Journal of Laboratory Medicine and Quality Assurance 2002;24(2):221-224
Evaluation of the HM-JACK Automatic Analyzer for Fecal Occult Blood Test.
Seung Bok HONG 1 ; Hyun Soo KIM ; Hyo Soon PARK ; Do Hoon LEE
Affiliations
Keywords
Fecal occult blood test; Colorectal cancer; HM-JACK; OC-Hemodia
Country
Republic of Korea
Language
Korean
MeSH
ACTIONS
ACTIONS
ACTIONS
ACTIONS
ACTIONS
Abstract
BACKGROUND: Fecal occult blood test is a useful method for mass screening of colorectal cancer. HM-JACK is automatic analyzer for detecting fecal occult blood with latex agglutination methods. We evaluated its apparatus for compatibility for clinical testing. METHODS: The linearity, precision and effect of temperature and container in stool storage were evaluated. The comparison study with OC-Hemodia and prozone phenomenon were evaluated. RESULTS: The linearity was good(R(2)=0.997) and coeffiecinet variation(CV) of within-day precision were 4.6% and 2.6% in low concentration and high concentration. The CV of between-day precision were 4.8% and 3.5%. The hemoglobin concentration of 0.1 g/dL(1x10(6) ng/mL) was measured to 651 ng/mL for Antigen excess zone. The hemoglobin were more decreased in room temperature storage and usual container storage than in low temperature storage and HM-JACK container storage. The concordance rate between HM-JACK and OC-Hemodia was 90%. CONCLUSION: HM-JACK showed good performance for linearity, precision and comparison study. Therefore, it can be used effectively in clinical laboratory due to convenience, and possibility of quantitation and fast reporting.