe strength, and stiffness were measured across various fixator types and tightening mechanisms, with or without the aid of a wrench. Statistical tools such as the t-test assuming equal variances, one-way analysis of variance, and Tukey-Kramer test with a 0.05 level of significance were used to determine any significant differences.
RESULTSThe mechanical performance of the external fixator prototype increased by a factor of 1.5 to 2.5 after the clamps were tightened with the wrench. However, when hand-tightened, it still performed sufficiently, showing a comparable mechanical performance with the Roger Anderson Fixator. In terms of the ultimate, safe, and yield strengths, it performed competitively in comparison with the Hoffmann system. However, there is a significant difference in stiffness between the prototype and the Hoffmann system.
CONCLUSIONThe locally-developed external fixator was comparable biomechanically to the commercially available external fixators and the prototypes in different studies.
Evaluation Studies As Topic ; Developing Countries ; External Fixators ; Fractures, Bone ; Philippines ; Patients
2.Biomechanical analysis of a novel bridging plate for treating Rockwood III acromioclavicular joint dislocation.
Yancai CHEN ; Gaofeng ZHANG ; Shubo LI ; Nianxiang LUO ; Yi ZHANG
Journal of Southern Medical University 2025;45(5):1103-1112
OBJECTIVES:
To assess the biomechanical performance of a novel bridging plate for treating Rockwood III acromioclavicular joint dislocation.
METHODS:
A novel bridging plate structure was designed based on CT data from a patient with Rockwood type III acromioclavicular joint dislocation, and a finite element model of the bridging plate-acromioclavicular joint interaction was constructed. The stress and deformation characteristics and biomechanical compatibility of the plate under post-reduction, normal loading, and impact loading conditions were analyzed to evaluate its fixation mechanism and clinical advantages.
RESULTS:
The stiffness of the bridging system was 27.78 N/mm, close to that of acromioclavicular joint ligaments (26.05 N/mm) and meeting the requirements for flexible deformation. Under normal loading, the maximum stress in the bridging system was 88.29 MPa to sustain physiological activities; under impact loading, the maximum stress reached 480 MPa, and the cable underwent plastic deformation to dissipate energy and effectively buffer local stress concentrations, thereby reducing the risk of rigid bone fractures. The high-stress regions in the bone primarily occurred at the edges of the C1-C4 screw holes. The maximum bone stress was 0.762 MPa under normal loading and 5.963 MPa under impact loading, accounting for 2.86% and 1.66% of the corresponding bolt stresses, respectively.
CONCLUSIONS
The novel bridging plate is better adapted to biomechanical characteristics of the acromioclavicular joint compared to traditional internal fixation. This fixation system provides sufficient stability while allowing physiological micromotion to facilitate postoperative rehabilitation. Significant flexible deformation can occur at the connection between the fixation ring and the cable, and brittle materials should not be used in this region. The issue of stress concentration at the C1-C4 screw holes requires special attention in its clinical application.
Acromioclavicular Joint/surgery*
;
Humans
;
Bone Plates
;
Biomechanical Phenomena
;
Finite Element Analysis
;
Joint Dislocations/surgery*
;
Fracture Fixation, Internal/methods*
3.Biomechanical analysis of three kinds of rigid internal fixation methods for condylar head fractures.
Junhui SUN ; Duoduo LAN ; Dong WANG ; Yao XU ; Zeyu WANG ; Chenchen ZHANG ; Kai ZHANG ; Tao XU
West China Journal of Stomatology 2025;43(1):126-132
OBJECTIVES:
This study aims to analyze the biomechanics of three kinds of rigid internal fixation methods for condylar head fractures.
METHODS:
A three dimensional finite element model of the normal mandible was constructed. It was then used to prepare condylar head fracture finite element model and three kinds of rigid internal fixation finite element model (unilateral tension screw, bilateral tension screw, tension screw+titanium plate). The mechanical characteristics and changes of the mandible condyle under the same mechanical conditions were compared among the three different rigid internal fixation methods.
RESULTS:
The maximum equivalent stress and displacement of the non-free end of condyle under the rigid internal fixation method of unilateral tension screw were 71.03 MPa and 4.72 mm, respectively. The maximum equivalent stress and displacement of the free end of condyle were 78.45 MPa and 4.50 mm, respectively. The maximum stress of fracture suture was 3.27 MPa. The maximum equivalent stress and displacement of the non-free end of condyle under the rigid internal fixation method of bilateral tension screw were 70.52 MPa and 4.00 mm, respectively. The maximum equivalent stress and displacement of the free end of condyle were 72.49 MPa and 3.85 mm, respectively. The maximum stress of fracture suture was 2.33 MPa. The maximum equivalent stress and maximum displacement of the non-free end of condyle under the rigid internal fixation method of tension screw+titanium plate were 67.26 MPa and 2.66 mm, respectively. The maximum equivalent stress and maximum displacement of the free end of condyle were 69.66 MPa and 2.50 mm, respectively. The maximum stress of fracture suture was 2.18 MPa.
