1.Influence of Patellar Implant Shape on Patellofemoral Contact Pressure Using Finite Element Analysis
Hun Sik CHO ; Hyoung-Taek HONG ; Hyuck Min KWON ; Yong-Gon KOH ; Seong-Mun HWANG ; Kwan Kyu PARK ; Kyoung-Tak KANG
Yonsei Medical Journal 2025;66(6):383-389
Purpose:
This study focused on analyzing the contact pressure and area on different patellar component designs in total knee arthroplasty (TKA) to evaluate biomechanics related to the patellofemoral (PF) joint.
Materials and Methods:
The patellar components studied included the dome design, modified dome design, and anatomical design implants. Using finite element analysis and mechanical testing, the pressure and area were evaluated. The first loading condition was simulated at flexion angles of 0°, 15°, 45°, 90°, 120°, and 150°. The second loading condition was simulated for a clinically relevant scenario, involving a 2-mm medial shift at a flexion angle of 45°.
Results:
For both the modified dome and anatomical designs, the contact area and pressure increased with the flexion angle. The dome design reached its maximum contact area at a flexion angle of 120°. Among the designs, the anatomical design had the largest contact area and a lower contact pressure compared to the dome and modified dome designs. However, when a medial shift of 2 mm was simulated at a 45° flexion angle, which can occur clinically, the anatomical design showed edge contact, leading to higher contact pressure and reduced contact area. In contrast, the modified dome design demonstrated the lowest contact pressure and the greatest contact area under the same shifted conditions.
Conclusion
These findings suggest that the design of the patellar component significantly affects patellar biomechanics and stability. Specifically, the modified dome design showed improved biomechanical effects in clinically relevant scenarios. Therefore, patellar components with a modified dome design are expected to better manage PF joint pain and reduce complications in TKA.
2.Influence of Patellar Implant Shape on Patellofemoral Contact Pressure Using Finite Element Analysis
Hun Sik CHO ; Hyoung-Taek HONG ; Hyuck Min KWON ; Yong-Gon KOH ; Seong-Mun HWANG ; Kwan Kyu PARK ; Kyoung-Tak KANG
Yonsei Medical Journal 2025;66(6):383-389
Purpose:
This study focused on analyzing the contact pressure and area on different patellar component designs in total knee arthroplasty (TKA) to evaluate biomechanics related to the patellofemoral (PF) joint.
Materials and Methods:
The patellar components studied included the dome design, modified dome design, and anatomical design implants. Using finite element analysis and mechanical testing, the pressure and area were evaluated. The first loading condition was simulated at flexion angles of 0°, 15°, 45°, 90°, 120°, and 150°. The second loading condition was simulated for a clinically relevant scenario, involving a 2-mm medial shift at a flexion angle of 45°.
Results:
For both the modified dome and anatomical designs, the contact area and pressure increased with the flexion angle. The dome design reached its maximum contact area at a flexion angle of 120°. Among the designs, the anatomical design had the largest contact area and a lower contact pressure compared to the dome and modified dome designs. However, when a medial shift of 2 mm was simulated at a 45° flexion angle, which can occur clinically, the anatomical design showed edge contact, leading to higher contact pressure and reduced contact area. In contrast, the modified dome design demonstrated the lowest contact pressure and the greatest contact area under the same shifted conditions.
Conclusion
These findings suggest that the design of the patellar component significantly affects patellar biomechanics and stability. Specifically, the modified dome design showed improved biomechanical effects in clinically relevant scenarios. Therefore, patellar components with a modified dome design are expected to better manage PF joint pain and reduce complications in TKA.
3.Influence of Patellar Implant Shape on Patellofemoral Contact Pressure Using Finite Element Analysis
Hun Sik CHO ; Hyoung-Taek HONG ; Hyuck Min KWON ; Yong-Gon KOH ; Seong-Mun HWANG ; Kwan Kyu PARK ; Kyoung-Tak KANG
Yonsei Medical Journal 2025;66(6):383-389
Purpose:
This study focused on analyzing the contact pressure and area on different patellar component designs in total knee arthroplasty (TKA) to evaluate biomechanics related to the patellofemoral (PF) joint.
