1.Additional Screw Added to the Femoral Neck System Could Enhance the Stability of Pauwel Type III Femoral Neck Fractures:a Finite Element Analysis
Yonghan CHA ; Sunghoon PARK ; Chang-Ho JUNG ; Jin-Woo KIM ; Jun-Il YOO ; Jung-Taek KIM ; Yongho JEON ; Kyeong Jin HAN
Clinics in Orthopedic Surgery 2025;17(2):204-215
Background:
This study explores effective fixation methods for Pauwel type III femoral neck fractures by evaluating the biomechanical benefits of adding a screw to the Femoral Neck System (FNS).
Methods:
Computed tomography (CT) scans of an 82-year-old female patient with an intertrochanteric fracture were used to establish a finite element femur model with heterogeneous material properties. Finite element models of Pauwel type III fractures were created with and without an additional screw. The central and inferior trajectories of the FNS bolt were examined separately and combined with an additional screw for virtual fixation. Walking and stair-climbing loads were applied.
Results:
With the addition of a screw, both peak maximum and minimum principal strains consistently stayed comparable or decreased in models with both central and inferior bolt trajectories, while the volume of elements with principal strain exceeding 1% decreased by more than half. The peak von Mises stress observed in the implants ranged from 215.7 to 359.3 MPa, remaining below the titanium alloy's yield strength of 800 MPa. For normal walking, the addition of a screw to the central bolt trajectory model decreased the fracture gap by 50.6% and reduced sliding distance by 8.6%. For the inferior bolt trajectory, the gap was reduced by 57.9% and sliding distance by 25.0%.Under stair-climbing conditions, these improvements were also evident; the central trajectory model saw a halved fracture gap and a 7.9% decrease in sliding distance, while the inferior trajectory model experienced a 55.7% gap reduction and a 27.2% decrease in sliding distance. The additional screw increased the area ratio of the fracture site experiencing interfragmentary compression 34%–39%, while the additional screw alleviated peak interfragmentary compression by 12%–18% under both normal walking and stair-climbing conditions.
Conclusions
The addition of a screw reduced the fracture gap, sliding distance, and peak interfragmentary compression, while increasing the area ratio of interfragmentary compression under both walking and stair-climbing loads, regardless of the FNS bolt trajectory, suggesting a better mechanical environment for fracture healing.
2.Powered Tool for the Removal of a Well-Fixed Acetabular Cup: A Comparative Experimental Study
Jun Young CHUNG ; Yonghan CHA ; Chang-Ho JUNG ; Jin-Woo KIM ; Jun-Il YOO ; Jung-Taek KIM ; Yongho JEON
Clinics in Orthopedic Surgery 2025;17(3):372-380
Background:
The removal of a well-fixed acetabular cup is a challenging, labor-intensive, and time-consuming step during revision hip arthroplasty. Although the advent of the manual osteotome, Explant, has simplified the procedure, it is still a stressful process as it dissipates the surgeon’s strength and time and risks an iatrogenic pelvic fracture. Recently, EZX, a powered tool for extraction of well-fixed acetabular cups with semicircular blade was invented. This study aimed to compare Explant and EZX in an experimental condition for their efficacy and safety.
Methods:
Cementless acetabular cups were press-fitted to 20 hemipelvic polyurethane models using foam adhesives. Ten cups were removed with each tool for comparison of the elapsed time, loads on the entire hemipelvis, periacetabular strain and temperature, volume of periacetabular bone removed, and diameter of the remaining acetabular rim. Strains and loads were quantitatively assessed using strain gauges and load cells for precise and reliable measurements.
Results:
The mean duration required to remove a well-fixed cup with EZX was 38.5 seconds (range, 25–55), whereas that with Explant was 543.7 seconds (range, 214–1,051) (p < 0.001). The load on the entire hemipelvis with EZX (mean, 9.1 kgf; range, 6.4–11.3) was 33% lower than that with Explant (mean, 13.6 kgf; range, 9.2–17.1) (p < 0.001). The periacetabular peak strains at the 3 positions with EZX were significantly lower than those with Explant (p < 0.001). The temperature during the removal did not differ significantly between the 2 tools. Although the mean volume of bone loss with Explant was 2.4 mL more than that with EZX (p < 0.001), the mean diameters of the remaining acetabular rim were not significantly different, measuring 54.1 mm with both tools.
