Journal of Modern Laboratory Medicine 2015;(1):159-163
doi:10.3969/j.issn.1671-7414.2015.01.048
Exploration of Rutual Recognition of Results of Protein and Lipid ExamInation among Various Clinical Laboratories
Dong LI ; Anyu BAO ; Lin SONG ; Zhen CHEN
Keywords
clinical chemistry; serum lipids; mutual recognition
Country
China
Language
Chinese
Abstract
Objective To explore the possibility and reliability of mutual recognition of renal indexs among 20 clinical labora-tories entitled with state key clinical laboratory,and to supply reference for future national mutual recognition of laboratory examination results.Methods Determined the concentrations of TP,ALB,TC,TG and Roche multiple calibrators and sub-mitted the results.The results were analyzed for robust Z score,percentage difference after calibration,bias at medical deci-sion level to observe the possibility and reliability.Results The bias of TG was out of the least allowable bias,thus they were not appropriate to mutual recognition.Conclusion It remains immature in the mutual recognition of lipids’determina-tions and much work needs to be done in field of internal quality control of the laboratory.