Drug Evaluation Research 2017;40(3):335-340

doi:10.7501/j.issn.1674-6376.2017.03.009

Determination of eight heavy metal elements in iron dextran by inductively coupled plasma optical emission spectrometry and correction of spectral interference

Chao MA ; Simeng FANG ; Nan MI ; Kexin ZANG ; Mujun SU

Keywords

inductively coupled plasma optical emission spectrometry; interference element correction; spectral interference; iron dextran; heavy metal elements

Country

China

Language

Chinese

Abstract

Objective To establish a method for the determination of eight heavy metal elements (Pb,Cd,As,Hg,Co,V,Se,and Mo) in iron dextran by inductively coupled plasma optical emission spectrometry (ICP-OES).Methods Through detection wavelength selection,optimization of instrument parameters and applying interference element correction (IEC) method to correction of spectral interference,the eight heavy metal elements were analyzed by ICP-OES.Results The elements recoveries were from 88.7% to 101.3% by correction of spectral interference with IEC method.The accuracy of the method was good.The linearities of the detected elements were good,and the correlation coefficients were all greater than 0.999.The detection limits were from 0.12 to 7.26 ng/rnL.The quantization limits were from 0.40 to 23.96 ng/mL.The precision of the method was good (RSD<3.5%,n=6).Conclusion The results of the spectral interference correction by IEC method are greatly superior to the results of conventional method.The established method is accurate,sensitive,and rapid,which can be applied to the determination of contents of eight heavy metal elements in iron dextran.