Conclusions
The present experiment revealed that a well-fixed cup could be removed using a powered tool with less strength and time and less load on the entire pelvis. Although the powered tool removed a larger volume of bone, the diameters of the remaining acetabular rims were equivalent. This tool may help surgeons remove well-fixed cups in a short time and reduce the deforming load on the bone around the cup without increasing the size of the subsequent reconstruction cup.
3.Clinical Practice Guidelines for Dementia: Recommendations for Cholinesterase Inhibitors and Memantine
Yeshin KIM ; Dong Woo KANG ; Geon Ha KIM ; Ko Woon KIM ; Hee-Jin KIM ; Seunghee NA ; Kee Hyung PARK ; Young Ho PARK ; Gihwan BYEON ; Jeewon SUH ; Joon Hyun SHIN ; YongSoo SHIM ; YoungSoon YANG ; Yoo Hyun UM ; Seong-il OH ; Sheng-Min WANG ; Bora YOON ; Sun Min LEE ; Juyoun LEE ; Jin San LEE ; Jae-Sung LIM ; Young Hee JUNG ; Juhee CHIN ; Hyemin JANG ; Miyoung CHOI ; Yun Jeong HONG ; Hak Young RHEE ; Jae-Won JANG ;
Dementia and Neurocognitive Disorders 2025;24(1):1-23
Background:
and Purpose: This clinical practice guideline provides evidence-based recommendations for treatment of dementia, focusing on cholinesterase inhibitors and N-methyl-D-aspartate (NMDA) receptor antagonists for Alzheimer’s disease (AD) and other types of dementia.
Methods:
Using the Population, Intervention, Comparison, Outcomes (PICO) framework, we developed key clinical questions and conducted systematic literature reviews. A multidisciplinary panel of experts, organized by the Korean Dementia Association, evaluated randomized controlled trials and observational studies. Recommendations were graded for evidence quality and strength using Grading of Recommendations, Assessment, Development, and Evaluation (GRADE) methodology.
Results:
Three main recommendations are presented: (1) For AD, cholinesterase inhibitors (donepezil, rivastigmine, galantamine) are strongly recommended for improving cognition and daily function based on moderate evidence; (2) Cholinesterase inhibitors are conditionally recommended for vascular dementia and Parkinson’s disease dementia, with a strong recommendation for Lewy body dementia; (3) For moderate to severe AD, NMDA receptor antagonist (memantine) is strongly recommended, demonstrating significant cognitive and functional improvements. Both drug classes showed favorable safety profiles with manageable side effects.
Conclusions
This guideline offers standardized, evidence-based pharmacologic recommendations for dementia management, with specific guidance on cholinesterase inhibitors and NMDA receptor antagonists. It aims to support clinical decision-making and improve patient outcomes in dementia care. Further updates will address emerging treatments, including amyloid-targeting therapies, to reflect advances in dementia management.
4.Study Protocol of Expanded Multicenter Prospective Cohort Study of Active Surveillance on Papillary Thyroid Microcarcinoma (MAeSTro-EXP)
Jae Hoon MOON ; Eun Kyung LEE ; Wonjae CHA ; Young Jun CHAI ; Sun Wook CHO ; June Young CHOI ; Sung Yong CHOI ; A Jung CHU ; Eun-Jae CHUNG ; Yul HWANGBO ; Woo-Jin JEONG ; Yuh-Seog JUNG ; Kyungsik KIM ; Min Joo KIM ; Su-jin KIM ; Woochul KIM ; Yoo Hyung KIM ; Chang Yoon LEE ; Ji Ye LEE ; Kyu Eun LEE ; Young Ki LEE ; Hunjong LIM ; Do Joon PARK ; Sue K. PARK ; Chang Hwan RYU ; Junsun RYU ; Jungirl SEOK ; Young Shin SONG ; Ka Hee YI ; Hyeong Won YU ; Eleanor WHITE ; Katerina MASTROCOSTAS ; Roderick J. CLIFTON-BLIGH ; Anthony GLOVER ; Matti L. GILD ; Ji-hoon KIM ; Young Joo PARK
Endocrinology and Metabolism 2025;40(2):236-246
Background:
Active surveillance (AS) has emerged as a viable management strategy for low-risk papillary thyroid microcarcinoma (PTMC), following pioneering trials at Kuma Hospital and the Cancer Institute Hospital in Japan. Numerous prospective cohort studies have since validated AS as a management option for low-risk PTMC, leading to its inclusion in thyroid cancer guidelines across various countries. From 2016 to 2020, the Multicenter Prospective Cohort Study of Active Surveillance on Papillary Thyroid Microcarcinoma (MAeSTro) enrolled 1,177 patients, providing comprehensive data on PTMC progression, sonographic predictors of progression, quality of life, surgical outcomes, and cost-effectiveness when comparing AS to immediate surgery. The second phase of MAeSTro (MAeSTro-EXP) expands AS to low-risk papillary thyroid carcinoma (PTC) tumors larger than 1 cm, driven by the hypothesis that overall risk assessment outweighs absolute tumor size in surgical decision-making.