CONCLUSIONS
The tension screw+titanium plate rigid internal fixation method is the most conducive to biomechanical distribution for condylar head fractures.
Fracture Fixation, Internal/instrumentation*
;
Mandibular Condyle/surgery*
;
Biomechanical Phenomena
;
Bone Screws
;
Finite Element Analysis
;
Humans
;
Mandibular Fractures/surgery*
;
Bone Plates
;
Titanium
;
Stress, Mechanical
4.Preliminary evaluation of modified anterior splint combined with anterior repositioning splint after successful mandibular manipulation in treatment of acute anterior disc displacement without reduction of temporomandibular joint.
Chao HUANG ; Xingsheng WU ; Zhen ZHAN ; Lin ZHANG ; Lianshui SHI
West China Journal of Stomatology 2025;43(2):262-268
OBJECTIVES:
Magnetic resonance imaging (MRI) was used to study the efficacy of modified anterior splint (MAS) combined with anterior repositioning splint (ARS) in maintaining the normal disk-condyle relationship.
METHODS:
We obtained clinical and imaging data of 10 patients with acute anterior disk displacement without reduction (ADDwoR) of the temporomandibular joint undergoing successful mandibular manipulation in the Affiliated Stomatological Hospital, Nanchang University from January to June 2024. The MAS and ARS were used for subsequent treatment. Maximum active mouth opening (MMO) and visual analogue score (VAS) were recorded before treatment and 2 weeks, 1 month, and 3 months after treatment. The changes in the disk-condyle relationship, disk-condyle angle, joint effusion, and double contour sign before and after treatment were analyzed through MRI.
RESULTS:
Three months after treatment, the MMO increased from 28.90 mm to 46.85 mm, VAS decreased from 4.68 to 0.61, and the disk-condyle angle decreased from 58.84° to 25.99°. MRI analysis found that 6 patients showed normal disk-condyle relationship, 2 patients exhibited displacement with reduction, and 2 patients exhibited ADDwoR. The disk-condyle angle was 58.84° before treatment and 25.99° after treatment.
CONCLUSIONS
The MAS combined with ARS could further stabilize the coordinated disk-condyle relationship and reduce the recurrence rate.
Humans
;
Male
;
Magnetic Resonance Imaging
;
Female
;
Adult
;
Temporomandibular Joint Disorders/diagnostic imaging*
;
Splints
;
Middle Aged
;
Mandible
;
Temporomandibular Joint Disc
;
Occlusal Splints
;
Temporomandibular Joint
;
Young Adult
;
Joint Dislocations/therapy*
5.Digital design and manufacturing method of double constrained split guide for orthodontic miniscrew implantation.
Xin DU ; Aonan WEN ; Zixiang GAO ; Zhihua LI ; Sheng ZHANG ; Yong WANG ; Yijiao ZHAO
West China Journal of Stomatology 2025;43(4):603-612
This study explored a novel digital design and fabrication method for a double constrained split orthodontic miniscrew guide to improve the accuracy and safety of clinical miniscrew implantation and reduce related complications. A patient requiring miniscrew implantation was selected, and data were acquired using cone beam computed tomography (CBCT) and intraoral optical scanning. For the construction of a double constrained split guide including a screw-hole guide and an insertion rod guide, different types of software such as Mimics 24.0, Geomagic wrap 2021, and Materialise magics 21.0 were utilized for 3D reconstruction, model integration, and guide design. The guide was then fabricated via laser metal 3D printing. Model and intraoral try-in results demonstrated that the guide fitted well and was stable. Postoperative CBCT verified that the final miniscrew implantation site was consistent with the preoperative design, and no related complications occurred. This double constrained split orthodontic miniscrew guide provides a precise and safe digital solution for clinical miniscrew implantation.
Humans
;
Bone Screws
;
Cone-Beam Computed Tomography
;
Printing, Three-Dimensional
;
Orthodontic Anchorage Procedures/instrumentation*
;
Imaging, Three-Dimensional
;
Computer-Aided Design
6.Early warning model of postoperative infection of internal fixation device in maxillofacial fracture based on the synthetic minority over-sampling technique algorithm.