Materials and Methods:
The patellar components studied included the dome design, modified dome design, and anatomical design implants. Using finite element analysis and mechanical testing, the pressure and area were evaluated. The first loading condition was simulated at flexion angles of 0°, 15°, 45°, 90°, 120°, and 150°. The second loading condition was simulated for a clinically relevant scenario, involving a 2-mm medial shift at a flexion angle of 45°.
Results:
For both the modified dome and anatomical designs, the contact area and pressure increased with the flexion angle. The dome design reached its maximum contact area at a flexion angle of 120°. Among the designs, the anatomical design had the largest contact area and a lower contact pressure compared to the dome and modified dome designs. However, when a medial shift of 2 mm was simulated at a 45° flexion angle, which can occur clinically, the anatomical design showed edge contact, leading to higher contact pressure and reduced contact area. In contrast, the modified dome design demonstrated the lowest contact pressure and the greatest contact area under the same shifted conditions.
Conclusion
These findings suggest that the design of the patellar component significantly affects patellar biomechanics and stability. Specifically, the modified dome design showed improved biomechanical effects in clinically relevant scenarios. Therefore, patellar components with a modified dome design are expected to better manage PF joint pain and reduce complications in TKA.
4.Influence of Patellar Implant Shape on Patellofemoral Contact Pressure Using Finite Element Analysis
Hun Sik CHO ; Hyoung-Taek HONG ; Hyuck Min KWON ; Yong-Gon KOH ; Seong-Mun HWANG ; Kwan Kyu PARK ; Kyoung-Tak KANG
Yonsei Medical Journal 2025;66(6):383-389
Purpose:
This study focused on analyzing the contact pressure and area on different patellar component designs in total knee arthroplasty (TKA) to evaluate biomechanics related to the patellofemoral (PF) joint.
Materials and Methods:
The patellar components studied included the dome design, modified dome design, and anatomical design implants. Using finite element analysis and mechanical testing, the pressure and area were evaluated. The first loading condition was simulated at flexion angles of 0°, 15°, 45°, 90°, 120°, and 150°. The second loading condition was simulated for a clinically relevant scenario, involving a 2-mm medial shift at a flexion angle of 45°.
Results:
For both the modified dome and anatomical designs, the contact area and pressure increased with the flexion angle. The dome design reached its maximum contact area at a flexion angle of 120°. Among the designs, the anatomical design had the largest contact area and a lower contact pressure compared to the dome and modified dome designs. However, when a medial shift of 2 mm was simulated at a 45° flexion angle, which can occur clinically, the anatomical design showed edge contact, leading to higher contact pressure and reduced contact area. In contrast, the modified dome design demonstrated the lowest contact pressure and the greatest contact area under the same shifted conditions.
Conclusion
These findings suggest that the design of the patellar component significantly affects patellar biomechanics and stability. Specifically, the modified dome design showed improved biomechanical effects in clinically relevant scenarios. Therefore, patellar components with a modified dome design are expected to better manage PF joint pain and reduce complications in TKA.
5.Influence of Patellar Implant Shape on Patellofemoral Contact Pressure Using Finite Element Analysis
Hun Sik CHO ; Hyoung-Taek HONG ; Hyuck Min KWON ; Yong-Gon KOH ; Seong-Mun HWANG ; Kwan Kyu PARK ; Kyoung-Tak KANG
Yonsei Medical Journal 2025;66(6):383-389
Purpose:
This study focused on analyzing the contact pressure and area on different patellar component designs in total knee arthroplasty (TKA) to evaluate biomechanics related to the patellofemoral (PF) joint.