Methods:
This protocol aims to address whether limiting AS to tumors smaller than 1 cm may result in unnecessary surgeries for low-risk PTCs detected during their rapid initial growth phase. By expanding the AS criteria to include tumors up to 1.5 cm, while simultaneously refining and standardizing the criteria for risk assessment and disease progression, we aim to minimize overtreatment and maintain rigorous monitoring to improve patient outcomes.
Conclusion
This study will contribute to optimizing AS guidelines and enhance our understanding of the natural course and appropriate management of low-risk PTCs. Additionally, MAeSTro-EXP involves a multinational collaboration between South Korea and Australia. This cross-country study aims to identify cultural and racial differences in the management of low-risk PTC, thereby enriching the global understanding of AS practices and their applicability across diverse populations.
5.Additional Screw Added to the Femoral Neck System Could Enhance the Stability of Pauwel Type III Femoral Neck Fractures:a Finite Element Analysis
Yonghan CHA ; Sunghoon PARK ; Chang-Ho JUNG ; Jin-Woo KIM ; Jun-Il YOO ; Jung-Taek KIM ; Yongho JEON ; Kyeong Jin HAN
Clinics in Orthopedic Surgery 2025;17(2):204-215
Background:
This study explores effective fixation methods for Pauwel type III femoral neck fractures by evaluating the biomechanical benefits of adding a screw to the Femoral Neck System (FNS).
Methods:
Computed tomography (CT) scans of an 82-year-old female patient with an intertrochanteric fracture were used to establish a finite element femur model with heterogeneous material properties. Finite element models of Pauwel type III fractures were created with and without an additional screw. The central and inferior trajectories of the FNS bolt were examined separately and combined with an additional screw for virtual fixation. Walking and stair-climbing loads were applied.
Results:
With the addition of a screw, both peak maximum and minimum principal strains consistently stayed comparable or decreased in models with both central and inferior bolt trajectories, while the volume of elements with principal strain exceeding 1% decreased by more than half. The peak von Mises stress observed in the implants ranged from 215.7 to 359.3 MPa, remaining below the titanium alloy's yield strength of 800 MPa. For normal walking, the addition of a screw to the central bolt trajectory model decreased the fracture gap by 50.6% and reduced sliding distance by 8.6%. For the inferior bolt trajectory, the gap was reduced by 57.9% and sliding distance by 25.0%.Under stair-climbing conditions, these improvements were also evident; the central trajectory model saw a halved fracture gap and a 7.9% decrease in sliding distance, while the inferior trajectory model experienced a 55.7% gap reduction and a 27.2% decrease in sliding distance. The additional screw increased the area ratio of the fracture site experiencing interfragmentary compression 34%–39%, while the additional screw alleviated peak interfragmentary compression by 12%–18% under both normal walking and stair-climbing conditions.
Conclusions
The addition of a screw reduced the fracture gap, sliding distance, and peak interfragmentary compression, while increasing the area ratio of interfragmentary compression under both walking and stair-climbing loads, regardless of the FNS bolt trajectory, suggesting a better mechanical environment for fracture healing.