Jinfeng JIANG ; Haiyan WANG ; Yanfeng SHI ; Ke XU
West China Journal of Stomatology 2025;43(6):837-844
OBJECTIVES:
This study investigates independent risk factors for postoperative internal fixation device infection in patients with maxillofacial fractures and proposes an early warning model based on the synthetic minority over-sampling technique (SMOTE) algorithm.
METHODS:
A total of 1 104 patients who underwent surgical treatment for maxillofacial fractures at Oral and Maxillofacial Surgery Department, Affiliated Hospital of Nantong University from January 2021 to December 2024 were retrospectively analyzed. The patients were divided into two groups based on the presence of postoperative internal fixation device infection: the infection group (27 cases) and non-infection group (1 077 cases). Clinical data from both groups were collected and subjected to statistical analysis. Univariate and binary Logistic regression analysis were used to identify risk factors for postoperative internal fixation device infection in maxillofacial fractures. Subsequently, a Logistic regression model was established, and the dataset was improved based on the SMOTE algorithm to construct an early warning model with the improved dataset. The prediction performance of the models was compared and validated.
RESULTS:
Among the 1 104 patients who underwent surgical treatment for maxillofacial fractures, 27 cases of postoperative internal fixation device infections were identified, corresponding to an infection rate of 2.45% (27/1 104). Age, diabetes history, fracture severity, and oral hygiene status were all identified as risk factors for postoperative internal fixation device infections in maxillofacial fractures (all P<0.05). The prediction model based on the original data (P1). The prediction model based on the SMOTE algorithm (P2). Receiver operating characteristic (ROC) curve analysis shows that the area under curve (AUC) for the P2 model was 0.882, the P1 model was 0.861, indicating the superior predictive performance of the P2 model. The DeLong test results show that the difference in AUC between the two models was statistically significant (P<0.05).
CONCLUSIONS
Age, diabetes history, postoperative fracture severity, and oral hygiene status are all risk factors for infections associated with internal fixation devices after maxillofacial fracture surgery. The proposed early warning model demonstrated good predictive performance. Medical professionals can utilize this model to effectively intervene and anticipate infections related to internal fixation devices after maxillofacial fracture surgery.
Humans
;
Algorithms
;
Retrospective Studies
;
Male
;
Female
;
Fracture Fixation, Internal/instrumentation*
;
Risk Factors
;
Middle Aged
;
Adult
;
Logistic Models
;
Surgical Wound Infection/epidemiology*
;
Aged
;
Internal Fixators/adverse effects*
;
Maxillofacial Injuries/surgery*
;
Adolescent
7.Biomechanical study on wing shaped titanium plate fixation of acetabular anterior column and posterior hemi-transverse fracture under multiple working conditions.
Jianwu ZHANG ; WURIKAIXI AIYITI ; Gang LYU ; MAIMAIAILI YUSHAN ; Zhiqiang MA ; Chao MA
Journal of Biomedical Engineering 2025;42(2):351-358
This article aims to compare and analyze the biomechanical differences between wing-shaped titanium plates and traditional titanium plates in fixing acetabular anterior column and posterior hemi-transverse (ACPHT) fracture under multiple working conditions using the finite element method. Firstly, four sets of internal fixation models for acetabular ACPHT fractures were established, and the hip joint stress under standing, sitting, forward extension, and abduction conditions was calculated through analysis software. Then, the stress of screws and titanium plates, as well as the stress and displacement of the fracture end face, were analyzed. Research has found that when using wing-shaped titanium plates to fix acetabular ACPHT fractures, the peak stress of screws decreases under all working conditions, while the peak stress of wing-shaped titanium plates decreases under standing and sitting conditions and increases under forward and outward extension conditions. The relative displacement and mean stress of the fracture end face decrease under all working conditions, but the values are higher under forward and outward extension conditions. Wing-shaped titanium plates can reduce the probability of screw fatigue failure when fixing acetabular ACPHT fractures and can bear greater loads under forward and outward extension conditions, improving the mechanical stability of the pelvis. Moreover, the stress on the fracture end surface is more conducive to stimulating fracture healing and promoting bone tissue growth. However, premature forward and outward extension rehabilitation exercises should not be performed.
Titanium
;
Bone Plates
;
Humans
;
Acetabulum/surgery*
;
Fracture Fixation, Internal/methods*
;
Biomechanical Phenomena
;
Finite Element Analysis
;
Bone Screws
;
Fractures, Bone/surgery*
;
Stress, Mechanical
;
Working Conditions
8.Effectiveness of modified single patellar tunnel medial patella femoral ligament reconstruction for recurrent patellar dislocation.