Materials and Methods:
The patellar components studied included the dome design, modified dome design, and anatomical design implants. Using finite element analysis and mechanical testing, the pressure and area were evaluated. The first loading condition was simulated at flexion angles of 0°, 15°, 45°, 90°, 120°, and 150°. The second loading condition was simulated for a clinically relevant scenario, involving a 2-mm medial shift at a flexion angle of 45°.
Results:
For both the modified dome and anatomical designs, the contact area and pressure increased with the flexion angle. The dome design reached its maximum contact area at a flexion angle of 120°. Among the designs, the anatomical design had the largest contact area and a lower contact pressure compared to the dome and modified dome designs. However, when a medial shift of 2 mm was simulated at a 45° flexion angle, which can occur clinically, the anatomical design showed edge contact, leading to higher contact pressure and reduced contact area. In contrast, the modified dome design demonstrated the lowest contact pressure and the greatest contact area under the same shifted conditions.
Conclusion
These findings suggest that the design of the patellar component significantly affects patellar biomechanics and stability. Specifically, the modified dome design showed improved biomechanical effects in clinically relevant scenarios. Therefore, patellar components with a modified dome design are expected to better manage PF joint pain and reduce complications in TKA.
6.Busulfan, Melphalan, and Etoposide (BuME) Showed an Equivalent Effect to Busulfan, Cyclophosphamide, and Etoposide (BuCE) as Conditioning Therapy for Autologous Stem Cell Transplantation in Patients with Relapsed or High-Risk Non-Hodgkin’s Lymphoma: A Multicenter Randomized Phase II Study bythe Consortium for Improving Survival of Lymphoma (CISL)
Kyoung Ha KIM ; Jae Hoon LEE ; Mark LEE ; Hoon-Gu KIM ; Young Rok DO ; Yong PARK ; Sung Yong OH ; Ho-Jin SHIN ; Won Seog KIM ; Seong Kyu PARK ; Jee Hyun KONG ; Moo-Rim PARK ; Deok-Hwan YANG ; Jae-Yong KWAK ; Hye Jin KANG ; Yeung-Chul MUN ; Jong-Ho WON
Cancer Research and Treatment 2023;55(1):304-313
Purpose:
High-dose chemotherapy followed by autologous stem cell transplantation (ASCT) is the standard management for relapsed or high-risk non-Hodgkin’s lymphoma (NHL). We reported the busulfan, melphalan, and etoposide (BuME) conditioning regimen was effective in patients with relapsed or high-risk NHL. Moreover, the busulfan, cyclophosphamide, and etoposide (BuCE) conditioning regimen has been used widely in ASCT for NHL. Therefore, based on these encouraging results, this randomized phase II multicenter trial compared the outcomes of BuME and BuCE as conditioning therapies for ASCT in patients with NHL.
Materials and Methods:
Patients were randomly assigned to receive either BuME (n=36) or BuCE (n=39). The BuME regimen was comprised of busulfan (3.2 mg/kg/day, intravenously) administered on days –7, –6, and –5, etoposide (400 mg/m2 intravenously) on days –5 and –4, and melphalan (50 mg/m2/day intravenously) on days –3 and –2. The BuCE regimen was comprised of busulfan (3.2 mg/kg/day intravenously) on days –7, –6, and –5, etoposide (400 mg/m2/day intravenously) on days –5 and –4, and cyclophosphamide (50 mg/kg/day intravenously) on days –3 and –2. The primary endpoint was 2-year progression-free survival (PFS).
Results:
Seventy-five patients were enrolled. Eleven patients (30.5%) in the BuME group and 13 patients (33.3%) in the BuCE group had disease progression or died. The 2-year PFS rate was 65.4% in the BuME group and 60.6% in the BuCE group (p=0.746). There were no non-relapse mortalities within 100 days after transplantation.
Conclusion
There were no significant differences in PFS between the two groups. Therefore, busulfan-based conditioning regimens, BuME and BuCE, may be important treatment substitutes for the BCNU-containing regimens.
