Toxicological Research 2013;29(3):211-215
Investigation of Water Safety in Non-treated Drinking Water with Trace Toxic Metals.
Suw Young LY 1 ; Dae Hong KIM ; Ga Eun LEE
Affiliations
Keywords
Trace; Toxic metal; Voltammetry; Copper; Diagnosis
Country
Republic of Korea
Language
English
MeSH
ACTIONS
ACTIONS
ACTIONS
ACTIONS
ACTIONS
ACTIONS
ACTIONS
ACTIONS
ACTIONS
ACTIONS
ACTIONS
ACTIONS
ACTIONS
Abstract
The trace toxic metal copper was assayed using mercury immobilized on a carbon nanotube electrode (MCW), with a graphite counter and a reference electrode. In this study, a macro-scale convection motor was interfaced with a MCW three-electrode system, in which a handmade MCW was optimized using cyclic-and square-wave stripping voltammetry. An analytical electrolyte for tap water was used instead of an expensive acid or base ionic solution. Under these conditions, optimum parameters were 0.09 V amplitude, 40 Hz frequency, 0.01 V incremental potential, and a 60-s accumulation time. A diagnostic working curve was obtained from 50.0 to 350 microg/L. At a constant Cu(II) concentration of 10.0 microg/L, the statistical relative standard deviation was 1.78% (RSD, n = 15), the analytical accumulation time was only 60 s, and the analytical detection limit approached 4.6 microg/L (signal/noise = 3). The results were applied to non-treated drinking water. The content of the analyzed copper using 9.0 and 4.0 microg/L standards were 8.68 microg/L and 3.96 microg/L; statistical values R2 = 0.9987 and R2 = 0.9534, respectively. This method is applicable to biological diagnostics or food surveys.
备案号: 11010502037788, 京ICP备10218182号-8)