1.Exploration on the influence of continuous infusion with long time on infusion precision of infusion pump
Changxin ZHANG ; Yu WANG ; Wenbo ZHOU ; Huiqing LIU ; Jing ZHAO
China Medical Equipment 2025;22(7):177-179,187
Objective:To investigate the influence of continuous infusion with long time on infusion precision of infusion pump,and establish a predictive model for flow velocity to predict the change of flow velocity of infusion pump after infusion with long time,so as to provide reference for the use of infusion pump in clinical treatment.Methods:Six infusion sets of different models,numbered A to F,were selected.A 24-hour continuous infusion experiment was conducted using an infusion pump with the above six infusion sets.The flow rate of the infusion pump was set to 100 mL/h,and real-time data of the infusion speed was recorded to evaluate the performance changes of the infusion pump under long-term continuous infusion.A flow rate prediction model for the infusion pump was established based on the research data,and the accuracy of the model was verified.Results:After 24 hours of continuous infusion,the infusion accuracy of infusion sets A to F showed varying degrees of decline over time.The final infusion accuracies were 96.47%,97.95%,92.56%,88.41%,89.26%,93.93%,respectively,with corresponding flow rate errors of 3.53%,2.05%,7.44%,11.59%,10.74%,6.07%.The prediction model established in this study was used to predict the flow rate of the infusion pump during 24-36 hours of continuous infusion.Compared with the actual flow rate under the same conditions,the difference was not statistically significant(P>0.05).Conclusion:Long-term continuous infusion affects the infusion accuracy of the infusion pump,resulting in a decrease in infusion speed.The established flow rate prediction model can accurately predict the flow rate of the infusion pump after long-term continuous infusion,providing a useful reference for the clinical application of infusion pumps.
2.Detection method for the performance of ultrasound probes based on pulse echo
Yu WANG ; Changxin ZHANG ; Jing ZHAO ; Guanghan LI ; Wenbo ZHOU ; Xinyuan DONG
China Medical Equipment 2025;22(3):157-159
The ultrasound probe is the most important accessory in ultrasound equipment,which adopted the piezoelectric vibrator as its core component,and it can complete ultrasonic emission and receiving echoes.A deep understanding for the operational characteristics of the probe's array elements is important premise that user can accurately judge the running situation of the probe.This study employed pulse-echo technique to detect sensitivity and capacitance of the array elements of ultrasonic probe.The detected data of 60 probes at China-Japan Friendship Hospital were collected,and 75%of the 60 probes existed residual couplant,and 80%of them existed indentation or damage to the outer skin.The tested results of leakage current indicated that 1 probe(1.7%)existed excessively high leakage current.In test for array elements,2 probes(3.4%)displayed partial damage at array elements,and 6 probes(10.2%)were irreparably damaged.The preventive test for the diagnostic equipment with ultrasound is helpful to establish a quality control system for ultrasound,and to provide analysis materials with reliability in using diagnostic equipment with ultrasound,and to also provide decision-making basis for equipment departments and clinical department of using equipment in hospital.
3.Correlation Between Traditional Chinese Medicine Syndrome Type of Unstable Angina Pectoris and Iron Metabolism Indices
Jiye CHEN ; Min WU ; Changxin SUN ; Xiaonan ZHANG ; Zeping WANG ; Longtao LIU
Chinese Journal of Experimental Traditional Medical Formulae 2025;31(20):126-136
ObjectiveTo investigate the correlation between iron metabolism parameters and various syndrome types of unstable angina pectoris (UAP). MethodsA cross-sectional study was conducted from October 2021 to October 2023, encompassing 213 patients diagnosed with UAP at Xiyuan Hospital of Chinese Academy of Chinese Medical Sciences. Additionally, 30 healthy individuals were selected as control cases. Single-factor analysis was used to investigate the differences in clinical data among different Traditional Chinese Medicine (TCM) syndrome types of UAP and their correlation with iron metabolism indices. The study conducted a comparative analysis of the aforementioned clinical data among patients with and without heat-toxic and blood-stasis syndrome. Logistic regression was used to analyze the correlation between TCM syndrome types and related factors. The receiver operator characteristic (ROC) curve was employed to assess the predictive value of iron metabolism indices, along with their sensitivity and specificity. ResultsCompared to those in the control group, serum iron (SI) and serum ferritin (SF) levels were significantly increased in the UAP group (P<0.01), while transferrin (TRF) and total iron binding capacity (TIBC) levels were decreased (P<0.01). However, there was no significant difference in unsaturated iron binding capacity (UIBC). Multivariate binary Logistic regression analysis identified apolipoprotein A1 (ApoA1), homocysteine (HCY), high-sensitivity C-reactive protein (hs-CRP), and SF as independent influencing factors for the UAP patients (P<0.05, P<0.01). Additionally, statistically significant differences were observed in SI, SF, TRF, and TIBC among 213 patients with different TCM types (P<0.01). Patients with heat-toxic and blood-stasis syndrome had higher SI and SF values than those without the syndrome (P<0.01), while their TIBC and TRF values were lower (P<0.01). Multivariate binary logistic regression analysis showed that SI and LDL-C levels were closely associated with the differentiation of heat-toxic and blood-stasis syndrome. ConclusionUAP patients often experience iron metabolism disorders, and the heat-toxic and blood-stasis syndrome are significantly correlated with iron metabolism parameters. The SI and LDL-C levels have high specificity and sensitivity in diagnosing heat-toxic and blood-stasis syndrome.
