Chinese Journal of Radiological Health 2023;32(1):1-5
doi:10.13491/j.issn.1004-714X.2023.01.001
Absolute measurement of surface particle emission rates from α and β plane sources using a digitizer
Tianxiang LU 1 ; Qianqian ZHOU 2 ; Ming ZHANG 2 ; Fei TUO 1 ; Shuying KONG 1
Affiliations
Keywords
Digitizer; Metrology; Proportional counter; Plane source; Emission rate
Country
China
Language
Chinese
Abstract
Objective:To prove the validity and accuracy of the digitizer instead of the conventional electronics plug-in
for radionuclide measurement.
Methods: Based on a large-area flow-gas multi-wire proportional counter for 2πα and 2πβ surface particle emission rate measurement, the DT5730 digital waveform sampler developed by CAEN was used for waveform signal acquisition, amplitude analysis, and data processing of the α-plane source 241Am and the β-plane source nuclides 14C, 36Cl, and 90Sr-90Y of different energies.
Results:The deviations between the α and β surface particle emission rate results obtained after dead time and background corrections and the measurements obtained based on the plug-in calibrator were
all within 0.6%, within the uncertainty range, under consistent experimental conditions such as electronics threshold and high pressure.
Conclusion:The digitizer is an effective alternative to conventional electronics plug-ins for α and β signal acquisition and processing and the accurate measurement of α and β emission rates.