7.A Novel Robot-Assisted Kinematic Measure for Children with Attention-Deficit/Hyperactivity Disorder: A Preliminary Study
Aran MIN ; Johanna Inhyang KIM ; Hak Jong NOH ; Moon Sang KIM ; Hyo-Shin LEE ; Mun-Taek CHOI ; Kyuha LEE ; Jun-Ho SEO ; Ga Hyun LEE ; Seong-kyu KANG ; Dong Hyun AHN
Psychiatry Investigation 2021;18(7):645-651
Objective:
Attention-deficit/hyperactivity disorder (ADHD) is a neurodevelopmental disorder characterized by inattention, hyperactivity, and impulsivity. In contrast to neurocognitive measurements of inattention and impulsivity, there has been limited research regarding the objective measurement of hyperactivity in youths with ADHD. The purpose of the present study was to investigate the clinical effectiveness of a newly developed Robot-assisted Kinematic Measure for ADHD (RAKMA) in children with ADHD.
Methods:
In total, 35 children with ADHD aged 5 to 12 years and 50 healthy controls (HCs) were recruited, and the parents completed the Child Behavior Checklist and the Korean ADHD Diagnostic Scale. RAKMA performance was represented by RAKMA stimulus–response and hyperactivity variables. We compared the RAKMA performance of those with ADHD and with that of HCs and also investigated the correlation between the RAKMA variables and ADHD clinical scale scores.
Results:
Significant differences between the ADHD and HC groups were observed regarding most RAKMA variables, including correct reactions, commission errors, omission errors, reaction times, migration distance, and migration speed scores. Significant correlations were detected between various ADHD clinical scale scores and RAKMA variables.
Conclusion
The RAKMA was a clinically useful tool for objectively measuring hyperactivity symptoms in children with ADHD. Further studies with larger samples are warranted.
8.A Novel Robot-Assisted Kinematic Measure for Children with Attention-Deficit/Hyperactivity Disorder: A Preliminary Study
Aran MIN ; Johanna Inhyang KIM ; Hak Jong NOH ; Moon Sang KIM ; Hyo-Shin LEE ; Mun-Taek CHOI ; Kyuha LEE ; Jun-Ho SEO ; Ga Hyun LEE ; Seong-kyu KANG ; Dong Hyun AHN
Psychiatry Investigation 2021;18(7):645-651
Objective:
Attention-deficit/hyperactivity disorder (ADHD) is a neurodevelopmental disorder characterized by inattention, hyperactivity, and impulsivity. In contrast to neurocognitive measurements of inattention and impulsivity, there has been limited research regarding the objective measurement of hyperactivity in youths with ADHD. The purpose of the present study was to investigate the clinical effectiveness of a newly developed Robot-assisted Kinematic Measure for ADHD (RAKMA) in children with ADHD.
Methods:
In total, 35 children with ADHD aged 5 to 12 years and 50 healthy controls (HCs) were recruited, and the parents completed the Child Behavior Checklist and the Korean ADHD Diagnostic Scale. RAKMA performance was represented by RAKMA stimulus–response and hyperactivity variables. We compared the RAKMA performance of those with ADHD and with that of HCs and also investigated the correlation between the RAKMA variables and ADHD clinical scale scores.
Results:
Significant differences between the ADHD and HC groups were observed regarding most RAKMA variables, including correct reactions, commission errors, omission errors, reaction times, migration distance, and migration speed scores. Significant correlations were detected between various ADHD clinical scale scores and RAKMA variables.
Conclusion
The RAKMA was a clinically useful tool for objectively measuring hyperactivity symptoms in children with ADHD. Further studies with larger samples are warranted.
9.Who Really Benefits from 3D-Based Planning of Brachytherapy for Cervical Cancer?.
