Korean Journal of Medical Physics 2010;21(4):354-359
The Dosimetric Effect on Real PTV and OARs at Various Image Fusion Protocol for Pituitary Adenomas.
Kyung Nam LEE 1 ; Dong Joon LEE ; Tae Suk SUH
Affiliations
Keywords
Pituitary adenomas; Image fusion protocol; PTV coverage; Safety of OARs
Country
Republic of Korea
Language
Korean
MeSH
ACTIONS
ACTIONS
ACTIONS
ACTIONS
Abstract
The purpose of this study is to verify the dosimetric effect on real PTV (planning target volume) coverage and safety of OARs (organs at risk) at various image fusion protocol-based radiosurgery plan for pituitary adenomas. Real PTV coverage and its variation was acquired and maximum dose and the volume absorbing above threshold dose were also measured for verifying the safety of optic pathway and brainstem. The protocol that can reduce superior-inferior uncertainty by using both axial and coronal MR (magnetic resonance) image sets shows relatively lower values than that of case using only axial image sets. As a result, the image fusion protocol with both axial and coronal image sets can be beneficial to generate OAR-weighted radiosurgery plan.