4.Preliminary study on the implementation of HT for segmented TMLI in adults
Haiyang WANG ; Gaoxiang CHEN ; Xiang HUANG ; Ruigang GE ; Shilei ZHANG ; Changxin YAN ; Chuanbin XIE
China Medical Equipment 2025;22(9):21-26
Objective:To investigate the feasibility of segmented total marrow and lymphoid irradiation(TMLI)using Helical TomoTherapy(HT)in adults,and to analyze the impact of positional errors in the treatment on the actually received dose.Methods:Imaging data from four adult patients with acute myeloid leukemia were treated at the First Medical Center of the Chinese PLA General Hospital between November 2022 and December 2024 were retrospectively selected.The prescribed dose was 12 Gy in 3 fractions,and the heights range of patient was from 162 to 178 cm.The planning target volume(PTV)included the entire bone marrow and lymphatic drainage regions.The mid-femur was used as the dividing landmark,and the segmented location of head-in first(upper segmentation)and foot-in first(lower segmentation)were adopted to design plan for each segmentation.The upper segmentation used helical tomotherapy,and the lower segmentation used tomo direct therapy.The dosimetric parameters of planning target volume,and organ at risk(OAR)included eye lens,testicles and lungs were analyzed.The located computed tomography(CT)images of patients were used to simulate the changes of dosimetric parameters of each planning target region and OAR when the positions of patients moved 3,5,10 and 15 mm to outside of rack aperture.The changes of dose distribution of connection region of upper and lower segmentation were further analyzed.Results:The segmented irradiation technique can better achieve the cover(95%)of whole target region of TMLI under the condition of protecting normal organs.Under the condition with different position errors,the cover rate of the most of target region significant decreased.The largest dose of right lens increased 9.272%than primary under the 5 mm position error,and that of left and right lens increased respectively 40.199%and 67.898%under the 10 mm position error,and the max dose and mean dose of testicles increased respectively 5.847%and 5.240%under the 3mm position error.The mean dose of whole lung increased all under the error at different grades,which increased 3.021%under the 10mm position error.The non-uniformity of dose distribution of connection region increased after the position error was introduced,and the largest variation range reached to 4 Gy.Conclusion:The technique of segmented radiotherapy of Helical Tomotherapy has feasibility for TMLI of adult.But the error of the position of patients can reduce the cover of target region,and decrease the control rate for target region,and can lead to the occurrences of excessive doses on eye lens,testicles,lung and other OAR at the same time,which can increase the probability of occurring OAR injury.It should be high pay attention in treatment and implementation.
5.Segmented Time Study and Optimization Strategy for Clinical Application of Ethos Online Adaptive Radiotherapy.
Dandan ZHANG ; Yuhan KOU ; Shilong ZHU ; Xiaoyu LIU ; Meng NING ; Peichao BAN ; Jinyuan WANG ; Changxin YAN ; Zhongjian JU
Chinese Journal of Medical Instrumentation 2025;49(2):134-140
OBJECTIVE:
To analyze the time characteristics of the Ethos online adaptive radiotherapy (OART) process in clinical practice and provide guidance for the comprehensive optimization of each stage of adaptive radiotherapy.
