Toxicological Research  2013;29(3):211-215

Investigation of Water Safety in Non-treated Drinking Water with Trace Toxic Metals.

Suw Young LY 1 ; Dae Hong KIM ; Ga Eun LEE

Affiliations

+expand

Keywords

Trace; Toxic metal; Voltammetry; Copper; Diagnosis

Country

Republic of Korea

Language

English

Abstract

The trace toxic metal copper was assayed using mercury immobilized on a carbon nanotube electrode (MCW), with a graphite counter and a reference electrode. In this study, a macro-scale convection motor was interfaced with a MCW three-electrode system, in which a handmade MCW was optimized using cyclic-and square-wave stripping voltammetry. An analytical electrolyte for tap water was used instead of an expensive acid or base ionic solution. Under these conditions, optimum parameters were 0.09 V amplitude, 40 Hz frequency, 0.01 V incremental potential, and a 60-s accumulation time. A diagnostic working curve was obtained from 50.0 to 350 microg/L. At a constant Cu(II) concentration of 10.0 microg/L, the statistical relative standard deviation was 1.78% (RSD, n = 15), the analytical accumulation time was only 60 s, and the analytical detection limit approached 4.6 microg/L (signal/noise = 3). The results were applied to non-treated drinking water. The content of the analyzed copper using 9.0 and 4.0 microg/L standards were 8.68 microg/L and 3.96 microg/L; statistical values R2 = 0.9987 and R2 = 0.9534, respectively. This method is applicable to biological diagnostics or food surveys.