6.Powered Tool for the Removal of a Well-Fixed Acetabular Cup: A Comparative Experimental Study
Jun Young CHUNG ; Yonghan CHA ; Chang-Ho JUNG ; Jin-Woo KIM ; Jun-Il YOO ; Jung-Taek KIM ; Yongho JEON
Clinics in Orthopedic Surgery 2025;17(3):372-380
Background:
The removal of a well-fixed acetabular cup is a challenging, labor-intensive, and time-consuming step during revision hip arthroplasty. Although the advent of the manual osteotome, Explant, has simplified the procedure, it is still a stressful process as it dissipates the surgeon’s strength and time and risks an iatrogenic pelvic fracture. Recently, EZX, a powered tool for extraction of well-fixed acetabular cups with semicircular blade was invented. This study aimed to compare Explant and EZX in an experimental condition for their efficacy and safety.
Methods:
Cementless acetabular cups were press-fitted to 20 hemipelvic polyurethane models using foam adhesives. Ten cups were removed with each tool for comparison of the elapsed time, loads on the entire hemipelvis, periacetabular strain and temperature, volume of periacetabular bone removed, and diameter of the remaining acetabular rim. Strains and loads were quantitatively assessed using strain gauges and load cells for precise and reliable measurements.
Results:
The mean duration required to remove a well-fixed cup with EZX was 38.5 seconds (range, 25–55), whereas that with Explant was 543.7 seconds (range, 214–1,051) (p < 0.001). The load on the entire hemipelvis with EZX (mean, 9.1 kgf; range, 6.4–11.3) was 33% lower than that with Explant (mean, 13.6 kgf; range, 9.2–17.1) (p < 0.001). The periacetabular peak strains at the 3 positions with EZX were significantly lower than those with Explant (p < 0.001). The temperature during the removal did not differ significantly between the 2 tools. Although the mean volume of bone loss with Explant was 2.4 mL more than that with EZX (p < 0.001), the mean diameters of the remaining acetabular rim were not significantly different, measuring 54.1 mm with both tools.
Conclusions
The present experiment revealed that a well-fixed cup could be removed using a powered tool with less strength and time and less load on the entire pelvis. Although the powered tool removed a larger volume of bone, the diameters of the remaining acetabular rims were equivalent. This tool may help surgeons remove well-fixed cups in a short time and reduce the deforming load on the bone around the cup without increasing the size of the subsequent reconstruction cup.
7.Clinical Practice Guidelines for Dementia: Recommendations for Cholinesterase Inhibitors and Memantine
Yeshin KIM ; Dong Woo KANG ; Geon Ha KIM ; Ko Woon KIM ; Hee-Jin KIM ; Seunghee NA ; Kee Hyung PARK ; Young Ho PARK ; Gihwan BYEON ; Jeewon SUH ; Joon Hyun SHIN ; YongSoo SHIM ; YoungSoon YANG ; Yoo Hyun UM ; Seong-il OH ; Sheng-Min WANG ; Bora YOON ; Sun Min LEE ; Juyoun LEE ; Jin San LEE ; Jae-Sung LIM ; Young Hee JUNG ; Juhee CHIN ; Hyemin JANG ; Miyoung CHOI ; Yun Jeong HONG ; Hak Young RHEE ; Jae-Won JANG ;
Dementia and Neurocognitive Disorders 2025;24(1):1-23
Background:
and Purpose: This clinical practice guideline provides evidence-based recommendations for treatment of dementia, focusing on cholinesterase inhibitors and N-methyl-D-aspartate (NMDA) receptor antagonists for Alzheimer’s disease (AD) and other types of dementia.
Methods:
Using the Population, Intervention, Comparison, Outcomes (PICO) framework, we developed key clinical questions and conducted systematic literature reviews. A multidisciplinary panel of experts, organized by the Korean Dementia Association, evaluated randomized controlled trials and observational studies. Recommendations were graded for evidence quality and strength using Grading of Recommendations, Assessment, Development, and Evaluation (GRADE) methodology.