In Bong HA ; Bae Kwon JEONG ; Ki Mun KANG ; Hojin JEONG ; Yun Hee LEE ; Hoon Sik CHOI ; Jong Hak LEE ; Won Jun CHOI ; Jeong Kyu SHIN ; Jin Ho SONG
Journal of Korean Medical Science 2018;33(18):e135-
BACKGROUND: Although intracavitary radiotherapy (ICR) is essential for the radiation therapy of cervical cancer, few institutions in Korea perform 3-dimensional (3D)-based ICR. To identify patients who would benefit from 3D-based ICR, dosimetric parameters for tumor targets and organs at risk (OARs) were compared between 2-dimensional (2D)- and 3D-based ICR. METHODS: Twenty patients with locally advanced cervical cancer who underwent external beam radiation therapy (EBRT) following 3D-based ICR were retrospectively evaluated. New 2D-based plans based on the Manchester system were developed. Tumor size was measured by magnetic resonance imaging. RESULTS: The mean high risk clinical target volume (HR-CTV) D90 value was about 10% lower for 2D- than for 3D-based plans (88.4% vs. 97.7%; P = 0.068). Tumor coverage did not differ between 2D- and 3D-based plans in patients with tumors ≤ 4 cm at the time of brachytherapy, but the mean HR-CTV D90 values in patients with tumors > 4 cm were significantly higher for 3D-based plans than for 2D-based plans (96.0% vs. 78.1%; P = 0.017). Similar results were found for patients with tumors > 5 cm initially. Other dosimetric parameters for OARs were similar between 2D- and 3D-based plans, except that mean sigmoid D2cc was higher for 2D- than for 3D-based plans (67.5% vs. 58.8%; P = 0.043). CONCLUSION: These findings indicate that 3D-based ICR plans improve tumor coverage while satisfying the dose constraints for OARs. 3D-based ICR should be considered in patients with tumors > 4 cm size at the time of brachytherapy or > 5 cm initially.
Brachytherapy*
;
Colon, Sigmoid
;
Humans
;
Imaging, Three-Dimensional
;
Korea
;
Magnetic Resonance Imaging
;
Organs at Risk
;
Radiotherapy
;
Retrospective Studies
;
Uterine Cervical Neoplasms*
10.Immunoglobulin E-mediated hypersensitivity reaction after intraperitoneal administration of vancomycin.
Mun Ju HWANG ; Jun Young DO ; Eun Woo CHOI ; Joon Hyuk SEO ; Yoon Jung NAM ; Kyung Woo YOON ; Jong Won PARK ; Kyu Hyang CHO ; Seok Hui KANG ; Hyun Jung JIN
Kidney Research and Clinical Practice 2015;34(1):57-59
Intraperitoneal (IP) vancomycin is widely used to treat Gram-positive peritonitis associated with peritoneal dialysis. There have been two cases of red man syndrome (RMS), a vancomycin-specific nonimmunologic reaction, associated with IP vancomycin. However, immune-mediated hypersensitivity reaction to IP vancomycin has not yet been reported. A 49 year old woman on continuous ambulatory peritoneal dialysis developed her first peritonitis episode. The patient was treated with IP vancomycin once/wk for 4 weeks. She experienced mild itching and flushing throughout her body for 1 day after the second treatment. Whenever vancomycin was administered, generalized urticaria and a prickling sensation developed, and the intensity increased gradually; however, these symptoms improved after vancomycin was discontinued. An allergic skin test was performed 6 weeks after the previous urticarial episode, and an intradermal skin test revealed a positive response to vancomycin. To our knowledge, this is the first case report of immunoglobulin E-mediated hypersensitivity reaction to IP vancomycin administration.
Female
;
Flushing
;
Humans
;
Hypersensitivity*
;
Immunoglobulins*
;
Peritoneal Dialysis
;
Peritoneal Dialysis, Continuous Ambulatory
;
Peritonitis
;
Pruritus
;
Sensation
;
Skin Tests
;
Urticaria
;
Vancomycin*

Result Analysis
Print
Save
E-mail