Journal of the Korean Fracture Society  2012;25(2):136-141

doi:10.12671/jkfs.2012.25.2.136

Repeated Metal Breakage in a Femoral Shaft Fracture with Lateral Bowing: A Case Report.

Dong Soo KIM 1 ; Yong Min KIM ; Eui Sung CHOI ; Hyun Chul SHON ; Kyoung Jin PARK ; Byung Ki CHO ; Ji Kang PARK ; Hyun Cheol LEE ; Kyung Ho HONG

Affiliations

+expand

Keywords

Femur shaft fracture; Lateral bowing; Metal breakage; Locking compression plate; Intramedullary nail

Country

Republic of Korea

Language

Korean

Abstract

Fractures of the femoral shaft with marked bowing face some obstacles in fixation of the fracture such as difficulty in insertion of the intramedullary nail (IM nail) or exact contouring plate. Locking compression plates (LCP) are an option to manage this problem. However, we experienced consecutive breakage of LCP twice and IM nail once in an 80-year-old female. Finally, union of the fracture was achieved after fixation of the IM nail and additional plate together. Fractures of the femur shaft with marked bowing are thought to have different biomechanical properties; therefore, we present this case with a review of the literature.