Results:
Three main recommendations are presented: (1) For AD, cholinesterase inhibitors (donepezil, rivastigmine, galantamine) are strongly recommended for improving cognition and daily function based on moderate evidence; (2) Cholinesterase inhibitors are conditionally recommended for vascular dementia and Parkinson’s disease dementia, with a strong recommendation for Lewy body dementia; (3) For moderate to severe AD, NMDA receptor antagonist (memantine) is strongly recommended, demonstrating significant cognitive and functional improvements. Both drug classes showed favorable safety profiles with manageable side effects.
Conclusions
This guideline offers standardized, evidence-based pharmacologic recommendations for dementia management, with specific guidance on cholinesterase inhibitors and NMDA receptor antagonists. It aims to support clinical decision-making and improve patient outcomes in dementia care. Further updates will address emerging treatments, including amyloid-targeting therapies, to reflect advances in dementia management.
8.Study Protocol of Expanded Multicenter Prospective Cohort Study of Active Surveillance on Papillary Thyroid Microcarcinoma (MAeSTro-EXP)
Jae Hoon MOON ; Eun Kyung LEE ; Wonjae CHA ; Young Jun CHAI ; Sun Wook CHO ; June Young CHOI ; Sung Yong CHOI ; A Jung CHU ; Eun-Jae CHUNG ; Yul HWANGBO ; Woo-Jin JEONG ; Yuh-Seog JUNG ; Kyungsik KIM ; Min Joo KIM ; Su-jin KIM ; Woochul KIM ; Yoo Hyung KIM ; Chang Yoon LEE ; Ji Ye LEE ; Kyu Eun LEE ; Young Ki LEE ; Hunjong LIM ; Do Joon PARK ; Sue K. PARK ; Chang Hwan RYU ; Junsun RYU ; Jungirl SEOK ; Young Shin SONG ; Ka Hee YI ; Hyeong Won YU ; Eleanor WHITE ; Katerina MASTROCOSTAS ; Roderick J. CLIFTON-BLIGH ; Anthony GLOVER ; Matti L. GILD ; Ji-hoon KIM ; Young Joo PARK
Endocrinology and Metabolism 2025;40(2):236-246
Background:
Active surveillance (AS) has emerged as a viable management strategy for low-risk papillary thyroid microcarcinoma (PTMC), following pioneering trials at Kuma Hospital and the Cancer Institute Hospital in Japan. Numerous prospective cohort studies have since validated AS as a management option for low-risk PTMC, leading to its inclusion in thyroid cancer guidelines across various countries. From 2016 to 2020, the Multicenter Prospective Cohort Study of Active Surveillance on Papillary Thyroid Microcarcinoma (MAeSTro) enrolled 1,177 patients, providing comprehensive data on PTMC progression, sonographic predictors of progression, quality of life, surgical outcomes, and cost-effectiveness when comparing AS to immediate surgery. The second phase of MAeSTro (MAeSTro-EXP) expands AS to low-risk papillary thyroid carcinoma (PTC) tumors larger than 1 cm, driven by the hypothesis that overall risk assessment outweighs absolute tumor size in surgical decision-making.
Methods:
This protocol aims to address whether limiting AS to tumors smaller than 1 cm may result in unnecessary surgeries for low-risk PTCs detected during their rapid initial growth phase. By expanding the AS criteria to include tumors up to 1.5 cm, while simultaneously refining and standardizing the criteria for risk assessment and disease progression, we aim to minimize overtreatment and maintain rigorous monitoring to improve patient outcomes.
Conclusion
This study will contribute to optimizing AS guidelines and enhance our understanding of the natural course and appropriate management of low-risk PTCs. Additionally, MAeSTro-EXP involves a multinational collaboration between South Korea and Australia. This cross-country study aims to identify cultural and racial differences in the management of low-risk PTC, thereby enriching the global understanding of AS practices and their applicability across diverse populations.
9.Additional Screw Added to the Femoral Neck System Could Enhance the Stability of Pauwel Type III Femoral Neck Fractures:a Finite Element Analysis
Yonghan CHA ; Sunghoon PARK ; Chang-Ho JUNG ; Jin-Woo KIM ; Jun-Il YOO ; Jung-Taek KIM ; Yongho JEON ; Kyeong Jin HAN
Clinics in Orthopedic Surgery 2025;17(2):204-215
Background:
This study explores effective fixation methods for Pauwel type III femoral neck fractures by evaluating the biomechanical benefits of adding a screw to the Femoral Neck System (FNS).
