Journal of Laboratory Medicine and Quality Assurance 2011;33(1):43-47
Evaluation of two Real-Time PCR Assays for the Detection of Chlamydia trachomatis.
Eun Young LEE 1 ; Sinyoung KIM ; Hyon Suk KIM
Affiliations
Keywords
Chlamydia trachomatis; Real-time polymerase chain reaction; Comparison
Country
Republic of Korea
Language
Korean
MeSH
ACTIONS
ACTIONS
ACTIONS
ACTIONS
ACTIONS
Abstract
BACKGROUND: Chlamydia trachomatis is common etiological agent of sexual transmitted disease. We evaluated two real-time PCR assays for the detection of C. trachomatis; Cobas TaqMan CT test v2.0 (Roche Diagnostics, Switzerland) and AdvanSure CT/NG real-time PCR (LG Life Sciences, Korea). And we compared performance of two real-time PCR assays with Cobas Amplicor CT test (Roche Diagnostics, Switzerland). METHODS: Seventy-nine cervical swab specimens collected from patients with suspected C. trachomatis infection were tested using CTM CT, ADVANSURE, and AMPLICOR for the detection of C. trachomatis. RESULTS: Twenty-five (31.6%) samples showed positive results in all three assays. The positive and negative concordance rates among three assays were 100%. CONCLUSIONS: CTM CT and ADVANSURE showed excellent agreement with AMPLICOR. Therefore, these assays were reliable for the evaluation of C. trachomatis infection.