Chinese Journal of Vaccines and Immunization 2008;0(01):-
Comparison of Two Commercial Enzyme-linked Immunosorbent Assays for Detection of IgG Antibody to Diphtheria
Xin ZHAO ; Dongsheng REN ; Li XU
Keywords
Diphtheria; IgG antibody; ELISA
Country
China
Language
Chinese
Abstract
Objective To compare the performance of two commercially available ELISA kits for detection of diphtheria IgG antibody to provide data for the proper kit choice. Methods Thirty-three serum samples from staff of the capital airdrome and two standard sera were tested for the evaluation of the kits.Commercial ELISAs were compared with respect to their reproducibility(intra-assay variation and inter-assay variation)and accuracy.Results For the detection of standard serum,intra-assay variation(Coefficient of Variation,CV)was 13.5507% for IBL kit and 5.9541% for Virion kit,the inter-assay variation(CV)was 9.9890% for IBL and 2.4728% for Virion kit.For the serum sample,intra-assay variation(CV)was 14.5270% for IBL and 11.0612% for Virion,the inter-assay variation(CV)11.0039% for IBL and 8.1506% for Virion kit.For the detection of standard serum,inter-assay variation(CV)was 9.9890% for IBL and 2.4728% for Virion.The accuracy was analyzed by variation between the actual concentration and measured concentration of standard serum,the CV was 17.6889% for IBL and 1.3966% for Virion.Conclusion The intra-assay and inter-assay variation of the two kits are generally acceptable(CV