Guoliang WANG ; Li LI ; Fan WANG ; Yixiang DAI ; Hua LI ; Qinglü SHI
Chinese Journal of Reparative and Reconstructive Surgery 2025;39(1):20-25
OBJECTIVE:
To investigate the effectiveness of modified single patellar tunnel medial patella femoral ligament (MPFL) reconstruction in the treatment of recurrent patellar dislocation.
METHODS:
Between January 2023 and June 2023, a total of 61 patients with recurrent patellar dislocation who underwent MPFL reconstruction with autologous semitendinosus were enrolled and divided into 2 groups using random number table method. In the patellar anchor group, 31 patients were treated with MPFL reconstruction with double medial patellar anchors, and 30 patients in the patellar tunnel group were treated with MPFL reconstruction with single patellar tunnel. The femoral ends of both groups were fixed with absorbable compression screws. There was no significant difference in baseline data such as gender, age, side, tibial tubercle-trochlear groove (TT-TG), Q angle, Caton-Deschamps index, number of dislocation, and preoperative Kujala score, preoperative patellar inclination angle ( P>0.05). Patellar tunnel, patellar anchor position, patellar reduction, and the patellar inclination angle were measured by CT scan after operation. Kujala score was used to evaluate the function of knee joint before operation, at 2 weeks and 1, 3, 6, 12 months after operation. Incision aesthetic satisfaction score was performed at 3 months after operation. The signal-to-noise quotient (SNQ) of the transplanted tendon was measured by knee MRI at 12 months after operation to compare the maturity of the graft between the two groups.
RESULTS:
There was no significant difference in operation time and intraoperative blood loss between the two groups ( P>0.05). Knee CT reexamination showed that the patellar tunnel and the patellar anchor position were consistent with the intraoperative fluoroscopy. There was no significant difference in the difference of the patellar inclination angle between the two groups before and after operation ( P>0.05). All patients were followed up 12-14 months (mean, 12.8 months). There was 1 case of patellar anchor suture rejection in patellar anchor group, and the wound healed after debridement and dressing change. During the follow-up, there was no complication such as recurrence of patellar dislocation, infection and postoperative stiffness. The Kujala scores of the two groups significantly improved at each time point after 1 month of operation when compared with those before operation ( P<0.05), and the Kujala scores of the two groups returned to normal levels at 3 months after operation. The Kujala score in the patellar tunnel group was significantly higher than that in the patellar anchor group in the very early stage (2 weeks) ( P<0.05), and there was no significant difference between the two groups at other time points ( P>0.05). Patients in the patellar tunnel group were significantly better than those in the patellar anchor group in the score of incision aesthetic satisfaction at 3 months after operation and the SNQ at 12 months after operation ( P<0.05).
CONCLUSION
Modified single patellar tunnel MPFL reconstruction was used to treat patients with recurrent patellar dislocation without pathological TT-TG. The slide-fixation structure formed by single patellar tunnel positioning provides a variable degree of freedom for the reconstructed MPFL, which shows good effectiveness in the very early stage of the rehabilitation process.
Humans
;
Patellar Dislocation/surgery*
;
Male
;
Female
;
Plastic Surgery Procedures/methods*
;
Adult
;
Patellar Ligament/surgery*
;
Recurrence
;
Treatment Outcome
;
Young Adult
;
Adolescent
;
Patella/surgery*
;
Suture Anchors
;
Hamstring Tendons/transplantation*
;
Knee Joint/surgery*
;
Transplantation, Autologous
9.Three-dimensional binding treatment for avulsion fractures of inferior pole of patella utilizing suture anchor.
Hongqing HE ; Ningkai LI ; Meng LIU ; Jiating LIN ; Qiang WANG ; Yinchang ZHANG
Chinese Journal of Reparative and Reconstructive Surgery 2025;39(1):26-31
OBJECTIVE:
To explore the feasibility and effectiveness of suture anchor double-pulley technique combined with suture three-dimensional binding via bone tunnel technique for avulsion fractures of the inferior pole of the patella.
METHODS:
A clinical data of 38 patients with avulsion fractures of the inferior pole of the patella, who met the selective criteria and were admitted between September 2021 and April 2023, was retrospectively analyzed. The fractures were treated with suture anchor double-pulley technique combined with suture three-dimensional binding via bone tunnel technique in 18 cases (group A) and steel wire tension-band fixation in 20 cases (group B). There was no significant difference in terms of age, gender, cause of fracture, side of fracture, and disease duration between the two groups ( P>0.05). The length of incision, operation time, occurrence of complications, the range of motion of knee joint, and Böstman score of knee joint at last follow-up were recorded. The fracture healing was evaluated through X-ray films and the time of fracture healing was recorded.