METHODS:
The study involved 61 patients with cervical, rectal, gastric, lung, esophageal, and breast cancers who underwent Ethos OART. The mean ± standard deviation of segmental time, total time, and target volume for these patients were tracked. The time characteristics for different cancer types were evaluated, and the average time for target and organ at risk (OAR) modifications was compared with the average target volume for each cancer type.
RESULTS:
Cervical cancer born the longest total treatment time, while breast cancer had the shortest. For all cancer types except breast cancer, the modification time for target and OAR was the most time-consuming segment. The average time for target and OAR modifications aligned with the trend of the average target volume.
CONCLUSION
The total treatment time for various cancers ranges from 15 to 35 minutes, indicating room for improvement.
Humans
;
Radiotherapy Planning, Computer-Assisted/methods*
;
Neoplasms/radiotherapy*
;
Female
6.Exploration on the influence of continuous infusion with long time on infusion precision of infusion pump
Changxin ZHANG ; Yu WANG ; Wenbo ZHOU ; Huiqing LIU ; Jing ZHAO
China Medical Equipment 2025;22(7):177-179,187
Objective:To investigate the influence of continuous infusion with long time on infusion precision of infusion pump,and establish a predictive model for flow velocity to predict the change of flow velocity of infusion pump after infusion with long time,so as to provide reference for the use of infusion pump in clinical treatment.Methods:Six infusion sets of different models,numbered A to F,were selected.A 24-hour continuous infusion experiment was conducted using an infusion pump with the above six infusion sets.The flow rate of the infusion pump was set to 100 mL/h,and real-time data of the infusion speed was recorded to evaluate the performance changes of the infusion pump under long-term continuous infusion.A flow rate prediction model for the infusion pump was established based on the research data,and the accuracy of the model was verified.Results:After 24 hours of continuous infusion,the infusion accuracy of infusion sets A to F showed varying degrees of decline over time.The final infusion accuracies were 96.47%,97.95%,92.56%,88.41%,89.26%,93.93%,respectively,with corresponding flow rate errors of 3.53%,2.05%,7.44%,11.59%,10.74%,6.07%.The prediction model established in this study was used to predict the flow rate of the infusion pump during 24-36 hours of continuous infusion.Compared with the actual flow rate under the same conditions,the difference was not statistically significant(P>0.05).Conclusion:Long-term continuous infusion affects the infusion accuracy of the infusion pump,resulting in a decrease in infusion speed.The established flow rate prediction model can accurately predict the flow rate of the infusion pump after long-term continuous infusion,providing a useful reference for the clinical application of infusion pumps.
7.Detection method for the performance of ultrasound probes based on pulse echo
Yu WANG ; Changxin ZHANG ; Jing ZHAO ; Guanghan LI ; Wenbo ZHOU ; Xinyuan DONG
China Medical Equipment 2025;22(3):157-159
The ultrasound probe is the most important accessory in ultrasound equipment,which adopted the piezoelectric vibrator as its core component,and it can complete ultrasonic emission and receiving echoes.A deep understanding for the operational characteristics of the probe's array elements is important premise that user can accurately judge the running situation of the probe.This study employed pulse-echo technique to detect sensitivity and capacitance of the array elements of ultrasonic probe.The detected data of 60 probes at China-Japan Friendship Hospital were collected,and 75%of the 60 probes existed residual couplant,and 80%of them existed indentation or damage to the outer skin.The tested results of leakage current indicated that 1 probe(1.7%)existed excessively high leakage current.In test for array elements,2 probes(3.4%)displayed partial damage at array elements,and 6 probes(10.2%)were irreparably damaged.The preventive test for the diagnostic equipment with ultrasound is helpful to establish a quality control system for ultrasound,and to provide analysis materials with reliability in using diagnostic equipment with ultrasound,and to also provide decision-making basis for equipment departments and clinical department of using equipment in hospital.
