Improve ment of Lunar DPX-L Dual Energy X-ray Absorptiometry Equipment
Bin YANG ; Zhongqing YUAN ; Mei ZHANG
Keywords
dual energy X-ray absorptiometry system; DOS; baryeenter
Country
China
Language
Chinese
Abstract
Objective To solve the encountered problem of lack of space and dyskinesia of detector in dual energy X-ray absorptiometry system, Methods The method used PartitionMagic8.0 software to partition space in hard disk and added the load to keep the detector balance for moving, Results The original software can run in hard disk, and the pulleys are not gotten force from side. Conclusion The reason of problem is analyzed to find the method that can reduce the percentage for system failure and improve precision and accuracy of data.