Methods:
Computed tomography (CT) scans of an 82-year-old female patient with an intertrochanteric fracture were used to establish a finite element femur model with heterogeneous material properties. Finite element models of Pauwel type III fractures were created with and without an additional screw. The central and inferior trajectories of the FNS bolt were examined separately and combined with an additional screw for virtual fixation. Walking and stair-climbing loads were applied.
Results:
With the addition of a screw, both peak maximum and minimum principal strains consistently stayed comparable or decreased in models with both central and inferior bolt trajectories, while the volume of elements with principal strain exceeding 1% decreased by more than half. The peak von Mises stress observed in the implants ranged from 215.7 to 359.3 MPa, remaining below the titanium alloy's yield strength of 800 MPa. For normal walking, the addition of a screw to the central bolt trajectory model decreased the fracture gap by 50.6% and reduced sliding distance by 8.6%. For the inferior bolt trajectory, the gap was reduced by 57.9% and sliding distance by 25.0%.Under stair-climbing conditions, these improvements were also evident; the central trajectory model saw a halved fracture gap and a 7.9% decrease in sliding distance, while the inferior trajectory model experienced a 55.7% gap reduction and a 27.2% decrease in sliding distance. The additional screw increased the area ratio of the fracture site experiencing interfragmentary compression 34%–39%, while the additional screw alleviated peak interfragmentary compression by 12%–18% under both normal walking and stair-climbing conditions.
Conclusions
The addition of a screw reduced the fracture gap, sliding distance, and peak interfragmentary compression, while increasing the area ratio of interfragmentary compression under both walking and stair-climbing loads, regardless of the FNS bolt trajectory, suggesting a better mechanical environment for fracture healing.
10.Powered Tool for the Removal of a Well-Fixed Acetabular Cup: A Comparative Experimental Study
Jun Young CHUNG ; Yonghan CHA ; Chang-Ho JUNG ; Jin-Woo KIM ; Jun-Il YOO ; Jung-Taek KIM ; Yongho JEON
Clinics in Orthopedic Surgery 2025;17(3):372-380
Background:
The removal of a well-fixed acetabular cup is a challenging, labor-intensive, and time-consuming step during revision hip arthroplasty. Although the advent of the manual osteotome, Explant, has simplified the procedure, it is still a stressful process as it dissipates the surgeon’s strength and time and risks an iatrogenic pelvic fracture. Recently, EZX, a powered tool for extraction of well-fixed acetabular cups with semicircular blade was invented. This study aimed to compare Explant and EZX in an experimental condition for their efficacy and safety.
Methods:
Cementless acetabular cups were press-fitted to 20 hemipelvic polyurethane models using foam adhesives. Ten cups were removed with each tool for comparison of the elapsed time, loads on the entire hemipelvis, periacetabular strain and temperature, volume of periacetabular bone removed, and diameter of the remaining acetabular rim. Strains and loads were quantitatively assessed using strain gauges and load cells for precise and reliable measurements.
Results:
The mean duration required to remove a well-fixed cup with EZX was 38.5 seconds (range, 25–55), whereas that with Explant was 543.7 seconds (range, 214–1,051) (p < 0.001). The load on the entire hemipelvis with EZX (mean, 9.1 kgf; range, 6.4–11.3) was 33% lower than that with Explant (mean, 13.6 kgf; range, 9.2–17.1) (p < 0.001). The periacetabular peak strains at the 3 positions with EZX were significantly lower than those with Explant (p < 0.001). The temperature during the removal did not differ significantly between the 2 tools. Although the mean volume of bone loss with Explant was 2.4 mL more than that with EZX (p < 0.001), the mean diameters of the remaining acetabular rim were not significantly different, measuring 54.1 mm with both tools.
Conclusions
The present experiment revealed that a well-fixed cup could be removed using a powered tool with less strength and time and less load on the entire pelvis. Although the powered tool removed a larger volume of bone, the diameters of the remaining acetabular rims were equivalent. This tool may help surgeons remove well-fixed cups in a short time and reduce the deforming load on the bone around the cup without increasing the size of the subsequent reconstruction cup.

Result Analysis