8.Preliminary study on the implementation of HT for segmented TMLI in adults
Haiyang WANG ; Gaoxiang CHEN ; Xiang HUANG ; Ruigang GE ; Shilei ZHANG ; Changxin YAN ; Chuanbin XIE
China Medical Equipment 2025;22(9):21-26
Objective:To investigate the feasibility of segmented total marrow and lymphoid irradiation(TMLI)using Helical TomoTherapy(HT)in adults,and to analyze the impact of positional errors in the treatment on the actually received dose.Methods:Imaging data from four adult patients with acute myeloid leukemia were treated at the First Medical Center of the Chinese PLA General Hospital between November 2022 and December 2024 were retrospectively selected.The prescribed dose was 12 Gy in 3 fractions,and the heights range of patient was from 162 to 178 cm.The planning target volume(PTV)included the entire bone marrow and lymphatic drainage regions.The mid-femur was used as the dividing landmark,and the segmented location of head-in first(upper segmentation)and foot-in first(lower segmentation)were adopted to design plan for each segmentation.The upper segmentation used helical tomotherapy,and the lower segmentation used tomo direct therapy.The dosimetric parameters of planning target volume,and organ at risk(OAR)included eye lens,testicles and lungs were analyzed.The located computed tomography(CT)images of patients were used to simulate the changes of dosimetric parameters of each planning target region and OAR when the positions of patients moved 3,5,10 and 15 mm to outside of rack aperture.The changes of dose distribution of connection region of upper and lower segmentation were further analyzed.Results:The segmented irradiation technique can better achieve the cover(95%)of whole target region of TMLI under the condition of protecting normal organs.Under the condition with different position errors,the cover rate of the most of target region significant decreased.The largest dose of right lens increased 9.272%than primary under the 5 mm position error,and that of left and right lens increased respectively 40.199%and 67.898%under the 10 mm position error,and the max dose and mean dose of testicles increased respectively 5.847%and 5.240%under the 3mm position error.The mean dose of whole lung increased all under the error at different grades,which increased 3.021%under the 10mm position error.The non-uniformity of dose distribution of connection region increased after the position error was introduced,and the largest variation range reached to 4 Gy.Conclusion:The technique of segmented radiotherapy of Helical Tomotherapy has feasibility for TMLI of adult.But the error of the position of patients can reduce the cover of target region,and decrease the control rate for target region,and can lead to the occurrences of excessive doses on eye lens,testicles,lung and other OAR at the same time,which can increase the probability of occurring OAR injury.It should be high pay attention in treatment and implementation.
9.Advance on research of Flash-RT technology
Xiangkun DAI ; Shaojuan WU ; Jinyuan WANG ; Wei YU ; Lehui DU ; Changxin YAN ; Shilei ZHANG ; Na MA ; Xiao LEI ; Baolin QU
China Medical Equipment 2024;21(1):2-8
At present,precise radiotherapy has been widely used through the development with many years,but the existing technique still is limited by the limitation of tolerance dose of normal tissues,which cannot achieve the optimal goal of treating tumor.Flash radiotherapy(Flash-RT)is one kind of radiotherapy technique that uses the beam with ultra-high dose rate(UHDR)to conduct irradiation,which can furthest treat tumors while significantly reduce radiation injury of normal tissues.But until now,the biological mechanism,key physical parameters and triggering mechanism of Flash-RT are still unclear,and its principle and clinical translational application are still in the stage of research.This review clarified the technological advance and clinical translational application of Flash-RT research through summarized the relevant research of Flash-RT.
10.The Application of “Diagnosing Diseases with Sinews” in Obstinate Facial Paralysis
Yuyan WANG ; Xiyou WANG ; Duoduo LI ; Eryang LIU ; Baolin JIA ; Changxin LIU ; Yuanyuan GONG
Journal of Traditional Chinese Medicine 2024;65(6):652-656
The theory of “diagnosing diseases with sinews” means that through the diagnosis and examination of the channel sinews at the site of the lesion, the surplus and deficit state of qi and blood in the channel sinews, vessles and channels, and the degree of damage to the organism caused by the disease and evils, can be determined, forming the three elements (the nature of the disease, the location of the disease, and the disease tendency) of the disease diagnosis can be closely integrated to form the trinity of diagnostic modes, which is “examining the disease nature by sinews, identifying the pattern by sinews, and determining the tendency by sinews”. For intractable facial paralysis, the method of “diagnosing diseases with sinews” can be adopted, in which the morphological changes of the channel sinews are judged through diagnosis by observation, the traditional Chinese patterns are identified through diagnosis by palpation, and the points of meridian tendons and the circulation of tendon and treatment lines are determined through diagnosis by circulation. The “diagnosing diseases with sinews” not only helps to accurately determine the disease condition, patterns and development trend, but also helps to adopt targeted treatment for the disease and prevent the disease from spreading, and providing ideas and methods for the clinical diagnosis and treatment of intractable facial paralysis.