RESULTS:
All incisions healed by first intention. The length of incision was significantly shorter in group A than in group B ( P<0.05). There was no significant difference in the operation time between the two groups ( P>0.05). All patients were followed up 12-24 months (mean, 16.1 months). X-ray films showed that all fractures healed and there was no significant difference in the healing time between the two groups ( P>0.05). At last follow-up, the range of motion and Böstman score of the knee joint in group A were significantly better than those in group B ( P<0.05). During follow-up, 1 patient (5.6%) in group A had one anchor mild prolapse and 3 patients (15.0%) occured internal fixation irritation in group B. But there was no significant difference in the incidence of complications between the two groups ( P>0.05).
CONCLUSION
For the avulsion fractures of the inferior pole of the patella, the suture anchor double-pulley technique combined with suture three-dimensional binding via bone tunnel technique has advantages of reliable fixation, small incision, avoidance of secondary operation to remove internal fixator, and fewer complications, with definite effectiveness.
Humans
;
Male
;
Female
;
Patella/surgery*
;
Suture Anchors
;
Fracture Fixation, Internal/instrumentation*
;
Adult
;
Retrospective Studies
;
Middle Aged
;
Fractures, Avulsion/surgery*
;
Treatment Outcome
;
Young Adult
;
Range of Motion, Articular
;
Fracture Healing
;
Adolescent
;
Suture Techniques
;
Knee Joint/physiopathology*
10.TiRobot-assisted minimally invasive treatment of coracoid process fractures of scapula.
Yonghong DAI ; Qingyu LI ; Yanhui ZENG ; Zhengjie WU ; Chunpeng ZHAO ; Junqiang WANG
Chinese Journal of Reparative and Reconstructive Surgery 2025;39(1):40-46
OBJECTIVE:
To explore effectiveness of TiRobot-assisted screw implantation in the treatment of coracoid process fractures of the scapula.
METHODS:
A retrospective analysis was conducted on the clinical data from 24 patients with coracoid process fractures of the scapula admitted between September 2019 and January 2024 and met selection criteria. Among them, 12 patients underwent TiRobot-assisted screw implantation (robot group) and 12 underwent manual screw implantation (control group) during internal fixation. There was no significant difference ( P>0.05) in baseline data such as gender, age, body mass index, disease duration, cause of injury, coracoid process fracture classification, and proportion of patients with associated injuries between the two groups. The incision length, operation time, intraoperative blood loss, hospital stay, accuracy of screw placement, coracoid process fracture healing time, and complications were recorded and compared, as well as pain visual analogue scale (VAS) score, and Constant-Murley score at last follow-up.
RESULTS:
The intraoperative blood loss and incision length in the robot group were significantly lower than those in the control group ( P<0.05); however, there was no significant difference in operation time and hospital stay between the two groups ( P>0.05). All patients were followed up 8-27 months (mean, 17.5 months), and the difference in follow-up time between the two groups was not significant ( P>0.05). At last follow-up, the VAS score for shoulder pain in the robot group was signifncatly lower compared to the control group, and the Constant-Murley score was significantly higher ( P<0.05). In the robot group, 16 screws were implanted intraoperatively, while 13 screws were implanted in the control group. Radiographic re-evaluation showed that the excellent and good rate of screw implantation was higher in the robot group (93.8%, 15/16) than in the control group (61.5%, 8/13), but the difference in the precision of screw implantation between the two groups was not significant ( P>0.05). Four patients in the robot group and 1 in the control group achieved double screws fixation; however, the difference in achieving double screws fixation between the two groups was not significant ( P>0.05). All fractures healed in both groups with 1 case of malunion in the control group. There was no significant difference in healing time between the two groups ( P>0.05). During follow-up, 1 patient in the control group experienced screw loosening and displacement. There was no significant difference in the incidence of screw loosening and fracture malunion between the two groups ( P>0.05).
CONCLUSION
Compared with manual screw implantation, TiRobot-assisted minimally invasive treatment of coracoid process fractures of the scapula can reduce intraoperative blood loss, shorten incision length, alleviate pain, and obtain better promote shoulder joint functional recovery.
Humans
;
Male
;
Female
;
Retrospective Studies
;
Fracture Fixation, Internal/instrumentation*
;
Minimally Invasive Surgical Procedures/instrumentation*
;
Adult
;
Middle Aged
;
Fractures, Bone/surgery*
;
Bone Screws
;
Coracoid Process/surgery*
;
Robotic Surgical Procedures/methods*
;
Scapula/surgery*
;
Treatment Outcome
;
Operative Time
;
Young Adult
;
Length of Stay
;
Blood Loss, Surgical


Result Analysis
Print
Save
E-mail